Skip to content

SIL. COATED I/A 45 MIL 1/B 85910S EARecalls & reported problems

Bausch & Lomb GmbH · Single Use Medical Device for Ophthalmic Surgery · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04058935000356), received between Mar 30, 2026 and Mar 30, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • Capsular Bag Tear1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 30, 2026Material Deformation

    THE USER FACILITY REPORTED THAT THAT THEY HAD AN ISSUE IN THE PAST WITH THE METAL ASPIRATION TUBE COMING OUT OF THE IRRIGATION HOLE IN THE HANDPIECE, CAUSING A CAPSULE BAG BREAK. THEY DID NOT REALIZE IT UNTIL AFTER THE MAIN USAGE WHEN THEY WENT BACK INTO REMOVE VISCOELASTIC. THIS EVENT WAS FROM A COUPLE YEARS AGO AND IT WAS JUST LEARNED ABOUT. THE DATE OF EVENT WAS NOT AVAILABLE. THE DOCTOR REMEMBERS THAT THERE WAS NO SIGNIFICANT DELAY, NO ADDITIONAL ANESTHESIA REQUIRED, AND THAT THE PATIENT WAS "FINE". NO FURTHER DETAILS ARE AVAILABLE.

    Patient problems: Capsular Bag Tear

    MAUDE report 24719785 ↗

Other Unit, Phacofragmentation devices

All Unit, Phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page