Skip to content

Solar GI LGI H20, Trolley system G3-6Recalls & reported problems

Medical Measurement Systems B.V. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717775955047), received between Jan 5, 2026 and Mar 30, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Decrease in Pressure” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Class IIOpen, Classified

    Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.

    Z-2669-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Decrease in Pressure3

Patient problems

  • Convulsion/Seizure3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Decrease in Pressure

    CUSTOMER RETURNED GI PUMP FOR REPAIR AS IT WAS REPORTED THE DEVICE WAS FAULTY.

    Patient problems: Convulsion/Seizure

    MAUDE report 24720437 ↗
  • MalfunctionJan 6, 2026Decrease in Pressure

    THE PUMP FAILED, RESULTING IN NEGATIVE PRESSURE READINGS.

    Patient problems: Convulsion/Seizure

    MAUDE report 23975313 ↗
  • MalfunctionJan 5, 2026Decrease in Pressure

    THE PUMP WAS FAULTY RESULTING IN NEGATIVE PRESSURE READINGS.

    Patient problems: Convulsion/Seizure

    MAUDE report 23967719 ↗