Solar GI LGI H20, Trolley system G3-6Recalls & reported problems
Medical Measurement Systems B.V. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717775955047), received between Jan 5, 2026 and Mar 30, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Decrease in Pressure” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.
Z-2669-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Decrease in Pressure3
Patient problems
- Convulsion/Seizure3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Decrease in Pressure
CUSTOMER RETURNED GI PUMP FOR REPAIR AS IT WAS REPORTED THE DEVICE WAS FAULTY.
Patient problems: Convulsion/Seizure
MAUDE report 24720437 ↗MalfunctionJan 6, 2026Decrease in Pressure
THE PUMP FAILED, RESULTING IN NEGATIVE PRESSURE READINGS.
Patient problems: Convulsion/Seizure
MAUDE report 23975313 ↗MalfunctionJan 5, 2026Decrease in Pressure
THE PUMP WAS FAULTY RESULTING IN NEGATIVE PRESSURE READINGS.
Patient problems: Convulsion/Seizure
MAUDE report 23967719 ↗
Other SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL) devices
- ManoScan 3890Given Imaging, Inc. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)90 reports
- Solar GI HRM H20, 24p, Trolley system G3-8Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall2 reports
- Solar GI HRM H2O, 36p, Trolley system G3-12Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall1 report
- Zvu® ZVU-3Diversatek Healthcare, Inc. · System, Gastrointestinal Motility (Electrical)1 recall0 reports
- EndoFLIP EF-100Covidien LP · System, Gastrointestinal Motility (Electrical)1 recall0 reports
- Solar GI HRAM H2O, 24p, Trolley system G3-14Medical Measurement Systems B.V. · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall0 reports
- Zvu® HRAM-PROC-10Diversatek Healthcare, Inc. · System, Gastrointestinal Motility (Electrical)1 recall0 reports
- Endoflip RFG-EF100Covidien LP · SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page