SIL. COATED I/A 45 STELLARIS 12/B 85910STRecalls & reported problems
Bausch & Lomb GmbH · Single Use Medical Device for Ophthalmic Surgery · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04058935000363), received between Nov 13, 2023 and May 13, 2025 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Defective Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Component4
- Product Quality Problem2
- Material Too Soft/Flexible1
- Particulates1
Patient problems
- Capsular Bag Tear5
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2025Particulates; Particulates
A USER FACILITY REPORTED THAT DURING CATARACT SURGERY, THE SURGEON NOTICED A WHITE SUBSTANCE ON THE TIP OF AN I/A 45 DEGREE HANDPIECE 85910ST. CURRENTLY, IT IS NOT KNOWN IF THERE IS ANY HARM TO THE PATIENT. THE DOCTOR WILL BE WATCHING THIS PATIENT CAREFULLY IN FOLLOW UP VISITS. THERE WERE NO DELAYS IN SURGERY NOR ADDITIONAL ANESTHESIA PROVIDED TO THE PATIENT. THE SURGERY WAS COMPLETED. THREE PHOTOS WERE PROVIDED, AND THE SUBSTANCE WAS SENT TO THE LAB FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22015485 ↗InjuryOct 3, 2024Defective Component; Defective Component
IT WAS REPORTED BY THE USER FACILITY AT THE VA IN (B)(6) THAT THE METAL TIP PROTRUDED OUT OF THE SILICONE AND MAY HAVE BEEN RESPONSIBLE FOR A PUNCTURE WOUND. THE PATIENT HAD A PUNCTURE TO THE BAG DURING A SURGERY ON (B)(6) 2024, WITH DR. (B)(6); OD (PHACO IOL WITH ANTERIOR VITRECTOMY). UPON ENTRY INTO THE INCISION, THE METAL TIP PROTRUDED OUT OF THE SILICONE SHEATH AND THE PUNCTURED THE POSTERIOR CAPSULE, RESULTING IN THE NEED TO PERFORM ANTERIOR VITRECTOMY. THE HANDPIECE WAS DISCARDED. ONE SUTURE WAS PLACED DUE TO ENLARGEMENT OF THE WOUND TO ACCOMMODATE A 3 PIECE IOL IN THE SULCUS DUE TO THE POSTERIOR CAPSULAR TEAR AND VITREOUS PROLAPSE. SURGERY WAS LONGER THAN USUAL. AFTER WE STARTED TOPICAL MOXIFLOXACIN, PREDNISOLONE AND DICLOFENAC DUE TO HIGH RISK OF ENDOPHTHALMITIS AND CYSTOID MACULA EDEMA FROM POSTERIOR CAPSULAR TEAR. PATIENT HAD A FOLLOW UP THREE DAYS AFTER SURGERY AND IS DOING WELL, VISION IS IMPROVING.
Patient problems: Capsular Bag Tear
MAUDE report 20361010 ↗InjurySep 4, 2024Defective Component; Defective Component
IT WAS REPORTED BY THE USER FACILITY IN THE UNITED KINGDOM THAT THE SLEEVES ARE MOVING AND THAT HAD CAUSED A POTERIOR CAPSULE RUPTURE. PROCEDURE CARRIED OUT SUCCESSFULLY WITH NO MEDICAL INTERVENTION REQUIRED. THE PATIENT HAD A SUCCESSFUL PROCEDURE AND IS HAPPY WITH OUTCOME.
Patient problems: Capsular Bag Tear
MAUDE report 20139357 ↗InjuryMay 23, 2024Material Too Soft/Flexible; Material Too Soft/Flexible
THE USER FACILITY REPORTED THAT THE CAPSULEGUARD WAS BUNCHING UP CAUSING IT TO GRAB THE BAG. THE SURGEON HAD TO PERFORM A VITRECTOMY, REQUIRING ADDITIONAL CASE TIME AND STITCHES TO CLOSE THE WOUND.
Patient problems: Capsular Bag Tear
MAUDE report 19381894 ↗InjuryNov 13, 2023Product Quality Problem; Defective Component; Product Quality Problem; Defective Component
A USER FACILITY IN THE UNITED KINGDOM REPORTED THAT DURING SURGERY THE SLEEVE IS COLLAPSING AND STICKING THE INNER ASPIRATION TUBE AT THE CORNEAL INCISION SITE. THERE WAS A SUDDEN ADVANCEMENT AND A POSTERIOR CAPSULE RUPTURE OCCURRED DURING IRRIGATION AND ASPIRATION. THE PROCEDURE WAS CHANGED AND A VITRECTOMY WAS PERFORMED ON THE PATIENTS EYE. SURGERY WAS INCREASED 15-30 MINUTES.
Patient problems: Capsular Bag Tear
MAUDE report 18122154 ↗InjuryNov 13, 2023Product Quality Problem; Defective Component; Product Quality Problem; Defective Component
A USER FACILITY IN THE UNITED KINGDOM REPORTED THAT DURING SURGERY THE SLEEVE IS COLLAPSING AND STICKING THE INNER ASPIRATION TUBE AT THE CORNEAL INCISION SITE. THERE WAS A SUDDEN ADVANCEMENT AND A POSTERIOR CAPSULE RUPTURE OCCURRED DURING IRRIGATION AND ASPIRATION. THE PROCEDURE WAS CHANGED AND A VITRECTOMY WAS PERFORMED ON THE PATIENTS EYE. SURGERY WAS INCREASED 15-30 MINUTES.
Patient problems: Capsular Bag Tear
MAUDE report 18118852 ↗
Other Unit, Phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page