Skip to content

Disposable Radiolucent Skull Pins (3 Pcs) 18910-01ARecalls & reported problems

Brainlab SE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04056481004767), received between May 30, 2018 and May 30, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure To Adhere Or Bond” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure To Adhere Or Bond1
  • Use of Device Problem1

Patient problems

  • Laceration(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 30, 2018Failure To Adhere Or Bond; Use of Device Problem; Failure To Adhere Or Bond; Use of Device Problem

    A PATIENT WAS FIXATED IN THE PMI DORO RADIOLUCENT SKULL CLAMP USING BRAINLAB RADIOLUCENT PINS DURING A CRANIAL PROCEDURE (A TRANSMASTOID CRANIOTOMY FOR RESECTION OF AN ENDOLYMPHATIC SAC TUMOR). THE SURGEON EXPERIENCED DIFFICULTIES IN PLACING THE PINS, ESP. TO VERIFY THAT THE PIN TIPS WERE IN COMPLETE CONTACT WITH THE SKULL, DUE TO A SIGNIFICANT AMOUNT OF FAT IN THE SUBCUTANEOUS LAYER OF THE PATIENT'S SCALP. SEVERAL TABLE MANEUVERS WERE PERFORMED TO CONFIRM THAT THERE WAS NO MOVEMENT OF THE PATIENT HEAD, AND STABILITY OF THE FIXATION WAS DETERMINED SATISFACTORY FOR THE PROCEDURE. DURING THE CASE, AFTER APPROX. 7H, WHILE THE SURGICAL RESECTION CAVITY WAS BEING INSPECTED FOR ADDITIONAL TUMOR (SURGEON WAS INSIDE WITH FORCEPS, AND CRITICAL BRAIN STRUCTURES WERE EXPOSED), THE PATIENT HEAD JERKED IN A SINGLE, RAPID MOVEMENT. THE HEAD SLIPPED OUT OF THE TWO PINS ON THE FOREHEAD. THIS CAUSED A LACERATION OF 1-2 CM ON THE RIGHT HAND SIDE WHICH NEEDED STITCHES, THE PUNCTURE ON THE LEFT HAND SIDE WAS LARGER THAN USUAL BUT DID NOT REQUIRE STITCHES. THE PATIENT HEAD COULD BE STABILIZED, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. ACCORDING TO THE HOSPITAL, THE SURGERY WAS COMPLETED SUCCESSFULLY,

    Patient problems: Laceration(s)

    MAUDE report 7552512 ↗

Other Holder, Head, Neurosurgical (Skull Clamp) devices

All Holder, Head, Neurosurgical (Skull Clamp) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page