Skip to content

Signia Siemens INSIO 1PXRecalls & reported problems

Ws Audiology Usa, Inc. · Signia Siemens INSIO 1PX Hearing Aid · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04056839035788), received between Jun 11, 2024 and Jun 11, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 11, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    IN THIS EVENT, CUSTOMER REPORTED A SHARP LOUD SOUND FROM LEFT HEARING AID WHEN HE WAS INSERTING THE AID. THIS WAS FOLLOWED BY PAIN IN LEFT SIDE OF HEAD AND ONGOING TINNITUS. CUSTOMER WAS ASSESSED BY HEARING CARE PROFESSIONAL (04/29/2024) AND LEFT EAR WAS SHOWING SOME MINOR CONDUCTIVE DECLINE, AS COMPARED TO HIS TESTING IN MARCH. CLINICIAN CONDUCTED HEARING ASSESSMENT AND TYMPANOMETRY REVEALED TYPE B TYMPANOGRAMS, CONSISTENT WITH A CONDUCTIVE COMPONENT TO HEARING LOSS. ALSO OBSERVED WAS AN ACTIVE EAR INFECTION AND THE CUSTOMER WAS PRESCRIBED ANTIBIOTICS TO TREAT INFECTION; CONSISTENT WITH THE TYMPANOMETRIC RESULTS. THE MAXIMUM OUTPUT SPL (SOUND PRESSURE LEVEL) FROM THE DEVICE IS LIMITED BY SAFETY LIMITER. RESULTS OF TECHNICAL INVESTIGATION: LEFT HEARING AID WAS RETURNED FOR INVESTIGATION. BASED ON ACOUSTIC TEST RESULT, HEARING AID IS FUNCTIONING WITHIN SPECIFICATION. THE MAXIMUM GAIN THAT THE HA CAN GIVE IS ONLY 52.1 DECIBEL AND MAX OSPL90 IS 111.5 DECIBEL, WHICH IS WITHIN THE DEVICE SPECIFICATION. THE MAX OUTPUT LEVEL IS LOW BY INDUSTRY STANDARDS AND WOULD NOT BE ASSOCIATED WITH A CONDUCTIVE HEARING LOSS. INTERNAL AUDIOLOGIST EVALUATION: THE TYPE B TYMPANOGRAM (CONSISTENT WITH A CO

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19510450 ↗