Skip to content

PIRANHA 49700113Recalls & reported problems

Richard Wolf GmbH · ROTATIONS-MORCELLATOR Ø 4.8MM WL 335MM for MORCESCOPE, for morcellation following laser prostate enucleation, sterile · Class 2

All models of PIRANHA (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04055207063422), received between Jul 18, 2024 and Jul 18, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Misassembly by Users” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Misassembly by Users1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 18, 2024Misassembly by Users; Misassembly by Users

    RICHARD WOLF GMBH HAS BEEN INFORMED OF AN ISSUE REGARDING A ROTATIONS-MORCELLATOR Ø 4.8MM WL 335MM, PART ID: 49700113, LOT# UNKNOWN. ACCORDING TO THE RECEIVED INFORMATION, A HOLMIUM LASER ENUCLEATION OF THE PROSTATE (HOLEP) PROCEDURE AND MORCELLATION WAS PLANNED. FOR THIS PURPOSE, A POWER CONTROL 2303 (RICHARD WOLF), PIRANHA CONTROL 2208 (RICHARD WOLF), PIRANHA HANDPIECE (RICHARD WOLF) AND A HOLEP NEPHROSCOPE SYSTEM FROM KARL STORZ WAS PROVIDED IN THE OPERATING THEATRE. WHEN USING THE HOLEP NEPHROSCOPE SYSTEM FROM KARL STORZ, THE PIRANHA BLADE VMAX ROTARY MORCELLATOR 49700103 (RICHARD WOLF) MUST BE USED. INSTEAD, A RICHARD WOLF INSTRUMENTS, PIRANHA BLADE VMAX ROTATION MORCELLATOR, 49700113 WERE SUPPLIED. THE RICHARD WOLF BLADE SUPPLIED IS NOT COMPATIBLE FOR KARL STORZ INSTRUMENTS. THE ENUCLEATION WAS COMPLETED. HOWEVER, THE MORCELLATION WAS UNABLE TO BE COMPLETED, BECAUSE THE INCORRECT SIZE MORCELLATOR BLADE (49700113) WAS APPLIED. THERE WAS NO REPORTED HARM TO THE PATIENT. THE PROCEDURE WAS COMPLETED 10-DAYS LATER UNDER A SECOND GENERAL ANESTHETIC SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19773010 ↗