Skip to content

MICS SIL I/A HDP 45° STELL 12B 85915STRecalls & reported problems

Bausch & Lomb GmbH · Single Use Medical Device for Ophthalmic Surgery · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04058935000516), received between Sep 25, 2023 and Feb 19, 2026 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Particulates” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Particulates3
  • Adverse Event Without Identified Device or Use Problem2
  • Detachment of Device or Device Component1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Capsular Bag Tear2
  • Endophthalmitis1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2026Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY IN AUSTRALIA REPORTED DURING THE POLISHING PORTION OF A CATARACT PROCEDURE, POSTERIOR CAPSULE (PC) TEAR OCCURRED. DUE TO THE PC TEAR AND THE NEED FOR VITRECTOMY, SURGERY WAS EXTENDED BY APPROXIMATELY 20 MINUTES, REQUIRING EXTRA TOPICAL ANESTHETIC. NO EXTRA MEDICATIONS OR TREATMENTS WERE NEEDED. PATIENT WAS REVIEWED THE FOLLOWING DAY, (B)(6) 2026, AND NO ADVERSE OUTCOMES WERE REPORTED.

    Patient problems: Capsular Bag Tear

    MAUDE report 24391095 ↗
  • MalfunctionOct 25, 2024Particulates; Particulates

    THE USER FACILITY IN THE UNITED KINGDOM REPORTED THE SURGEON NOTICED FIBERS FOUND ON THE I/A HANDPIECE WERE IN THE ANTERIOR CHAMBER. NO MEDICAL INTERVENTION/TREATMENT REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20532335 ↗
  • MalfunctionOct 15, 2024Particulates; Particulates

    THE USER FACILITY IN UNITED KINGDOM REPORTED PLASTIC FIBERS WERE COMING OFF THE I/A HANDPIECES (BLUE PLASTIC TIPS) ON TWO PATIENTS. NO MEDICAL INTERVENTION/TREATMENT WAS REQUIRED OTHER THAN REMOVING THE FIBERS FROM THE PATIENTS EYE. THE SURGERY INCREASE DURATION WAS 0-15 MINUTES. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20450925 ↗
  • MalfunctionOct 15, 2024Particulates; Particulates

    THE USER FACILITY IN UNITED KINGDOM REPORTED PLASTIC FIBERS WERE COMING OFF THE I/A HANDPIECES (BLUE PLASTIC TIPS) ON TWO PATIENTS. NO MEDICAL INTERVENTION/TREATMENT WAS REQUIRED OTHER THAN REMOVING THE FIBERS FROM THE PATIENTS EYE. THE SURGERY INCREASE DURATION WAS 0-15 MINUTES. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20450875 ↗
  • InjurySep 28, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE USER FACILITY IN THE UNITED KINGDOM REPORTED A PATIENT DEVELOPED AN INFECTION IN THEIR EYE AND DIAGNOSED WITH ENDOPHTHALMITIS. MEDICAL INTERVENTION: POST OPERATION MEDICATION.

    Patient problems: Endophthalmitis

    MAUDE report 17838556 ↗
  • InjurySep 25, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY IN THE UNITED KINGDOM REPORTS THAT A SPLIT WAS IDENTIFIED IN THE TIP OF THE IA HANDPIECE FOLLOWING A CATARACT PROCEDURE THAT RESULTED IN A POSTERIOR CAPSULAR RUPTURE. PATIENT RECEIVED AN ANTERIOR VITRECTOMY.

    Patient problems: Capsular Bag Tear

    MAUDE report 17806473 ↗

Other Unit, Phacofragmentation devices

All Unit, Phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page