Skip to content

ERBEFLO CleverCap® 20325-207Recalls & reported problems

Erbe Medical, LLC · Hybrid CO2 Tubing/Cap Set for Pentax® Scopes · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04065655000689), received between Jun 8, 2026 and Jun 8, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 31, 2026Open, Classified

    Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

    Z-2062-2026 · Root cause: Component change control · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 8, 2026Leak/Splash

    IT WAS REPORTED THAT THE HYBRID CO2 TUBING/CAP SET FOR PENTAX® LEAKED FROM THE IRRIGATION LINE. NO INFORMATION WAS PROVIDED IN REGARDS TO THE EQUIPMENT OR ANY OTHER ACCESSORY USED WHEN THE SET LEAKED. THERE WAS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25434081 ↗

Other Pump, Air, Non-Manual, For Endoscope devices

All Pump, Air, Non-Manual, For Endoscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page