ManoScan 3890Recalls & reported problems
Given Imaging, Inc. · ESO Z Catheter · Class 2
All models of ManoScan (8)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Given Imaging ManoScan catheter appears in 89 adverse event reports, with no deaths and 31 injuries. Reports mostly describe the catheter getting stuck during removal, sometimes in the nose, and parts becoming loose or deformed. Some patient problems listed include bleeding, pain, and airway obstruction. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 90
- Deaths reported
- 0
- Injuries reported
- 31
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 90 adverse event reports that list this exact device (DI 04260167482446), received between Jun 11, 2018 and Sep 18, 2026 — classified as 31 injuries, 59 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (78 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove78
- Material Integrity Problem14
- Material Deformation11
- Insufficient Device Problem Information9
- Adverse Event Without Identified Device or Use Problem5
- Poor Quality Image4
- Break2
- Failure to Calibrate2
Patient problems
- No Clinical Signs, Symptoms or Conditions31
- Hemorrhage/Blood Loss/Bleeding13
- Pain7
- Airway Obstruction6
- No Consequences Or Impact To Patient6
- Unspecified Tissue Injury6
- Blood Loss5
- Discomfort5
Recent adverse event reports
The 25 most recent of 90 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Difficult to Insert; Difficult to Remove; Loss of Data
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829201 ↗InjuryJun 20, 2025Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE IMPEDANCE CATHETER GOT STUCK IN A PATIENT¿S NARES. IT WAS INSERTED EASILY AND THE TEST COMPLETED, BUT UPON REMOVAL IT ONLY CAME OUT TO THE 41CM MARK (ABOUT 10 CM). IT COULD NOT BE REMOVED BY EITHER THE NURSES OR DOCTOR WITH THE PATIENT AWAKE AS SIGNIFICANT NASAL PAIN AND EMOTIONAL DISTRESS EXPERIENCED. PATIENT WAS THUS SEDATED IN THE ENDOSCOPY SUITE AND A GASTROSCOPY WAS DONE TO ENSURE THE CATHETER WAS NOT ¿STUCK¿ IN THE ESOPHAGUS OR STOMACH. EVERYTHING LOOKED NORMAL BELOW THE UPPER ESOPHAGEAL SPHINCTER, SO THE ISSUE OF RESISTANCE WAS IN THE PATIENT'S NASAL CAVITY. AFTER SEVERAL ATTEMPTS THE PHYSICIAN WAS ABLE TO REMOVE IT AND THE PATIENT WAS RECOVERED FROM SEDATION.
Patient problems: Distress; Pain
MAUDE report 22297286 ↗MalfunctionJun 18, 2025Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE PROBE BECAME STUCK IN THE PATIENT'S NASAL PASSAGE UPON REMOVAL. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22283316 ↗MalfunctionMar 4, 2025Difficult to Remove; Insufficient Device Problem Information; Difficult to Remove; Insufficient Device Problem Information
IT WAS REPORTED THAT THE CATHETER WHITE PLASTIC HAS DISLODGED AND, ALTHOUGH IT CALIBRATES AND WORKS CORRECTLY, WHEN THE CATHETER WAS REMOVED FROM THE PATIENT¿S NOSE, IT BECOMES STUCK. THE CATHETER WAS REMOVED CAREFULLY. THERE WAS NO ANY ADDITIONAL DEVICE NEEDED TO REMOVE THE CATHETER FROM THE PATIENT. THERE WAS NO PATIENT OR USER HARM. .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21516318 ↗MalfunctionJan 24, 2025Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT A CATHETER BECAME STUCK IN THE PATIENT'S NARES DURING ITS REMOVAL. THERE WAS NO PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21228819 ↗InjuryOct 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ESOPHAGOGASTRODUODENOSCOPY ASSIST. IT WAS CONFIRMED TO BE IN THE STOMACH. WHILE THE PATIENT WAS IN THE RECOVERY WAITING ROOM, THE "VOMITING" STARTED. IT WAS OBSERVED THAT THE CATHETER COILED UP IN THE BACK OF THE PATIENT¿S THROAT, FURTHER STIMULATING THE GAG REFLEX. THERE WAS NO DIFFICULTY PULLING THE CATHETER FROM THE PATIENT. NO RESISTANCE FELT. THE PATIENT WAS FINE, AND THE VOMITING STOPPED AFTER THE CATHETER WAS REMOVED. NO OTHER DEVICES WERE USED TO REMOVE THE CATHETER. IT WAS PULLED OUT IN THE MANNER OF REMOVING AN NG TUBE. AN X-RAY WAS ORDERED TO CONFIRM THAT THERE WAS NO INJURY.
Patient problems: Vomiting
MAUDE report 20510247 ↗InjurySep 25, 2024Difficult to Remove; Insufficient Device Problem Information; Difficult to Remove; Insufficient Device Problem Information
ACCORDING TO THE REPORTER, THE CATHETER GOT STUCK IN PATIENT'S NOSE. ENT (EAR, NOSE, THROAT) TEAM TOOK ONE OF ENT (EAR, NOSE, THROAT) SCOPES AND CUT THE DEVICE TO GET IT OUT OF PATIENT'S NOSE. FORCEPS WAS ALSO USED TO REMOVE THE CATHETER FROM THE PATIENT. THE CATHETER WAS DAMAGED.
Patient problems: Airway Obstruction
MAUDE report 20305248 ↗InjurySep 24, 2024Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, THE CATHETER GOT STUCK IN A PATIENT'S NOSE AND HAD TO BE REMOVED. THE CATHETER WAS CUT, PUSHED THROUGH THE PATIENT'S NOSE AND PULLED OUT OF MOUTH WITH FORCEPS. THE EAR, NOSE AND THROAT SPECIALIST TOOK A SCOPE TO LOOK IN PATIENT'S NOSE PRIOR TO PULLING DEVICE OUT AND AFTER REMOVAL. THE CATHETER WAS DAMAGED.
Patient problems: Airway Obstruction
MAUDE report 20294610 ↗MalfunctionJul 22, 2024Poor Quality Image; Difficult to Remove; Poor Quality Image; Difficult to Remove
ACCORDING TO THE REPORTER, THE CATHETER HAD GREY LINES ON STUDY. IT WAS STUCK ON PATIENT WHEN BEING REMOVED. PATIENT TURNED PROBE AND IT CAME OUT. PATIENT HAD A BLOODY NOSE.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 19792778 ↗MalfunctionMay 15, 2024Poor Quality Image; Difficult to Remove; Material Deformation; Poor Quality Image; Difficult to Remove; Material Deformation
ACCORDING TO THE REPORTER, THE CATHETER WAS DAMAGED DURING A DIFFICULT EXTUBATION. WHEN CALIBRATION WAS ATTEMPTED, IT SHOWED MULTIPLE PRESSURE BANDS. THE TPE SENSORS ALSO FELT TO BE STICKING OUT FURTHER THAN THE OTHER CATHETER. THERE WAS NO PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19325619 ↗MalfunctionApr 16, 2024Difficult to Remove; Insufficient Device Problem Information; Difficult to Remove; Insufficient Device Problem Information
ACCORDING TO THE REPORTER, THE CATHETER WAS DAMAGED. THE INNER WHITE LINING IS DEGRADED <(>&<)> STRUCTURE IS COMPROMISED. A PROBLEM WAS ENCOUNTERED AT THE END OF THE STUDY AT NASAL CANAL, THE CATHETER WAS DIFFICULT TO EXTUBATE, IT WAS STUCK. THERE WAS NO MEDICAL INTERVENTION PERFORMED TO REMOVE THE CATHETER. THERE WAS NO PATIENT AND USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19116484 ↗MalfunctionMar 29, 2024Difficult to Remove; Manufacturing, Packaging or Shipping Problem; Insufficient Device Problem Information; Difficult to Remove; Manufacturing, Packaging or Shipping Problem; Insufficient Device Problem Information
ACCORDING TO THE REPORTER, THE CATHETER WAS DIFFICULT TO INTUBATE AND EXTUBATE AS THE SENSORS APPEARED TO BE MISALIGNED, THIS WAS THE FIRST USE OF THE CATHETER. THERE WAS NO PATIENT AND USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19003750 ↗InjuryJan 31, 2024Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, THE CATHETER WAS UNABLE TO BE REMOVED FROM PATIENT'S NARE. IT WENT IN WITHOUT RESISTANCE. THEY ATTEMPTED MULTIPLE TIMES TO REMOVE IT WITH THE POOR PATIENT WRITHING IN PAIN. THE PHYSICIAN CAME TO ASSIST AND COULD NOT GET IT EITHER. THEY ENDED UP HAVING TO CUT THE CATHETER AND RETRIEVED IT ORALLY. PATIENT IS FINE.
Patient problems: Pain
MAUDE report 18616019 ↗InjuryJan 22, 2024Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, THE CATHETER WAS STUCK IN PATIENT'S NOSTRIL. RESISTANCE WAS FELT UPON REMOVAL OF PROBE. PATIENT COMPLAINED OF PAIN. PATIENT'S ENT PHYSICIAN WAS CONTACTED AND PATIENT WAS BROUGHT IMMEDIATELY TO ENT'S OFFICE. THE PROBE WAS REMOVED UNDER DIRECT VISUALIZATION OF NASAL SCOPE. CUSTOMER ADVISED THAT THERE WAS A BLOOD STAIN INSIDE THE CATHETER. IT WAS CAUSED BY A TEAR OF A VERY SMALL PIECE OF TISSUE THAT CAME OUT FROM PATIENT'S NOSTRIL UPON REMOVAL OF THE PROBE. NO BLEEDING WAS NOTED. PATIENT WAS STABLE.
Patient problems: Pain
MAUDE report 18552764 ↗MalfunctionDec 13, 2023Material Integrity Problem; Insufficient Device Problem Information; Material Integrity Problem; Insufficient Device Problem Information
ACCORDING TO THE REPORTER, THE CATHETER HAD A ROUGH SPOT ON THE CATHETER ABOVE THE METAL RING. THE CUSTOMER WAS AFRAID IT WILL INJURE A PATIENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18317528 ↗MalfunctionNov 6, 2023Difficult to Remove; Failure to Calibrate; Difficult to Remove; Failure to Calibrate
ACCORDING TO THE REPORTER, THE CATHETER GOT STUCK IN THE PATIENT BUT WAS ABLE TO GET IT OUT. THE CUSTOMER WANTED TO CHECK THE CATHETER BECAUSE IT FAILED TO CALIBRATE. THERE WAS NO PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18079075 ↗MalfunctionOct 3, 2023Difficult to Remove; Material Integrity Problem; Difficult to Remove; Material Integrity Problem
ACCORDING TO THE REPORTER, THE CATHETER GOT STUCK IN A PATIENT. ABOUT 1 INCH FROM END FELT LIKE THERE WAS A ROUGH SPOT. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Obstruction/Occlusion
MAUDE report 17866032 ↗InjuryOct 3, 2023Difficult to Remove; Material Integrity Problem; Difficult to Remove; Material Integrity Problem
ACCORDING TO THE REPORTER, THE CATHETER GOT STUCK IN A PATIENT. ABOUT 1 INCH FROM END FELT LIKE THERE WAS A ROUGH SPOT. SURGERY WAS DONE TO HAVE THE CATHETER REMOVED.
Patient problems: Obstruction/Occlusion
MAUDE report 17863051 ↗MalfunctionSep 14, 2023Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, THE CATHETER STUCK IN A PATIENT THAT HAD TO BE REMOVED BY THE EAR, NOSE AND THROAT CONSULTANT. THE PATIENT HAS A SMALL NASAL SEPTUM THE CATHETER WENT DOWN WITH NO PROBLEMS HOWEVER UPON REMOVAL THE CATHETER BECOME STRUCK IN THE TURBINATE¿S. THE PATIENT DID NOT SUFFER ANY OBSTRUCTIONS IN THE NOSE AND DID NOT SUFFER ANY BLEEDING DURING THE REMOVAL OF THE CATHETER. THE PATIENT RETURNED AND HAD A REPEAT PROCEDURE WITH NO PROBLEMS WENT DOWN THE OTHER NASAL PASSAGE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17745487 ↗MalfunctionAug 16, 2023Material Integrity Problem; Material Integrity Problem
ACCORDING TO THE REPORTER, THE TPE OF THE CATHETER FELT ROUGH ON CERTAIN SPOTS. THERE WAS NO PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17560057 ↗MalfunctionAug 3, 2023Difficult to Remove; Material Integrity Problem; Difficult to Remove; Material Integrity Problem
ACCORDING TO THE REPORTER, THE CATHETER HAD A ROUGH SPOT ON THE SENSOR PORTION AND IT WAS DIFFICULT TO INSERT AND REMOVE FROM THE PATIENT. THERE WAS NO PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17453672 ↗MalfunctionJul 3, 2023Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, THE CATHETER GOT STUCK IN THE PATIENT'S NOSE AND HAD TO BE REMOVED. IT WAS SHOWING A BAD SENSOR 8. THERE WAS NO PATIENT OR USER HARM.
Patient problems: Airway Obstruction; Nasal Obstruction
MAUDE report 17246846 ↗MalfunctionJun 21, 2023Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, THE PATIENT HAD A CATHETER STUCK IN NOSE. AFTER COMPLETION OF THE STUDY, THEY WERE TRYING TO REMOVE THE PROBE AND COULD NOT REMOVED AT FIRST. THE NURSE HELPED THE PATIENT RELAX IN ORDER FOR THE PROBE TO BE PROPERLY REMOVED AND CAUSED NOSEBLEED. IT DID NOT CAUSE TRAUMA ON BOTH PATIENT AND THE PROBE. THE PATIENT WAS STABLE AND DID NOT REQUIRE EXTRA MED INTERVENTION.
Patient problems: Airway Obstruction; Hemorrhage/Blood Loss/Bleeding
MAUDE report 17176201 ↗MalfunctionJun 12, 2023Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation
ACCORDING TO THE REPORTER, THE CATHETER WAS STUCK IN THE PATIENT NOSE. LUBRICANT WAS USED AND HAD TO TWIST THE PROBE AND CONTINUOUSLY LUBRICATE. CUSTOMER STATED IT WAS NOTICED THAT TPE WAS DAMAGED ONCE THE CATHETER WAS REMOVED FROM THE PATIENT NOSE. THERE WAS NO BLEEDING AND NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17113515 ↗MalfunctionJun 12, 2023Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, THE CATHETER WAS VERY CHALLENGING TO REMOVE. IT HAD TO BE FORCEFULLY REMOVED AND EACH RIDGE WAS HEARD COMING OUT OF THE PATIENT'S NOSE. THE PATIENT HAD DISCOMFORT WHEN THE CATHETER WAS REMOVED AND MINIMAL AMOUNT OF BLOOD WAS NOTED. THE PATIENT WAS DOING FINE AT DISCHARGE.
Patient problems: Discomfort; Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 17112044 ↗
FDA clearance and classification
The FDA regulates product code FFX (system, gastrointestinal motility (electrical)) under 21 CFR 876.1725 as a Class II device. 21 CFR 876.1725 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ManoScan 3890
Has the ManoScan 3890 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04260167482446), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ManoScan 3890?
The FDA's MAUDE database lists 90 adverse event reports for this device, received between Jun 11, 2018 and Sep 18, 2026. In those reports, the most common specific device problems are “Difficult to Remove” (78), “Material Integrity Problem” (14) and “Material Deformation” (11); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (13), “Pain” (7) and “Airway Obstruction” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ManoScan 3890?
None of the reports are classified as deaths, 31 as injuries and 59 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ManoScan 3890 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the ManoScan 3890, and is it still sold?
The FDA's device database lists Given Imaging, Inc. as the company responsible for it (brand ManoScan, model 3890). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL) devices
- EndoFLIP EF-100Covidien LP · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall3 reports
- Solar GI LGI H20, Trolley system G3-6Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall3 reports
- Solar GI HRM H20, 24p, Trolley system G3-8Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall2 reports
- Solar GI HRM H2O, 36p, Trolley system G3-12Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall1 report
- Zvu® ZVU-3Diversatek Healthcare, Inc. · System, Gastrointestinal Motility (Electrical)1 recall0 reports
- Solar GI HRAM H2O, 24p, Trolley system G3-14Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall0 reports
- Endoflip RFG-EF100Covidien LP · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall0 reports
- EndoFLIP Loaner Unit RFG-EF100Covidien LP · System, Gastrointestinal Motility (Electrical)1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page