SHARK/S-LINE 46221313USRecalls & reported problems
Richard Wolf GmbH · CUTTING ELECTRODE BIPO 24FR 12/30° · Class 2
All models of SHARK/S-LINE (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04055207066027), received between Oct 3, 2024 and Oct 3, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Melted” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Melted1
- Overheating of Device1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 3, 2024Melted; Overheating of Device; Melted; Overheating of Device
RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) RECEIVED A MEDSUN REPORT: (B)(4), REGARDING AN ISSUE WITH A CUTTING ELECTRODE BIPO 24FR 12/30°, PART ID: 46221313US, LOT# 51018536. ACCORDING TO THE MEDSUN REPORT, "PROVIDER NOTED ISSUES WITH FOOT PEDAL FOR ERBE NOT CHARGING ELECTRODE, THEN WOULD HEAT SUDDENLY. THE TECH NOTED THAT ELEMENT BURNED "WHITE HOT". WHEN THE LENS CRACKED, SCOPE SET UP WAS REMOVED. THE TECH NEEDED A KELLY CLAMP TO UNSEAT ELECTRODE FROM RESECTOSCOPE. PATIENT EXPERIENCED AGGRESSIVE BILATERAL LEG MOVEMENTS DURING LOOP CAUTERY, REQUIRING MORE SEDATION. TECH NOTED WHAT APPEARED TO BE METAL FLUSH OUT OF PATIENT DURING PROCEDURE AND ASKED PROVIDER; PROVIDER BELIEVED IT MAY HAVE BEEN FROM UROCLIPS THAT WERE BEING REMOVED DURING PROCEDURE. ONCE REMOVED, UROCLIPS APPEARED TO BE COMPLETELY INTACT." ORIGINAL INTENDED PROCEDURE: CYSTOSCOPY WITH TRANSURETHRAL PROSTATECTOMY, CYSTOSCOPY LITHOLAPAXY.
Patient problems: Insufficient Information
MAUDE report 20358712 ↗
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page