Skip to content

Biotronik Fortress Introducer Sheath System 6F; 45 cm; crossoverRecalls & reported problems

Contract Medical International GmbH · Introducer Sheath System containing One (1) introducer Sheath with haemostasis valve and One (1) dilator. · Class 2

All models of Biotronik Fortress Introducer Sheath System (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04251244503727), received between Oct 25, 2024 and Feb 24, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K153197); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Difficult to Remove1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Fluid/Blood Leak

    EVENT DETAILS: DURING THE PROCEDURE, IT WAS OBSERVED THAT THE HEMOSTATIC VALVE DID NOT SEAL COMPLETELY. WHEN ADDITIONAL GUIDEWIRES, BALLOONS, OR OTHER INSTRUMENTS WERE INTRODUCED, EXCESSIVE BLOOD FLOW OCCURRED. AS THIS ISSUE HAS NOW APPEARED FOR THE THIRD TIME WITHIN THE SAME LOT, THE ENTIRE LOT HAS BEEN COLLECTED FOR FURTHER INSPECTION. PRODUCTS THAT WERE USED INSIDE PATIENTS WILL NOT BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434372 ↗
  • MalfunctionOct 25, 2024Break; Difficult to Remove; Break; Difficult to Remove

    THE 6FR FORTRESS SHEATH WAS ATTEMPTED TO BE REMOVED BUT WAS FOUND TO HAVE RESISTANCE. THE SHEATH WAS NOTED TO HAVE BROKEN WHILE IN THE VESSEL. DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO WORK OR STOPPED WORKING). PATIENT INJURY: NO. DID DEVICE PARTS REMAIN IN THE PATIENT? NO. EVENT OCCURRED: INSIDE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20531400 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K153197. One submission often covers many models, so check the product and lot details in each record.

  • Z-2532-2025 · Jul 30, 2025 · Open, Classified
    Brand Name: Fortress Introducer Sheath System Product Name: Fortress Introducer Sheath System Model/Catalog Number: 386594 Product Description: Fortress Introducer Sheath System, 6

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page