ACM 0306563000Recalls & reported problems
Stryker Corporation · Bowl and Cement Cartridge with Breakaway Femoral Nozzle · Class 1
All models of ACM (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Stryker ACM bone cement mixer, used with a breakaway femoral nozzle during hip surgery, is described in 44 adverse event reports, with 3 injuries and no deaths. Reports mostly describe the applicator nozzle coming off, cracking, or failing to lock onto the cement cartridge, plus units arriving contaminated or with debris in the sterile packaging. One open recall from August 2023 covers the possibility that the nozzle tip breaks off during use and pieces fall into the surgical site.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 44
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 04546540055408), received between Jun 13, 2018 and Aug 12, 2025 — classified as 3 injuries, 41 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (25 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 7, 2023Class IIOpen, Classified
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
Z-2598-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component25
- Delivered as Unsterile Product15
- Device Contaminated During Manufacture or Shipping3
- Biocompatibility1
- Crack1
- Pressure Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions38
- No Patient Involvement5
- Insufficient Information4
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A BHA, WHEN A SURGEON WAS INJECTING BONE CEMENT, THE NOZZLE CAME OFF FROM THE SYRINGE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22783963 ↗MalfunctionApr 21, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY CONTAMINATED DURING MANUFACTURE OR SHIPPING. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21878617 ↗MalfunctionMar 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE NOZZLE OF DEVICE WAS NOT ABLE TO LOCK WITH CARTRIDGE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21733769 ↗MalfunctionJan 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE STOPPER TAB OF THE NOZZLE DID NOT GET OVER THE BUMP OF SYRINGE DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196237 ↗MalfunctionOct 22, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY CONTAMINATED DURING MANUFACTURE OR SHIPPING. 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20504488 ↗MalfunctionJul 24, 2024Delivered as Unsterile Product; Device Contaminated During Manufacture or Shipping; Delivered as Unsterile Product; Device Contaminated During Manufacture or Shipping
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19823142 ↗MalfunctionApr 15, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19113323 ↗MalfunctionJan 22, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18556595 ↗MalfunctionDec 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AFTER MIXING CEMENT USING ACM SYSTEM, CEMENT WAS FILLED INTO THE FEMUR BY THE 2 SPEED CEMENT GUN. DURING FILLING, THE NOZZLE WAS BROKEN AT THE CONNECTION PART BETWEEN THE ACM SYSTEM AND THE CEMENT GUN. THERE WAS NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18366467 ↗MalfunctionOct 23, 2023Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17987879 ↗MalfunctionSep 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
CEMENT NOZZLE TAB WOULD NOT ENGAGE ON CARTRIDGE. SCRUB NURSE USED MALLET TO PUSH TAB OVER. NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17740848 ↗MalfunctionAug 31, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE CEMENT MIXING BOWL LOT 23010012 COULDN'T TIGHTEN UP PROPERLY WITH THE NOZZLE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17660541 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17515515 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17515513 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17515364 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17515029 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17515027 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17515026 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17512445 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17508277 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE NOZZLE TAB DID NOT REACH THE LOCK POSITION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17508276 ↗MalfunctionAug 10, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE NOZZLE DISENGAGED DURING CEMENTING. NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17497331 ↗MalfunctionAug 9, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE CEMENT CARTRIDGE TAB DID NOT LINE UP WITH TRIANGLE FOR FIRM CONNECTION WHEN PRESSURIZING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17487529 ↗MalfunctionAug 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE ACM NOZZLE NOT CONNECTING TO CARTRIDGE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17478897 ↗MalfunctionAug 7, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE ACM NOZZLE NOT CORRECTLY ATTACHING TO CEMENT CARTRIDGE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17474326 ↗
FDA clearance and classification
The FDA regulates product code JDZ (mixer, cement, for clinical use) under 21 CFR 888.4210 as a Class I device. 21 CFR 888.4210 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ACM 0306563000
Has the ACM 0306563000 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2598-2023 (Class II), began on Aug 7, 2023: Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ACM 0306563000?
The FDA's MAUDE database lists 44 adverse event reports for this device, received between Jun 13, 2018 and Aug 12, 2025. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (25), “Delivered as Unsterile Product” (15) and “Device Contaminated During Manufacture or Shipping” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ACM 0306563000?
None of the reports are classified as deaths, 3 as injuries and 41 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ACM 0306563000 FDA approved?
Devices of this type (product code JDZ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker Corporation.
Who makes the ACM 0306563000, and is it still sold?
The FDA's device database lists Stryker Corporation as the company responsible for it (brand ACM, model 0306563000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other MIXER, CEMENT, FOR CLINICAL USE devices
- SMARTSET 5450-35-500Depuy (Ireland) · MIXER, CEMENT, FOR CLINICAL USE2 deaths964 reports
- NA 5049-11Zimmer, Inc. · MIXER, CEMENT, FOR CLINICAL USE1 recall388 reports
- ACM 0306573000Stryker Corporation · MIXER, CEMENT, FOR CLINICAL USE2 recalls9 reports
- NA 0206530000Stryker Corporation · MIXER, CEMENT, FOR CLINICAL USE2 recalls2 reports
- NA 0206512000Stryker Corporation · MIXER, CEMENT, FOR CLINICAL USE2 recalls2 reports
- AutoPlex, VertaPlex 0607687000Stryker Corporation · MIXER, CEMENT, FOR CLINICAL USE1 recall4 reports
- Revolution 0606573000Stryker Corporation · MIXER, CEMENT, FOR CLINICAL USE1 recall2 reports
- Revolution 0606516000Stryker Corporation · MIXER, CEMENT, FOR CLINICAL USE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page