Catapult Guide Sheath 5F, 90cm, straight, hydrophilic coatedRecalls & reported problems

Contract Medical International GmbH · Introducer Sheath System containing One (1) introducer Sheath with haemostasis valve and One (1) dilator. Introducer Sheath is hydrophilic coated on the distal end. · Class 2

All models of Catapult Guide Sheath (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04251244502072), received between Sep 24, 2024 and Sep 24, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Material Separation1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2024Fracture; Material Separation; Fracture; Material Separation

    THE PROBLEM WAS DESCRIBED AS: "I DID NOTICE PHYSICIAN STRUGGLING TO ADVANCE SHEATH UPON INSERTION. WE DID NOT SEE ANY OBVIOUS ANATOMICAL CHALLENGES ON ANGIO. PHYSICIAN PERFORMED INTERVENTION SUCCESSFULLY. UPON REMOVAL, PHYSICIAN SAID THE SHEATH BECAME "STUCK" AND HE HAD TO USE FORCE TO PULL IT OUT AND THEN THE PLASTIC PORTION TORE AND THE BRAIDED PORTION BEGAN TO UNRAVEL. THE DOCTOR DECIDED TO USE A SCALPEL TO CUT THE SHEATH ABOVE THE DAMAGED PORTION IN ORDER TO INSERT A WIRE TO KEEP ACCESS. HE SUCCESSFULLY DID THIS AND WAS ABLE TO PUT A SHORT 7 FRENCH SHEATH IN. NO COMPLICATIONS, NOTHING LEFT INSIDE PATIENT. MINIMAL BLOOD LOSS." WHAT TROUBLESHOOTING STEPS, IF ANY, RESOLVED THE ISSUE? HE TRIED TO INSERT A BALLOON IN TO TRAP HIS WIRE SO HE WOULDN'T LOSE ACCESS. THAT DID NOT WORK. PATIENT PRESENT AT TIME OF EVENT? YES, PATIENT COMPLICATIONS: HEMATOMA, BLOOD LOSS / HEMORRHAGE IN THE PHYSICIAN'S OPINION, DID THE DEVICE OR PROCEDURE CAUSE OR CONTRIBUTE TO THE PATIENT COMPLICATION? YES, PLEASE DESCRIBE THE WAY IN WHICH THE DEVICE OR PROCEDURE CAUSED OR CONTRIBUTED TO THE PATIENT COMPLICATION? PHYSICIAN SAID IT LOST ITS INTEGRITY AND BEGAN TO UNRAVEL. MEDICAL OR SURGICAL INTERVENTIONS: PH

    Patient problems: Hemorrhage/Blood Loss/Bleeding