Howmedica Osteonics Corp. · LFIT V40 Femoral Head · Class 2 · Implant

Adverse event reports
11
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07613327013009), received between Oct 31, 2017 and Jan 6, 2025 — classified as 9 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information3
  • Corroded2
  • Adverse Event Without Identified Device or Use Problem1
  • Degraded1
  • Detachment of Device or Device Component1
  • Device Contamination with Chemical or Other Material1
  • Device Dislodged or Dislocated1
  • Fitting Problem1

Patient problems

  • Injury6
  • Inflammation2
  • Joint Dislocation2
  • No Consequences Or Impact To Patient2
  • Pain2
  • Reaction2
  • Ambulation or Postural Difficulties1
  • Foreign Body Reaction1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY COUNSEL THAT CLAIMANT UNDERWENT RIGHT THA ON (B)(6) 2012 AND WAS IMPLANTED WITH AN LFIT V40 FEMORAL HEAD AND ACCOLADE TMZF HIP STEM. HER RIGHT HIP WAS REVISED ON (B)(6) 2022 ALLEGEDLY DUE TO ELEVATED METAL IONS AND MRI FINDINGS CONSISTENT WITH ADVERSE LOCAL TISSUE REACTION.

    Patient problems: Foreign Body Reaction

    MAUDE report 21081848 ↗
  • InjuryApr 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PATIENT REQUIRED A TOTAL HIP REVISION SURGERY DUE TO FAILURE OF THE HEAD-TRUNNION JUNCTION. THE PATIENT HAD PRESENTED TO THE ER WITH DISLOCATION OF THE FEMORAL HEAD WHICH WAS CONFIRMED UPON ADMITTANCE. IT WAS FURTHER REPORTED THAT THE REVISION INVOLVED REMOVAL OF THE TRIDENT ACETABULAR LINER, V40 FEMORAL HEAD AND ACCOLADE TMZF STEM. THE ORIGINAL SHELL WAS LEFT IN PLACE.

    Patient problems: Ambulation or Postural Difficulties; Joint Dislocation

    MAUDE report 16713028 ↗
  • InjuryJun 10, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    REP WAS NOTIFIED BY SURGEON THAT A PATIENT HAS HIP PAIN AND WANTED TO KNOW IF THE PATIENT'S FEMORAL HEAD WAS SUBJECT TO RECALL. REP WAS NOTIFIED HEAD IS NOT SUBJECT OF RECALL, SURGEON WAS NOTIFIED BY THE REP. IT IS UNKNOWN AT THE TIME OF REPORT IF THERE IS ANY TREATMENT PLAN FOR THIS PATIENT. REP CONFIRMED THAT NO FURTHER INFORMATION IS AVAILABLE FROM THE HOSPITAL OR SURGEON AT THE TIME OF REPORT, BUT WILL UPDATE IF ANY NEW INFORMATION BECOMES AVAILABLE.

    Patient problems: Pain

    MAUDE report 10137709 ↗
  • InjuryApr 2, 2020Corroded; Corroded

    PATIENTS HIP WAS REVISED FOR SUSPECTED INFECTION AND ACETABULAR LOOSENING. INFECTION LABS WERE FOUND TO BE NEGATIVE, BUT THERE WAS MARKED CORROSION ON THE FEMORAL STEM AND HEAD. PATIENTS ACETABULUM WAS REPLACED WITH A TRIDENT II MULTIHOLE SHELL AND 40MM LINER. FEMORAL HEAD WAS REPLACED WITH +0 40MM BIOLOX UNIVERSAL HEAD AND SLEEVE.

    Patient problems: Inadequate Osseointegration; Injury

    MAUDE report 9918105 ↗
  • InjuryOct 8, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO RECURRENT DISLOCATIONS OF A STRYKER FEMORAL HEAD AND A COMPETITOR LINER. THE LFIT V40 FEMORAL HEAD AND COMPETITOR DEVICES WERE REVISED TO STRYKER DEVICES. REP REPORTED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Injury; Joint Dislocation

    MAUDE report 9165812 ↗
  • InjuryOct 4, 2019Degraded; Insufficient Device Problem Information; Degraded; Insufficient Device Problem Information

    IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HER LEFT HIP ON OR ABOUT (B)(6) 2006 AND WAS REVISED ON (B)(6) 2018. IT IS FURTHER ALLEGED THAT SHE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE AT ISSUE, DEVICE RECALL AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HER BLOOD.

    Patient problems: Inflammation; Injury; Metal Related Pathology; Reaction

    MAUDE report 9155850 ↗
  • InjuryFeb 25, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    DURING AN INFECTION CASE IT WAS NOTICED METALLOSIS AROUND THE HEAD AND ACCORDING TO SURGEON ALVAL CASE. THE HEAD WAS REVISED. THE PATIENT HAD AN ABG2 WITH LFIT V40 HEAD. THE PATIENT GOT A UNIVERSAL BIOLOX DELTA HEAD WITH V40 SLEEVE.

    Patient problems: Inflammation; Injury; Reaction; Unspecified Infection

    MAUDE report 8367087 ↗
  • InjuryAug 28, 2018Corroded; Corroded

    IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED DUE TO SUSPECTED CORROSION (ION LEVELS NOT REPORTED TO REP). REP WAS NOT PRESENT FOR THE PROCEDURE.

    Patient problems: Injury

    MAUDE report 7823434 ↗
  • InjuryDec 6, 2017Material Discolored; Material Discolored

    PATIENT HAD A TOTAL HIP IN (B)(6) 2011 AND RECENTLY COMPLAINED OF PAIN. A BONE SCAN SHOWED THAT THE CUP WAS LOOSE AND IT WAS REVISED. THERE WAS EVIDENCE OF BLACK STAINING IN THE HEAD AND AROUND THE NECK OF THE IMPLANT.

    Patient problems: Injury; Pain

    MAUDE report 7089489 ↗
  • MalfunctionNov 15, 2017Device Contamination with Chemical or Other Material; Packaging Problem; Device Contamination with Chemical or Other Material; Packaging Problem

    AN IMPLANT WAS OPENED AND HAIR WAS FOUND. THE NURSE TOOK IT AWAY SO IT NEVER CAME IN CONTACT WITH PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7032586 ↗
  • MalfunctionOct 31, 2017Fitting Problem; Fitting Problem

    SURGEON REPORTED THAT WHEN HE PLACED DEFINITIVE HEAD ON TO EXETER STEM, THE HEAD FIT LOOSELY. AFTER IMPACTING WITH HEAD IMPACTOR AND MALLET, THE HEAD WAS ABLE TO BE REMOVED BY HAND. ANOTHER HEAD WAS OPENED AND USED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6991535 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.