The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04543527181792), received between Jan 2, 2019 and Dec 5, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Contaminated During Manufacture or Shipping” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contaminated During Manufacture or Shipping2
  • Break1
  • Device-Device Incompatibility1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 5, 2023Difficult to Remove; Device-Device Incompatibility; Difficult to Remove; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SHEATH WAS DIFFICULT TO REMOVE. A 5FR INTRODUCER SHEATH WAS SELECTED FOR USE IN A TRANS-ARTERIAL CHEMOEMBOLIZATION (TACE) PROCEDURE. VASCULAR ACCESS WAS OBTAINED VIA A PUNCTURE OF THE RIGHT COMMON FEMORAL ARTERY (CFA). THE INTRODUCER SHEATH WAS ADVANCED INTO THE ARTERY OVER A STANDARD 0.035 GUIDEWIRE. A NON-BSC DIAGNOSTIC CATHETER WAS ADVANCED OVER THE GUIDEWIRE AND TORQUED 360 DEGREES. AT THIS POINT IT WAS OBSERVED THAT THE CATHETER COULD NOT BE RETRACTED, AND A GUIDEWIRE COULD NOT PASS THROUGH IT. A 0.018 NON-BSC GUIDEWIRE WAS ATTEMPTED FOR USE, BUT WAS UNSUCCESSFUL. THE CATHETER WAS MANAGED TO BE REMOVED WITH DIFFICULTY, AND A NEW .035 NON-BSC GUIDEWIRE WAS ADVANCED INTO THE AORTA. IT WAS THEN OBSERVED THAT THE CATHETER WAS COMPLETELY TWISTED ON ITSELF. THE SHEATH WAS THEN REMOVED, HOWEVER, IT WAS CRUMPLED AND DIFFICULT TO REMOVE. A NEW 5FR INTRODUCER SHEATH WAS SELECTED FOR USE. THERE WAS DIFFICULTY ADVANCING THE SHEATH INTO THE VESSEL. THE SHEATH WAS OBSERVED TO HAVE COILED UP UNDER THE SKIN WHICH RESULTED IN A HEMATOMA. A NEW 5FR SHEATH WAS SELECTED FOR USE. THE PROCEDURE WAS THEN ABLE TO BE COMPLETED WITHOUT FURTHER INCIDENT. THERE WERE NO FURTHER PA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18268060 ↗
  • MalfunctionApr 29, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT THERE WAS FLUID WITHIN THE STERILE PACKAGE. FLUID WAS OBSERVED IN THE SUPER SHEATH INTRODUCER SHEATH STERILE PACKAGE NEAR THE END OF THE SHEATH. THE FLUID WAS VISIBLE WITHOUT OPENING THE PACKAGE. THE PACKAGING APPEARED NORMAL AND WAS INTACT. THERE WERE NUMEROUS SUPER SHEATH INTRODUCER SHEATH DEVICES OBSERVED TO HAVE THIS ISSUE. ONE OF THE PACKAGES WAS OPENED AND THE SUBSTANCE WAS NOTED TO BE OILY. THE DEVICES ARE ALL IN THEIR PACKAGES AND WERE NOT USED WITH A PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8562235 ↗
  • MalfunctionApr 29, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT THERE WAS FLUID WITHIN THE STERILE PACKAGE. FLUID WAS OBSERVED IN THE SUPER SHEATH INTRODUCER SHEATH STERILE PACKAGE NEAR THE END OF THE SHEATH. THE FLUID WAS VISIBLE WITHOUT OPENING THE PACKAGE. THE PACKAGING APPEARED NORMAL AND WAS INTACT. THERE WERE NUMEROUS SUPER SHEATH INTRODUCER SHEATH DEVICES OBSERVED TO HAVE THIS ISSUE. ONE OF THE PACKAGES WAS OPENED AND THE SUBSTANCE WAS NOTED TO BE OILY. THE DEVICES ARE ALL IN THEIR PACKAGES AND WERE NOT USED WITH A PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 8562176 ↗
  • MalfunctionJan 2, 2019Break; Break

    IT WAS REPORTED THAT THE DISTAL END OF THE SHEATH WAS DETACHED. THE TARGET LESION WAS LOCATED AT SUPERFACIAL FEMORAL ARTERY. A 5FR X 11CM SUPER SHEATH WAS SELECTED FOR USE. DURING THE PROCEDURE, THE PHYSICIAN INSERTED A NON BSC CLOSURE DEVICE. AFTER WHICH, A NON BSC WIRE WAS INSERTED AND THE SHEATH WAS REMOVED OVER THE WIRE. HOWEVER, IT WAS NOTED THAT THE DISTAL END OF THE SHEATH HAD SEPARATED. THE LEG WAS CHECKED AND IMAGED AND IT WAS NOTED THAT THE DISTAL SHEATH TIP MARK ON THE SHEATH HAD BROKEN OFF IN THE SFA AND TRAVELLED DOWN TO THE POPLITEAL ARTERY. THE PHYSICIAN THEN USED A MACH 1 GUIDE CATHETER TO ASPIRATE THE DISTAL TIP OF THE SHEATH OUT IN THE BODY AND IT WAS THEN LODGED IN THE SOFT TISSUE TRACK. FINALLY, THE PHYSICIAN RETRIEVED THE DISTAL END OF THE SHEATH OUT OF THE SOFT TISSUE TRACK OF THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO WIRE LEFT IN THE SHEATH. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.

    Patient problems: No Code Available

    MAUDE report 8208813 ↗