V40 Head 6260-9-032Recalls & reported problems
Howmedica Osteonics Corp. · LFIT V40 Femoral Head · Class 2 · Implant
All models of V40 Head (30)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07613327013009), received between Oct 31, 2017 and Jan 6, 2025 — classified as 9 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information3
- Corroded2
- Adverse Event Without Identified Device or Use Problem1
- Degraded1
- Detachment of Device or Device Component1
- Device Contamination with Chemical or Other Material1
- Device Dislodged or Dislocated1
- Fitting Problem1
Patient problems
- Injury6
- Inflammation2
- Joint Dislocation2
- No Consequences Or Impact To Patient2
- Pain2
- Reaction2
- Ambulation or Postural Difficulties1
- Foreign Body Reaction1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY COUNSEL THAT CLAIMANT UNDERWENT RIGHT THA ON (B)(6) 2012 AND WAS IMPLANTED WITH AN LFIT V40 FEMORAL HEAD AND ACCOLADE TMZF HIP STEM. HER RIGHT HIP WAS REVISED ON (B)(6) 2022 ALLEGEDLY DUE TO ELEVATED METAL IONS AND MRI FINDINGS CONSISTENT WITH ADVERSE LOCAL TISSUE REACTION.
Patient problems: Foreign Body Reaction
MAUDE report 21081848 ↗InjuryApr 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PATIENT REQUIRED A TOTAL HIP REVISION SURGERY DUE TO FAILURE OF THE HEAD-TRUNNION JUNCTION. THE PATIENT HAD PRESENTED TO THE ER WITH DISLOCATION OF THE FEMORAL HEAD WHICH WAS CONFIRMED UPON ADMITTANCE. IT WAS FURTHER REPORTED THAT THE REVISION INVOLVED REMOVAL OF THE TRIDENT ACETABULAR LINER, V40 FEMORAL HEAD AND ACCOLADE TMZF STEM. THE ORIGINAL SHELL WAS LEFT IN PLACE.
Patient problems: Ambulation or Postural Difficulties; Joint Dislocation
MAUDE report 16713028 ↗InjuryJun 10, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
REP WAS NOTIFIED BY SURGEON THAT A PATIENT HAS HIP PAIN AND WANTED TO KNOW IF THE PATIENT'S FEMORAL HEAD WAS SUBJECT TO RECALL. REP WAS NOTIFIED HEAD IS NOT SUBJECT OF RECALL, SURGEON WAS NOTIFIED BY THE REP. IT IS UNKNOWN AT THE TIME OF REPORT IF THERE IS ANY TREATMENT PLAN FOR THIS PATIENT. REP CONFIRMED THAT NO FURTHER INFORMATION IS AVAILABLE FROM THE HOSPITAL OR SURGEON AT THE TIME OF REPORT, BUT WILL UPDATE IF ANY NEW INFORMATION BECOMES AVAILABLE.
Patient problems: Pain
MAUDE report 10137709 ↗InjuryApr 2, 2020Corroded; Corroded
PATIENTS HIP WAS REVISED FOR SUSPECTED INFECTION AND ACETABULAR LOOSENING. INFECTION LABS WERE FOUND TO BE NEGATIVE, BUT THERE WAS MARKED CORROSION ON THE FEMORAL STEM AND HEAD. PATIENTS ACETABULUM WAS REPLACED WITH A TRIDENT II MULTIHOLE SHELL AND 40MM LINER. FEMORAL HEAD WAS REPLACED WITH +0 40MM BIOLOX UNIVERSAL HEAD AND SLEEVE.
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 9918105 ↗InjuryOct 8, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO RECURRENT DISLOCATIONS OF A STRYKER FEMORAL HEAD AND A COMPETITOR LINER. THE LFIT V40 FEMORAL HEAD AND COMPETITOR DEVICES WERE REVISED TO STRYKER DEVICES. REP REPORTED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Injury; Joint Dislocation
MAUDE report 9165812 ↗InjuryOct 4, 2019Degraded; Insufficient Device Problem Information; Degraded; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HER LEFT HIP ON OR ABOUT (B)(6) 2006 AND WAS REVISED ON (B)(6) 2018. IT IS FURTHER ALLEGED THAT SHE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE AT ISSUE, DEVICE RECALL AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HER BLOOD.
Patient problems: Inflammation; Injury; Metal Related Pathology; Reaction
MAUDE report 9155850 ↗InjuryFeb 25, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
DURING AN INFECTION CASE IT WAS NOTICED METALLOSIS AROUND THE HEAD AND ACCORDING TO SURGEON ALVAL CASE. THE HEAD WAS REVISED. THE PATIENT HAD AN ABG2 WITH LFIT V40 HEAD. THE PATIENT GOT A UNIVERSAL BIOLOX DELTA HEAD WITH V40 SLEEVE.
Patient problems: Inflammation; Injury; Reaction; Unspecified Infection
MAUDE report 8367087 ↗InjuryAug 28, 2018Corroded; Corroded
IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED DUE TO SUSPECTED CORROSION (ION LEVELS NOT REPORTED TO REP). REP WAS NOT PRESENT FOR THE PROCEDURE.
Patient problems: Injury
MAUDE report 7823434 ↗InjuryDec 6, 2017Material Discolored; Material Discolored
PATIENT HAD A TOTAL HIP IN (B)(6) 2011 AND RECENTLY COMPLAINED OF PAIN. A BONE SCAN SHOWED THAT THE CUP WAS LOOSE AND IT WAS REVISED. THERE WAS EVIDENCE OF BLACK STAINING IN THE HEAD AND AROUND THE NECK OF THE IMPLANT.
Patient problems: Injury; Pain
MAUDE report 7089489 ↗MalfunctionNov 15, 2017Device Contamination with Chemical or Other Material; Packaging Problem; Device Contamination with Chemical or Other Material; Packaging Problem
AN IMPLANT WAS OPENED AND HAIR WAS FOUND. THE NURSE TOOK IT AWAY SO IT NEVER CAME IN CONTACT WITH PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7032586 ↗MalfunctionOct 31, 2017Fitting Problem; Fitting Problem
SURGEON REPORTED THAT WHEN HE PLACED DEFINITIVE HEAD ON TO EXETER STEM, THE HEAD FIT LOOSELY. AFTER IMPACTING WITH HEAD IMPACTOR AND MALLET, THE HEAD WAS ABLE TO BE REMOVED BY HAND. ANOTHER HEAD WAS OPENED AND USED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6991535 ↗
FDA clearance and classification
The FDA regulates product code MAY (prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish) under 21 CFR 888.3353 as a Class II device. 21 CFR 888.3353 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the V40 Head 6260-9-032
Has the V40 Head 6260-9-032 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613327013009), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the V40 Head 6260-9-032?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Oct 31, 2017 and Jan 6, 2025. In those reports, the most common specific device problems are “Corroded” (2), “Degraded” (1) and “Detachment of Device or Device Component” (1); the most common patient problems are “Injury” (6), “Inflammation” (2) and “Joint Dislocation” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the V40 Head 6260-9-032?
None of the reports are classified as deaths, 9 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the V40 Head 6260-9-032 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the V40 Head 6260-9-032, and is it still sold?
The FDA's device database lists Howmedica Osteonics Corp. as the company responsible for it (brand V40 Head, model 6260-9-032). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish devices
- Arcos Modular Revision Hip System 11-301310Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish1 recall8 reports
- Arcos Modular Revision Hip System 11-301342Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish1 recall4 reports
- Tritanium 502-03-52DHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish39 reports
- V40 Head 6260-5-332Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish12 reports
- EXETER 0580-1-443Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish11 reports
- Trident 509-02-56EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish10 reports
- Taperloc Complete Primary Femoral 51-104180Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish9 reports
- Taperloc Complete Primary Femoral 51-103200Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page