Skip to content

EIT CIF Trial CET2870SRecalls & reported problems

EIT Emerging Implant Technologies GmbH · Trial, H 7mm, 8°, S · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04260397081617), received between Aug 18, 2021 and Feb 28, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K170503); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 28, 2023Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6), 2023, THE HANDLE FOR TRIAL BROKE AT THE NECK, LEAVING THE TRIAL HEAD IN THE PATIENT'S DISC SPACE. THE CASPAR PINS WERE INSERTED TO OPEN THE DISC SPACE AND THE TRIAL HEAD WAS RETRIEVED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH TEN(10) MINUTES OF SURGICAL DELAY. NO FRAGMENTS WERE GENERATED. THIS REPORT IS FOR ONE (1) TRIAL H 7MM 8DEG S. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16455655 ↗
  • MalfunctionAug 18, 2021Break; Break

    IT WAS FURTHER REPORTED THAT THERE WAS AN UNKNOWN SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12340209 ↗

FDA clearance and classification

Cleared through 510(k) K170503 on Jul 11, 2017 (decision: Substantially Equivalent). Submitted by Eit Emerging Implant Technologies GmbH as “EIT Cellular Titanium Cervical Cage, EIT Cellular Titanium PLIF Cages, EIT Cellular Titanium TLIF Cages, and EIT Cellular Titanium ALIF Cages”. The FDA review took 140 days. FDA record ↗

The FDA regulates product code LXH (orthopedic manual surgical instrument) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K170503. One submission often covers many models, so check the product and lot details in each record.

  • Z-1323-2022 · May 20, 2022 · Open, Classified
    EIT CIF cage, H 6mm, 8degree, L- an intervertebral body fusion devices indicated for use with autograft when used as an adjunct to fusion in patients with cervical disc disease (DD
  • Z-1324-2022 · May 20, 2022 · Open, Classified
    EIT CIF cage, H 7mm, 8degree, S- an intervertebral body fusion devices indicated for use with autograft when used as an adjunct to fusion in patients with cervical disc disease (DD

Common questions about the EIT CIF Trial CET2870S

Has the EIT CIF Trial CET2870S been recalled?

No FDA recall record in our data names this exact device identifier (DI 04260397081617), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the EIT CIF Trial CET2870S?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Aug 18, 2021 and Feb 28, 2023. In those reports, the most common specific device problem is “Break” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the EIT CIF Trial CET2870S?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the EIT CIF Trial CET2870S FDA approved?

It was cleared through 510(k) K170503 on Jul 11, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the EIT CIF Trial CET2870S, and is it still sold?

The FDA's device database lists EIT Emerging Implant Technologies GmbH as the company responsible for it (brand EIT CIF Trial, model CET2870S). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates