Skip to content

BIXCUT 0227-6100Recalls & reported problems

Stryker GmbH · IM Reamer, mod. Trinkle Fitting · Class 1

All models of BIXCUT (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540145840), received between Jul 27, 2022 and Jun 23, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation2
  • Break1
  • Fracture1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2025Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    AS REPORTED: "THE REAMER WAS USED TO WIDEN THE FEMORAL MARROW CAVITY. IT WAS DEFEATED BY THE HARD BONE AND THE SHAFT PORTION WAS BENT. THE REAMER WAS NOT BROKEN, SO NO METAL OR OTHER OBJECT WAS IMPLANTED IN THE BODY.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22310689 ↗
  • MalfunctionDec 20, 2024Material Deformation; Material Deformation

    AS REPORTED: "MECHANICAL DAMAGE OF REAMERS DURING REAMING OF TIBIAL BONE. EVERY NEXT REAMER WERE BLOCKING IN MARROW CAVITY. THEY WERE USED ACCORDING TO INSTRUCTION FOR USE, STARTING FROM THE SMALLEST. IN THE END REPLACEMENT DEVICE FROM DIFFERENT COMPANY HAS BEEN USED. NOTHING LEFT IN PATIENT BODY." (B)(6) 2024 - ADDITIONAL INFORMATION RECEIVED: DEFORMATION OF THE DEVICE, DEVICE WAS CHANGED DUE TO POSSIBILITY OF BREAKING. ALL INSTRUMENTS USED TO DRILL WERE COMPATIBLE WITH BIXCUT REAMERS. THERE WERE USED STRYKER¿S POWER TOOLS WITH ALL APPROPRIATE ATTACHMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20995779 ↗
  • MalfunctionJul 27, 2022Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT IM REAMERS BROKE DURING USE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15113984 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page