The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540195913), received between Sep 29, 2015 and May 12, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component1
  • Device Contamination with Chemical or Other Material1
  • Difficult to Advance1
  • Difficult to Remove1
  • Fracture1
  • Improper or Incorrect Procedure or Method1
  • Physical Resistance/Sticking1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 12, 2025Break; Fracture; Break; Fracture

    AS REPORTED:"THE SURGEON ASKED FOR A SLIGHT BEND IN THE GUIDE, WHICH ENDED BROKEN IN THE IBODE'S FINGERS." ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22003510 ↗
  • MalfunctionMar 8, 2019Difficult to Remove; Difficult to Advance; Physical Resistance/Sticking

    DURING PRIMARY PROCEDURE OF LEFT HUMERUS. IT WAS REPORTED THAT WHEN REAMING, THE REAMERS (SIZE 6-8 IN .5MM INCREMENTS) WERE DIFFICULT TO ADVANCE OVER THE 3.0 800MM GUIDE WIRE. SURGEON REAMED TO SIZE 8 AND MOVED TO MODULAR REAMERS TO MORE EASILY REAM. AFTER THE NAIL WAS PLACED WITH THE TARGETER, A SCREW HOLE WAS DRILLED USING THE CANNULA. AFTER DRILLING, HE INSERTED THE FIRST SCREW AND THE CANNULA BECAME STUCK ON THE SCREW HEAD AND WAS DIFFICULT TO REMOVE. THEN THE SECOND PROXIMAL SCREW HOLE WAS DRILLED. AS THE SURGEON ADVANCED THE SCREW DOWN THE CANNULA, HE THOUGHT THE SCREW HEAD WAS GOING TO BECOME STUCK AGAIN. SURGEON REMOVED THE CANNULA TO ATTEMPT TO IMPLANT THE SCREW, COULD NOT FIND THE DRILLED HOLE, AND HAD TO MAKE AN INCISION TO REMOVE THE SCREW. SURGERY WAS COMPLETED WITH AN OVERALL DELAY OF APPROXIMATELY 20 MINUTES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8405949 ↗
  • MalfunctionJul 12, 2018Break; Detachment of Device or Device Component

    CUSTOMER REPORTS THAT 1 CM GUIDEWIRE INCL HEAD BROKE OFF IN THE STEM OF THE HUMERUS (DURING UPREAMING). A PART REMAINED IN THE HUMERUS.

    Patient problems: Foreign Body In Patient

    MAUDE report 7682840 ↗
  • MalfunctionOct 27, 2016Device Contamination with Chemical or Other Material

    IN IN-COMING INSPECTION AT DISTRIBUTION, IT WAS FOUND THAT THERE WAS A FOREIGN MATERIAL (HAIR) IN BLISTER. THIS EVENT WAS FOUND ON 1 OF 20 UNITS, LOT#K09888B RECEIVED ON OUR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6062287 ↗
  • MalfunctionSep 29, 2015Shelf Life Exceeded; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT DURING A TOTAL HIP ARTHROPLASTY, AN OUT OF DATE GUIDEWIRE WAS MISTAKENLY USED.