Skip to content

iGUIDE P10603-110Recalls & reported problems

3C-Medical Intelligence GmbH · iGUIDE Software 2.2.1 is the software which controls the movement of the HexaPOD™ evo Module and monitors the correct position of the module. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04056719001742), received between Jun 12, 2017 and Jun 12, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

2 FDA recall records name this device's identifier.

1 other recall cover devices cleared under the same FDA submission (K143485); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jul 18, 2019Terminated

    Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.

    Z-2082-2019 · Root cause: Software design · FDA record ↗

  2. Jun 19, 2019Terminated

    Potential for iGUIDE to incorrectly monitor the 3D position.

    Z-2230-2019 · Root cause: Software Design Change · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2017Computer Software Problem; Computer Software Problem

    THE MANUFACTURER DISCOVERED DURING AN IN-HOUSE INVESTIGATION THAT IF THE PEC DATA DIALOG AFTER A VERIFICATION SCAN IS CANCELLED, A RE-SEND OF PEC DATA WILL LEAD TO INCORRECT POSITIONING BECAUSE THE 2ND PEC VALUES SUPERSEDE THE 1ST PEC VALUES INSTEAD OF BEING ADDED.

    Patient problems: No Patient Involvement

    MAUDE report 6633208 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K143485. One submission often covers many models, so check the product and lot details in each record.

  • Z-3017-2017 · Jun 19, 2017 · Terminated
    iGUIDE Software 2.2 Product Usage: Control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.

Other Accelerator, Linear, Medical devices

All Accelerator, Linear, Medical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page