Skip to content

FDR Go 566-20000-18Recalls & reported problems

Shimadzu Corporation · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04540217059379), received between Mar 20, 2023 and Jul 1, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Unstable” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unstable3
  • Installation-Related Problem2
  • Arcing1
  • Insufficient Device Problem Information1
  • No Apparent Adverse Event1
  • Radiation Overexposure1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2025Arcing; Arcing

    WE RECEIVED THE FOLLOWING REPORT FROM FUJIFILM. ON APRIL 16TH, 2025, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PLUS E. THE ISSUE WAS THAT THE PLUG MAY NOT BE GROUNDED, SPARKING INSIDE OF CLEAR HOUSING AND CAUSING THE PLUG TO HEAT UP. THE INJURY TO THE USER OR ANY PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED AT THIS TIME. THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22368624 ↗
  • MalfunctionMay 16, 2025Sparking; Sparking

    ON APRIL 16TH, 2025, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PLUS E. THE ISSUE WAS THAT THE PLUG MAY NOT BE GROUNDED, SPARKING INSIDE OF CLEAR HOUSING AND CAUSING THE PLUG TO HEAT UP. THE INJURY TO THE USER OR ANY PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED AT THIS TIME. THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035997 ↗
  • MalfunctionMay 9, 2024Installation-Related Problem; No Apparent Adverse Event; Insufficient Device Problem Information; Installation-Related Problem; No Apparent Adverse Event; Insufficient Device Problem Information

    ON APRIL 12, WE WERE NOTIFIED BY THE IMPORTER OF THE DEVICE, FUJIFILM HEALTHCARE AMERICAS, THAT IT HAD SUBMITTED AN MDR (REPORT NO: 1000513161-2024 00025) TO THE FDA WITH THE FOLLOWING INFORMATION: ON MARCH 27TH 2024 FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PLUS E. IT WAS REPORTED THAT THE PEDS ABDOMEN IS DEFAULTING TO ADULT ABDOMEN TECHNIQUES. THERE IS NO ADDITIONAL INFORMATION PROVIDED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19276059 ↗
  • MalfunctionApr 11, 2024Installation-Related Problem; Installation-Related Problem

    ON (B)(6) 2024 FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PLUS E. IT WAS REPORTED THAT THE PEDS ABDOMEN IS DEFAULTING TO ADULT ABDOMEN TECHNIQUES. THERE IS NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19091213 ↗
  • MalfunctionMar 28, 2023Unstable; Unstable

    ON (B)(6) 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PLUS E. IT WAS REPORTED THAT THE SCREWS ON THE COLLIMATOR RING ARE BECOMING LOOSE OVER TIME CAUSING THE COLLIMATOR TO BECOME UNSTABLE. THERE IS NO PATIENT INVOLVEMENT OR REPORT OF SERIOUS INJURY OR DEATH ASSOCIATED WITH THE EVENT. THEREFORE, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16629272 ↗
  • MalfunctionMar 28, 2023Unstable; Unstable

    ON MARCH 13, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PLUS E. IT WAS REPORTED THAT THE SCREWS ON THE COLLIMATOR RING ARE BECOMING LOOSE OVER TIME CAUSING THE COLLIMATOR TO BECOME UNSTABLE. THERE IS NO PATIENT INVOLVEMENT OR REPORT OF SERIOUS INJURY OR DEATH ASSOCIATED WITH THE EVENT. THEREFORE, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16629270 ↗
  • MalfunctionMar 28, 2023Unstable; Unstable

    ON (B)(6) 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PLUS E. IT WAS REPORTED THAT THE SCREWS ON THE COLLIMATOR RING ARE BECOMING LOOSE OVER TIME CAUSING THE COLLIMATOR TO BECOME UNSTABLE. THERE IS NO PATIENT INVOLVEMENT OR REPORT OF SERIOUS INJURY OR DEATH ASSOCIATED WITH THE EVENT. THEREFORE, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16629261 ↗
  • MalfunctionMar 20, 2023Radiation Overexposure; Radiation Overexposure

    ON FEBRUARY 21, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PLUS E. IT WAS REPORTED THAT A 6-MONTH-OLD INFANT RECEIVED ONE ADDITIONAL IMAGING SCAN AT A HIGHER EXPOSURE RATE. THE TECHNICIAN PERFORMED THE INITIAL SCAN AT A RATE OF 80KV AND 14 MAS WHICH WAS THE RATE USED ON AN ADULT PATIENT PRIOR. THE TECHNICIAN NOTICED THAT THE KV AND MAS WERE INCORRECTLY SET FOR AN INFANT SCAN AND A SECOND EXPOSURE WAS RETAKEN AT THE CORRECTED RATE OF 63KV AND 1.6 MAS. IT IS UNCLEAR IF THE ADDITIONAL EXPOSURE COULD LEAD TO A SAFETY RISK. THEREFORE, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION. THERE IS NO SERIOUS INJURY OR DEATH ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16573944 ↗

Other System, X-Ray, Mobile devices

All System, X-Ray, Mobile devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page