Impella 0052-3018Recalls & reported problems
Abiomed, Inc. · Lake Region 0.027" PTFE Guidewire · Class 3
All models of Impella Heart Pump (34)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04260113630327), received between Jan 14, 2026 and Mar 30, 2026 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Insert2
- Patient Device Interaction Problem2
- Adverse Event Without Identified Device or Use Problem1
- Device Damaged Prior to Receipt by User1
- Difficult to Advance1
- Failure to Advance1
- Malposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Vascular Dissection2
- Great Vessel Perforation1
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2026Malposition of Device; Patient Device Interaction Problem
AN IMPELLA CP DEVICE WAS INSERTED INTO THE RIGHT FEMORAL ARTERY AND AN IMPELLA RP FLEX WAS INSERTED INTO THE RIGHT FEMORAL VEIN IN A 64-YEAR-OLD MALE PATIENT PRESENTING IN AN OUT-OF-HOSPITAL CARDIAC ARREST REQUIRING CARDIOPULMONARY RESUSCITATION (CPR), IN ACUTE MYOCARDIAL INFARCTION (AMI) AND CARDIOGENIC SHOCK (CGS), AND IN SCAI STAGE E SHOCK, ON MULTIPLE INOTROPES AND VASOPRESSORS, AND ON A VENTILATOR FOR RESPIRATORY SUPPORT, PRIOR TO INITIATION OF SUPPORT. THE IMPELLA WAS INSERTED FOR VENTRICULAR SUPPORT FOR A HIGH-RISK PERCUTANEOUS CORONARY INTERVENTION (PCI). DURING THE PROCEDURE, ACCESS WAS OBTAINED TO THE RIGHT JUGULAR VEIN BUT WAS LOST. THE SHEATH AND GUIDEWIRE CAME OUT. THE PHYSICIAN ATTEMPTED TO REGAIN ACCESS BUT WAS UNSUCCESSFUL. ACCESS WAS THEN OBTAINED IN THE RIGHT FEMORAL VEIN, AND THE RP FLEX WAS INSERTED. THIRTY SECONDS LATER, THE PATIENT'S PRESSURES BEGAN TO DROP, AND WHEN VERIFYING POSITION, THE OUTLET CAGE SLIPPED INTO THE RIGHT VENTRICLE. SEVERAL ATTEMPTS WERE MADE TO PUSH THE RP FLEX BACK ACROSS THE PULMONARY ARTERY VALVE, WITHOUT SUCCESS. THE PATIENT CONTINUED TO CODE, AND THE PATIENT EXPIRED. THE DEVICE IS UNLIKELY TO HAVE CONTRIBUTED TO THE PATIENT'S DEATH, W
Patient problems: Vascular Dissection
MAUDE report 24724023 ↗MalfunctionMar 18, 2026Difficult to Insert
THIS IMPELLA GUIDEWIRE DEVICE REPORT IS ONE OF TWO DEVICES ASSOCIATED WITH THE EVENT. REFER TO THE RELATED MANUFACTURER REPORT NUMBER AS NOTED IN SECTION H10 FOR THE MEDICAL DEVICE REPORT ON THE IMPELLA RP FLEX PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24629598 ↗InjuryMar 2, 2026Failure to Advance
AN IMPELLA CP WAS INSERTED IN A 65 YEAR OLD MALE ADMITTED FOR ACUTE MYOCARDIAL INFARCTION WITH CARDIOGENIC SHOCK. THE PATIENT¿S COMORBIDITIES WERE UNKNOWN. THE PATIENT'S CLINICAL STATE PRIOR TO INITIATION OF SUPPORT WAS SCAI STAGE E. THE PATIENT WAS BRIDGED TO 5.5 FOR AORTIC VALVULOPLASTY. ON DAY 7 OF 5.5 SUPPORT, AN ATTEMPT WAS MADE TO ESCALATE THE PATIENT TO BIPELLA SUPPORT. AN IMPELLA RP FLEX WAS ATTEMPTED VIA RIGHT FEMORAL VEIN. THE DEVICE WAS UNABLE TO PASS DUE TO THE PATIENT'S VASCULATURE, WHICH APPEARED TO CAUSE THE WIRE TO PROLAPSE AND POSSIBLE DISSECTION OF THE VEIN. IT WAS REPORTED A TEMPORARY PACER WAS ALSO PREVIOUSLY PLACED IN THAT VESSEL. THE RP FLEX IMPLANT WAS ABORTED. VASCULAR INJURY IS A KNOWN POTENTIAL ADVERSE EVENT OF THE IMPELLA RP FLEX PER THE INSTRUCTIONS FOR USE. THE COMPLAINANT HAS ALSO REPORTED THAT THE DEVICE HAS SINCE BEEN DISCARDED AND THAT THE PATIENT'S OUTCOME WAS GOOD. THE DEVICES WERE EXPLANTED MONDAY 2/22 AND THE PATIENT WAS DOWNGRADED OUT OF THE ICU.
Patient problems: Vascular Dissection
MAUDE report 24483352 ↗MalfunctionFeb 23, 2026Device Damaged Prior to Receipt by User
ADDITIONAL INFORMATION INDICATES THAT THE 027 WIRE WAS NOT USED AND WAS SHIPPED BACK IN A RETURN KIT. THE ABIOMED 018 WIRE WAS USED INSTEAD FOR SUCCESSFUL DELIVERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24424588 ↗InjuryFeb 5, 2026Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem
CLINICIAN NARRATIVE: AN IMPELLA RP FLEX WITH SMARTASSIST WAS PLACED VIA RIGHT FEMORAL VENOUS ACCESS IN A 73-YEAR-OLD FEMALE WITH A MEDICAL HISTORY SIGNIFICANT FOR CORONARY ARTERY DISEASE, PRIOR MYOCARDIAL INFARCTION WITH STENTS, SEVERE MITRAL REGURGITATION STATUS POST MITRAL VALVE REPAIR, AND END-STAGE RENAL DISEASE ON HEMODIALYSIS. THE PATIENT PRESENTED IN CARDIOGENIC SHOCK SECONDARY TO ACUTE MYOCARDIAL INFARCTION AND WAS RECEIVING INOTROPES, VASOPRESSORS, AND MECHANICAL VENTILATORY SUPPORT. SHE WAS CLASSIFIED AS SCAI STAGE E AT THE TIME OF INTERVENTION. DURING PROCEDURAL PREPARATION, THE RIGHT INTERNAL JUGULAR VEIN WAS ACCESSED WITH THE INTENT TO PLACE THE INTRODUCER FOR IMPELLA RP FLEX SUPPORT. THE EXISTING HEMODIALYSIS CATHETER WAS REMOVED OVER A WIRE, AND SERIAL DILATION WAS PERFORMED USING AN 8 FR DILATOR FOLLOWED BY A 20 FR DILATOR. DURING ATTEMPTED ADVANCEMENT OF A 23 FR PEEL-AWAY INTRODUCER SHEATH, THE GUIDEWIRE WAS NOTED TO BE INSUFFICIENTLY POSITIONED WITHIN THE CENTRAL VENOUS SYSTEM. SUBSEQUENT ADVANCEMENT RESULTED IN LOSS OF WIRE POSITION, AND IMAGING REVEALED THE WIRE HAD ENTERED THE PLEURAL SPACE WITH INFERIOR TRACKING NEAR THE DIAPHRAGM. THE WIRE WAS REMOVED, AND TH
Patient problems: Great Vessel Perforation
MAUDE report 24266884 ↗InjuryJan 14, 2026Difficult to Insert; Difficult to Advance
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH AN IMPELLA RP FLEX EXPERIENCED LOSS OF FLOWS. THE PUMP WAS EXPLANTED AND THE PATIENT WAS IN STABLE CONDITION.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24067204 ↗
Other Temporary Non-Roller Type Left Heart Support Blood Pump devices
- Impella 0048-0045Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump5 recalls1,269 deaths11,112 reports
- Impella 1000100Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump5 recalls848 deaths6,093 reports
- Impella 1000080Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump7 recalls987 deaths5,084 reports
- Impella 0042-0000-USAbiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump12 recalls120 deaths2,456 reports
- Impella 0550-0008Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump4 recalls87 deaths1,005 reports
- Impella Impella CP Pump SetAbiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump9 recalls9 deaths272 reports
- Impella 0042-0010-USAbiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump7 recalls6 deaths284 reports
- Impella 0048-0024Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump2 recalls40 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.