Skip to content

ACM 0306573000Recalls & reported problems

Stryker Corporation · Bowl and Cement Cartridge with Breakaway Femoral Nozzle and Medium Pressurizer · Class 1

All models of ACM (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04546540055415), received between Jul 27, 2020 and Oct 30, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (6 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 7, 2023Class IIOpen, Classified

    Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site

    Z-2600-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 9, 2017Class IITerminated

    Lack of sterility assurance as bioburden levels were found to be higher than internal acceptable rates.

    Z-1248-2018 · Root cause: Environmental control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component6
  • Delivered as Unsterile Product1
  • Device Contaminated During Manufacture or Shipping1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A BHA, WHEN A SURGEON WAS INJECTING BONE CEMENT, THE NOZZLE CAME OFF FROM THE SYRINGE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23424455 ↗
  • MalfunctionJan 22, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY CONTAMINATED DURING MANUFACTURE OR SHIPPING. - 1 EVENT HAD PATIENT INVOLVEMENT, NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21207624 ↗
  • MalfunctionJul 26, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE CEMENT MIXER WAS EXPIRED, IT WAS USED IN SURGERY WITH PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19837099 ↗
  • MalfunctionMar 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE DEVICE WAS LEAKING BETWEEN THE CARTRIDGE AND THE NOZZLE DURING A PROCEDURE. THE CASE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18855981 ↗
  • MalfunctionJul 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 10 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 8 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17397204 ↗
  • InjuryJun 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE NOZZLE OF THE CEMENT MIXING BOWL WAS NOT ABLE TO BE CONNECTED PROPERLY CAUSING A DETACHMENT AND FINALLY A LEAKAGE OF CEMENT OUT THE GUN AND THEREFORE IT HAD TO BE APPLIED BY HAND. THE PROCEDURE WAS COMPLETED BY PLACING THE CEMENT BY HAND; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES HAVE BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17164620 ↗
  • InjuryJun 2, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE REPORTED EVENT OCCURRED DURING A RIGHT HEMI HIP PROCEDURE WITH A CEMENTED STEM. THE CEMENT WAS MIXED AND THE ACM DELIVERY SYSTEM WAS USED TO BEGIN INJECTION OF THE CEMENT INTO THE MEDULLARY CANAL. DURING CEMENT INJECTION, THE NOZZLE OF THE ACM CAME OFF. THE SURGEON RE-ATTACHED THE NOZZLE OF THE ACM, WHICH TOOK 1-2 MINUTES. THE SURGEON WAS ABLE TO INJECT THE REMAINING VIABLE CEMENT IN THE ACM AND CONTINUED WITH THE PROCEDURE. IT WAS REPORTED THERE WERE CHALLENGES WHEN PLACING THE SIZE 5 STEM, AS IT WAS SITTING 2CM PROUD. THE SURGICAL PROCEDURE REQUIRED MODIFICATION AND BECAME MORE COMPLEX WHEN THE SURGEON REMOVED THE IMPLANT AND USED A CURETTE TO RETRIEVE CEMENT FROM THE CANAL. ALL CEMENT THAT HAD BEEN INJECTED COULD NOT BE REMOVED. A SIZE 2 IMPLANT WAS PLACED, WHICH SAT 1CM PROUD. THE SIZE 2 STEM WAS THEN REMOVED. THESE STEPS OF THE SURGICAL PROCEDURE OCCURRED OVER APPROXIMATELY 60 MINUTES, AT WHICH POINT THE PROCEDURE WAS CONCLUDED WITHOUT A STEM BEING IMPLANTED. IT WAS REPORTED THE PATIENT HAD DEVELOPED AN UNSPECIFIED CARDIAC ISSUE AND LATER EXPIRED. THIS REPORT IS FOR THE MALFUNCTION OF THE ACM NOZZLE DETACHING DURING USE, AND THE ASSOCIATED MEDICAL INTERVENTION. MFR REPORT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17050312 ↗
  • MalfunctionApr 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16802916 ↗
  • MalfunctionJul 27, 2020Delivered as Unsterile Product; Delivered as Unsterile Product

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10329745 ↗

FDA clearance and classification

The FDA regulates product code JDZ (mixer, cement, for clinical use) under 21 CFR 888.4210 as a Class I device. 21 CFR 888.4210 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ACM 0306573000

Has the ACM 0306573000 been recalled?

Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-2600-2023 (Class II), began on Aug 7, 2023: Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ACM 0306573000?

The FDA's MAUDE database lists 9 adverse event reports for this device, received between Jul 27, 2020 and Oct 30, 2025. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (6), “Delivered as Unsterile Product” (1) and “Device Contaminated During Manufacture or Shipping” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ACM 0306573000?

None of the reports are classified as deaths, 2 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ACM 0306573000 FDA approved?

Devices of this type (product code JDZ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker Corporation.

Who makes the ACM 0306573000, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand ACM, model 0306573000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other MIXER, CEMENT, FOR CLINICAL USE devices

All MIXER, CEMENT, FOR CLINICAL USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page