Skip to content

SHARK/S-LINE 46221313USRecalls & reported problems

Richard Wolf GmbH · CUTTING ELECTRODE BIPO 24FR 12/30° · Class 2

All models of SHARK/S-LINE (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04055207066027), received between Oct 3, 2024 and Oct 3, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Melted” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Melted1
  • Overheating of Device1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 3, 2024Melted; Overheating of Device; Melted; Overheating of Device

    RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) RECEIVED A MEDSUN REPORT: (B)(4), REGARDING AN ISSUE WITH A CUTTING ELECTRODE BIPO 24FR 12/30°, PART ID: 46221313US, LOT# 51018536. ACCORDING TO THE MEDSUN REPORT, "PROVIDER NOTED ISSUES WITH FOOT PEDAL FOR ERBE NOT CHARGING ELECTRODE, THEN WOULD HEAT SUDDENLY. THE TECH NOTED THAT ELEMENT BURNED "WHITE HOT". WHEN THE LENS CRACKED, SCOPE SET UP WAS REMOVED. THE TECH NEEDED A KELLY CLAMP TO UNSEAT ELECTRODE FROM RESECTOSCOPE. PATIENT EXPERIENCED AGGRESSIVE BILATERAL LEG MOVEMENTS DURING LOOP CAUTERY, REQUIRING MORE SEDATION. TECH NOTED WHAT APPEARED TO BE METAL FLUSH OUT OF PATIENT DURING PROCEDURE AND ASKED PROVIDER; PROVIDER BELIEVED IT MAY HAVE BEEN FROM UROCLIPS THAT WERE BEING REMOVED DURING PROCEDURE. ONCE REMOVED, UROCLIPS APPEARED TO BE COMPLETELY INTACT." ORIGINAL INTENDED PROCEDURE: CYSTOSCOPY WITH TRANSURETHRAL PROSTATECTOMY, CYSTOSCOPY LITHOLAPAXY.

    Patient problems: Insufficient Information

    MAUDE report 20358712 ↗

Other HYSTEROSCOPE (AND ACCESSORIES) devices

All HYSTEROSCOPE (AND ACCESSORIES) devices →