Super Sheath (15-732B) BT70Z25PPW038Recalls & reported problems
Togo Medikit CO.,Ltd. · Class 2
All models of Super Sheath (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04543527181709), received between Apr 17, 2025 and Apr 17, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2025Break; Break
IT WAS REPORTED THAT A HOLE IN THE SHEATH OCCURRED. A 7 FR X 25 CM SUPER SHEATH INTRODUCER SHEATH WAS PLACED INTO THE ARM OVER THE WIRE FOR TREATING SLOW FLOW FISTULA PROCEDURE. THE WIRE WAS REMOVED AND TOOK AN ANGIOGRAPHY OF THE ARM BY INJECTING THE CONTRAST THROUGH THE SHEATH. HOWEVER, THE CONTRAST WAS FOUND COMING OUT OF THE TIP OF THE SHEATH. ANOTHER CONTRAST WAS SHOT, AND IT WAS CONFIRMED THAT THE CONTRAST WAS COMING OUT IN THE MIDDLE PART OF THE SHEATH. THE DEVICE WAS REMOVED AND FOUND A HOLE IN IT. ANOTHER SHEATH OF THE SAME SIZE AND TYPE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21854136 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page