ASNIS 604660SRecalls & reported problems
Stryker GmbH · Cannulated Screw · Class 2 · Implant
All models of ASNIS (51)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540089908), received between Oct 29, 2025 and Oct 29, 2025 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Fracture1
Patient problems
- Medical device site infection4
- Foreign Body In Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2025Fracture; Fracture
THE CUSTOMER REPORTED THAT: "PATIENT OPERATED ON (B)(6) 2024 FOR A BIMALLEOLAR FRACTURE WITH INSERTION OF TWO STRYKER SCREWS REF. 604660S BATCH K31308 AND ONE PLATE REF. 40-20903S BATCH 1000591836. SEVERAL ORTHOPEDIC CONSULTATIONS, INCLUDING ON (B)(60 2024 "MATERIAL VISIBLE TO THE NAKED EYE AND INFECTION ON MATERIAL" PROPOSAL TO REMOVE THE MATERIAL; LAST CONSULTATION ON (B)(6) 2025 WITH SCHEDULING OF OUTPATIENT REMOVAL (INFECTION WAS CONTROLLED BETWEEN THE PREVIOUS CONSULTATION AND THE LAST ONE WITH TRACEABILITY OF COMPLEX DRESSINGS). (PLEASE REFER TO COMPLAINTS (B)(4) ON (B)(6) 2025, REMOVAL IN DAY HOSPITAL IN ACCORDANCE WITH DECISION; (PLEASE REFER TO COMPLAINTS (B)(4). DURING EXPLANTATION OF THE MATERIAL INCOMPLETE, A SCREW BROKE IN THE BONE OF THE MALLEOLUS DURING ITS EXTRACTION. A PIECE OF THE SCREW COULD NOT BE REMOVED AND REMAINS IN THE TISSUE.
Patient problems: Foreign Body In Patient
MAUDE report 23418095 ↗InjuryOct 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED THAT : "PATIENT OPERATED ON (B)(6) 2024 FOR A BIMALLEOLAR FRACTURE WITH INSERTION OF TWO STRYKER SCREWS REF. 604660S BATCH K31308 AND ONE PLATE REF. 40-20903S BATCH 1000591836. SEVERAL ORTHOPEDIC CONSULTATIONS, INCLUDING ON (B)(6) 2024 "MATERIAL VISIBLE TO THE NAKED EYE AND INFECTION ON MATERIAL" PROPOSAL TO REMOVE THE MATERIAL; LAST CONSULTATION ON (B)(6) 2025 WITH SCHEDULING OF OUTPATIENT REMOVAL (INFECTION WAS CONTROLLED BETWEEN THE PREVIOUS CONSULTATION AND THE LAST ONE WITH TRACEABILITY OF COMPLEX DRESSINGS). (PLEASE REFER TO COMPLAINTS (B)(4). ON (B)(6) 2025, REMOVAL IN DAY HOSPITAL IN ACCORDANCE WITH DECISION. DURING EXPLANTATION OF THE MATERIAL INCOMPLETE, A SCREW BROKE IN THE BONE OF THE MALLEOLUS DURING ITS EXTRACTION. (PLEASE REFER TO COMPLAINT (B)(4).
Patient problems: Medical device site infection
MAUDE report 23417937 ↗InjuryOct 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED THAT: "PATIENT OPERATED ON (B)(6) 2024 FOR A BIMALLEOLAR FRACTURE WITH INSERTION OF TWO STRYKER SCREWS REF. 604660S BATCH K31308 AND ONE PLATE REF. (B)(4) BATCH 1000591836. SEVERAL ORTHOPEDIC CONSULTATIONS, INCLUDING ON (B)(6) 2024 "MATERIAL VISIBLE TO THE NAKED EYE AND INFECTION ON MATERIAL" PROPOSAL TO REMOVE THE MATERIAL; LAST CONSULTATION ON (B)(6) 2025 WITH SCHEDULING OF OUTPATIENT REMOVAL (INFECTION WAS CONTROLLED BETWEEN THE PREVIOUS CONSULTATION AND THE LAST ONE WITH TRACEABILITY OF COMPLEX DRESSINGS). (PLEASE REFER TO COMPLAINTS (B)(4)). ON (B)(6) 2025, REMOVAL IN DAY HOSPITAL IN ACCORDANCE WITH DECISION. DURING EXPLANTATION OF THE MATERIAL INCOMPLETE, A SCREW BROKE IN THE BONE OF THE MALLEOLUS DURING ITS EXTRACTION. (PLEASE REFER TO PI (B)(4)).
Patient problems: Medical device site infection
MAUDE report 23417932 ↗InjuryOct 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED THAT: "PATIENT OPERATED ON (B)(6) 2024 FOR A BIMALLEOLAR FRACTURE WITH INSERTION OF TWO STRYKER SCREWS REF. 604660S, BATCH K31308 AND ONE PLATE REF. 40-20903S BATCH 1000591836. SEVERAL ORTHOPEDIC CONSULTATIONS, INCLUDING ON (B)(6) 2024 "MATERIAL VISIBLE TO THE NAKED EYE AND INFECTION ON MATERIAL" PROPOSAL TO REMOVE THE MATERIAL; LAST CONSULTATION ON (B)(6) 2025 WITH SCHEDULING OF OUTPATIENT REMOVAL (INFECTION WAS CONTROLLED BETWEEN THE PREVIOUS CONSULTATION AND THE LAST ONE WITH TRACEABILITY OF COMPLEX DRESSINGS). ON 10/02/2025, REMOVAL IN DAY HOSPITAL IN ACCORDANCE WITH DECISION; (PLEASE REFER TO COMPLAINTS (B)(4) DURING EXPLANTATION OF THE MATERIAL INCOMPLETE, A SCREW BROKE IN THE BONE OF THE MALLEOLUS DURING ITS EXTRACTION. (PLEASE REFER TO (B)(4).
Patient problems: Medical device site infection
MAUDE report 23417574 ↗InjuryOct 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED THAT: "PATIENT OPERATED ON (B)(6) 2024 FOR A BIMALLEOLAR FRACTURE WITH INSERTION OF TWO STRYKER SCREWS REF. 604660S, BATCH: K31308 AND ONE PLATE REF. 40-20903S, BATCH: 1000591836. SEVERAL ORTHOPEDIC CONSULTATIONS, INCLUDING ON (B)(6) 2024 "MATERIAL VISIBLE TO THE NAKED EYE AND INFECTION ON MATERIAL" PROPOSAL TO REMOVE THE MATERIAL; LAST CONSULTATION ON (B)(6) 2025 WITH SCHEDULING OF OUTPATIENT REMOVAL (INFECTION WAS CONTROLLED BETWEEN THE PREVIOUS CONSULTATION AND THE LAST ONE WITH TRACEABILITY OF COMPLEX DRESSINGS). ON (B)(6) 2025, REMOVAL IN DAY HOSPITAL IN ACCORDANCE WITH DECISION; (PLEASE REFER TO COMPLAINTS (B)(4). DURING EXPLANTATION OF THE MATERIAL INCOMPLETE, A SCREW BROKE IN THE BONE OF THE MALLEOLUS DURING ITS EXTRACTION. (PLEASE REFER TO (B)(4)).
Patient problems: Medical device site infection
MAUDE report 23417537 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page