Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04546540153326), received between Sep 10, 2018 and Feb 25, 2025 — classified as 10 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Break3
- Insufficient Device Problem Information2
- Naturally Worn2
- Adverse Event Without Identified Device or Use Problem1
- Use of Device Problem1
Patient problems
- Ambulation or Postural Difficulties4
- Injury4
- Insufficient Information4
- Fall2
- Bone Fracture(s)1
- No Consequences Or Impact To Patient1
- Reaction1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryFeb 25, 2025Fracture; Fracture
EXETER V40 44 #3 STEM SNAPPED AT SHOULDER LEADING TO IMPLANT FAILURE AND REVISION SURGERY. REVISION SURGERY TOOK PLACE ON (B)(6) 2025 WHERE BROKEN EXETER STEM WAS REMOVED AND REPLACED WITH THE SAME SIZE STEM.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 21462143 ↗InjuryNov 15, 2022Fracture; Fracture
THE FOLLOWING WAS REPORTED: HIP REVISION SURGERY DUE TO PATIENT'S EXETER STEM 44 N3 BEING BROKEN AT NECK LEVEL.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 15798863 ↗InjuryOct 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT REQUIRED A REVISION SURGERY FOR PERIPROSTHETIC FRACTURE OF HIP. PATIENT ADMITTED INTO HOSPITAL AFTER A TRIP AND FALL AT HOME.
Patient problems: Bone Fracture(s); Fall
MAUDE report 15672153 ↗InjurySep 23, 2021Fracture; Fracture
THE SURGEON REPORTED "AN EVENT TOOK PLACE IN 2019: PATIENT 130KILOS, IMPLANT BROKE AT THE EXETER'S NECK (WITHOUT ANY NOTION OF PARTICULAR TRAUMA)".
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 12518682 ↗InjuryMar 29, 2021Break; Break
(B)(6)-YR OLD MALE HAD A FALL BACKWARDS. X-RAY SHOWED HIS RIGHT FEMORAL STEM WAS IN TWO PIECES. REVISION SURGERY WAS REQUIRED. 44MM#3 EXETER STEM WAS REMOVED AND REPLACED WITH ANOTHER INTO EXISTING CEMENT MANTLE. OPERATING SURGEON COMMENTED THAT THE 36MM +5 FEMORAL HEAD APPEARED TO BE ¿TWISTED AND STUCK¿ INSIDE THE ACETABULAR COMPONENT AND HE COMMENTED THAT HE BELIEVES THE STEM BROKE AT THE NECK AS A RESULT OF THE MECHANISM OF THE FALL.
Patient problems: Fall
MAUDE report 11581244 ↗InjuryMar 2, 2021Break; Fracture; Break; Fracture
REVISION HIP SURGERY CASE. EXETER STEM REMOVED. IMPLANT HAD BROKEN AT THE NECK.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 11398902 ↗InjuryJun 11, 2020Naturally Worn; Naturally Worn
REVISION OF EXETER STEM. FAILED DUE TO SUSPECTED POLY WEAR (NON STRYKER). UPDATE: NO ALLEGATIONS WERE NOTED AGAINST THE EXETER STEM.
Patient problems: Injury
MAUDE report 10142880 ↗InjuryJun 12, 2019Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information
REVISION OF EXETER TOTAL HIP REPLACEMENT. UPDATE 05/JUNE/2019 WG: AS REPORTED IN HOW WAS ISSUE NOTICED "PATIENT PRESENTED TO SURGEON WITH PAIN IN RIGHT HIP. RTHR PREVIOUS 5 YEARS". ADDITIONAL DETAILS PROVIDED BY REP: "DURING THE REVISION OPERATION THE REASON WAS EVIDENT WITH METALOSIS. FROM MY OBSERVATION THE FEMORAL HEAD WAS SITTING LOOSELY ON THE FEMORAL COMPONENT".
Patient problems: Injury
MAUDE report 8691300 ↗InjuryNov 9, 2018Break; Fracture; Break; Fracture
MalfunctionNov 7, 2018Use of Device Problem; Use of Device Problem
SURGEON WAS PLACING DEFINITIVE EXETER IMPLANT ONTO THE STEM INTRODUCER. I NOTICED HE HAD PLACED THE WINGED CENTRALISER ONTO THE STEM FOR A SMALL CANAL. WE DISCUSSED HIS OPTION TO USE THE STRAIGHT CENTRALISER TO USE WITH ALONGSIDE THE SMALL ARTISAN PLUG. HE ACKNOWLEDGED MY SUGGESTION BUT CHOSE TO GO AHEAD WITH THE WINGED CENTRALISER. HE THEN ATTACHED THE IMPLANT ONTO THE STEM INTRODUCER AND WRAPPED A PACK OVER THE END OF THE STEM AND PLACED IT ON THE TABLE STILL WRAPPED IN THE PACK. HE THEN PICKED UP THE STEM, WRAPPED IN THE PACK SO THAT THE TIP WAS NOT VISIBLE AND PROCEEDED TO INSERT THE STEM INTO THE PATIENT/CEMENT. IT WAS THEN NOTICED BY THE STRYKER REPRESENTATIVE THAT THE CENTRALISER WAS ON THE TABLE BEHIND SOME RETRACTORS. THE REP BROUGHT THIS TO THE IMMEDIATE ATTENTION OF THE SURGEON TO WHICH HE REPLIED, ¿OH WELL, TOO LATE NOW¿.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8047657 ↗InjurySep 10, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
NOTIFICATION OF LFIT HAZARD ALERT GIVEN TO OPERATING SURGEON AND HE CONTACTED PATIENT TO INFORM. SURGEON WAS ADVISED THAT THE PATIENT REQUIRED AN EARLY REVISION AS HAD A TISSUE REACTION. SURGEON REPORTS THAT THE PATIENT HAS HAD SUBSEQUENT PROBLEMS.
Patient problems: Injury; Reaction
MAUDE report 7862251 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish devices
- Arcos Modular Revision Hip System 11-301310Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish1 recall9 reports
- Arcos Modular Revision Hip System 11-301342Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish1 recall4 reports
- Tritanium 502-03-52DHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish39 reports
- V40 Head 6260-5-332Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish12 reports
- V40 Head 6260-9-032Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish11 reports
- Trident 509-02-56EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish10 reports
- Taperloc Complete Primary Femoral 51-104180Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish9 reports
- Taperloc Complete Primary Femoral 51-103200Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page