ERBEFLO CleverCap recallZ-2062-2026 · Erbe Medical, LLC · initiated Mar 31, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Erbe Medical, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
FDA root cause category: Component change control
What the company is doing
On 3/31/2026, correction notices were emailed to Medical Directors, Risk Managers, Medical Device Safety Officers, or Healthcare Professionals informing them to take the following actions: 1) Discontinue use of unsupported third-party Scope Port Connectors. 2) Provide this notice to all personnel within your organization who use, service, purchase, stock, or otherwise manage. 3) If you further distributed any affected devices, promptly forward this notice to those accounts and request that they take the same actions. 4) Continue following the instructions for use, including verifying connector compatibility, ensuring that all connectors are fully seated, confirming that all joints and interfaces are secure, and verifying correct placement and alignment of the tubing within the pump head. 5) Utilize the compatibility table in the notice until updated Notes on Use is received. 6) Complete and return the response form via email to [email protected]. Firm contacts: Correction: [email protected], 480-564-1899. Routine Customer Service and Complaints: [email protected], 1-800-778-3723
Recall details
- Recall number
- Z-2062-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 31, 2026
- Posted by the FDA
- May 1, 2026
- Recalling firm
- Erbe Medical, LLC
- Quantity
- 260,962
- Product code
- FEQ
- FDA submissions
- K132340, K162152
- Distribution
- US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
Product description
ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
REF/UDI-DI/Lots: 20325-239/04065655000375/W4464506, W4464507, W4464512, W4464514, W4464515, W4464516, W4464517, W4465114, W4465115, W4465116, W4465702, W4465709, W4465991, W4466004, W4466266, W4466267, W4466268, W4466269, W4466271, W4466272, W4466273, W4466275, W4466278, W4466279, W4466813, W4466814, W4466815, W4466816, W4467057, W4467058, W4467059, W4467680, W4467681, W4468147, W4468148, W4468273, W4468274, W4468275, WO468776, WO468777, WO468778, WO468779, WO468780, WO469443, WO469444, WO469448, WO469449, WO469450, WO469451, WO469452, WO469933, WO469934, WO469947, WO469948, WO469949, WO469950, WO469951, WO469952, WO470768, WO470774, WO470775, WO470776, WO470777, WO470778, WO470896, WO470897, WO470913, WO470914, WO470915, WO470922, WO471236, WO471237, WO471238, WO471239, WO471240, WO471241, WO471242, WO473987, WO473988, WO473989, WO473991, WO475246, WO475247, WO475248, WO475249; 20325-240/04065655000405/W4466020, W4466021, WO468768, WO468769, WO468770, WO468771, WO469445, WO469446, WO469447, WO469936, WO469937, WO471244, WO471246, WO473170, WO473171, WO473172, WO473173, WO473174, WO473982, WO473983, WO473984, WO473985, WO474409, WO474410, WO474411, WO474412, WO474413, WO474414, WO475243; 20325-206/04065655000436/W4467039, W4467049, W4467678, W4467679, WO468582, WO470769, WO470770, WO470771, WO470772, WO470773, WO471076, WO471077, WO471078, WO471079, WO471390, WO478365, WO478366, WO478367; 20325-207/04065655000689/W4469939, W4469940, W4469941; 20325-248/04065655001006/W4467050, W4467051, W4467055, W4467056, W4467682, WO470898, WO470899, WO470900, WO470901, WO470902, WO470903, WO471252, WO471253, WO471254, WO471255, WO471256; 20325-208/04065655001099/W4467040, W4468271, WO468760
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- ERBEFLO CleverCap® 20325-239Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall24 reports
- ERBEFLO CleverCap® 20325-206Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall5 reports
- ERBEFLO CleverCap® 20325-240Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall3 reports
- ERBEFLO CleverCap® 20325-248Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall3 reports
- ERBEFLO CleverCap® 20325-207Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall1 report
- ERBEFLO CleverCap® 20325-208Erbe Medical, LLC · Pump, Air, Non-Manual, For Endoscope1 recall1 report
Product lines
Other devices under the same FDA submission
1 device cleared under K132340, K162152 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ERBE 20325-239Erbe Elektromedizin GmbH · PUMP, AIR, NON-MANUAL, FOR ENDOSCOPE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98674).
- Z-2063-2026 · Open, ClassifiedERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;
- Z-2064-2026 · Open, ClassifiedERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page