Skip to content

BIXCUT 0225-8115Recalls & reported problems

Stryker GmbH · IM Reamer, AO Fitting · Class 1

All models of BIXCUT (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540145413), received between Nov 15, 2018 and Nov 15, 2018 — classified as 1 death.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Death1
  • Pulmonary Embolism1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathNov 15, 2018Adverse Event Without Identified Device or Use Problem

    THE SURGEON WAS UNDERTAKING A T2 NAILING PROCEDURE IN A (B)(6) FEMALE PATIENT . AT THE START OF THE CASE DURING REAMING WITH BIXCUT REAMERS, THE PATIENT HAD A PULMONARY EMBOLISM. THE SURGEONS ATTEMPTED RESUSCITATION AND TRANSFERRED THE PATIENT TO ITU BUT THE PATIENT PASSED AWAY SOME TIME LATER. THERE WERE NO REPORTS OF ANY ISSUES WITH THE BIXCUT REAMER DURING USE AND THE T2 NAIL HAD NOT BEEN IMPLANTED UPDATE: THE PATIENT WAS A (B)(6) FEMALE, OVERWEIGHT (HIGH BMI). SHE ATTEMPTED A HALF MARATHON ON (B)(6) 2018 BUT DROPPED OUT. SHE COMPLAINED TO A&E OF THIGH PAIN, A STRESS FRACTURE WAS LATER NOTED AFTER 3 VISITS TO HOSPITAL AND WAS ADMITTED FOR A IM NAIL PROCEDURE. THE CASE WAS PROGRESSING AS PLANNED WITHOUT INCIDENT, HOWEVER THERE WAS A SPIRAL FRACTURE WHICH THE SURGEONS FACE DIFFICULTY IN REDUCING. FOLLOWING REDUCTION THE SURGEON STARTED REAMING USING THE BIXCUT REAMERS. THE ANAESTHETIST REPORTED A PE AND THE NEED TO INTUBATE AND THE SURGEON STOPPED THE PROCEDURE. FOLLOWING THIS THEY STARTED TO STABILISE THE PATIENT AND PERFORM CPR. AFTER 15 MINS THE ANAESTHETIST CONFIRMED THAT SHE WOULD NOT BE FIT FOR CONTINUATION WITH THE NAIL EVEN IF SHE WAS STABILIZED. ALL OF THESE DEVICES HAS B

    Patient problems: Death; Pulmonary Embolism

    MAUDE report 8075100 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page