T2 1806-0163Recalls & reported problems

Stryker GmbH · Nail Holding Screw, Humerus 10 mm · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540196095), received between Oct 26, 2018 and Feb 9, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Connection Problem1
  • Fracture1
  • Material Fragmentation1

Patient problems

  • Cardiac Arrest1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2026Break; Fracture

    AS REPORTED: "BREAKING OF THE LOCKING PINS ON A DIAPHYSEAL HUMERAL NAIL." ADDITIONAL FEEDBACK RECEIVED: "HOW WAS THE EVENT DETECTED? DURING SURGERY, WHEN IT BECAME IMPOSSIBLE TO CONTINUE THE PROCEDURE. WHEN INSERTING THE HUMERAL NAIL, THE NAIL WAS PLACED ON THE NAIL HOLDER, THE SCREW WAS TIGHTENED CORRECTLY, AND THE NAIL ANCHOR WAS CHECKED. THE ANVIL WAS POSITIONED TO IMPACT THE NAIL. A ROTATIONAL MOVEMENT WAS PERFORMED TO ALIGN THE NAIL AND POSITION IT CORRECTLY. DURING THE ROTATION, A CRACK WAS HEARD, WITH ABNORMAL MOBILITY BETWEEN THE NAIL HOLDER AND THE NAIL. DID THE EVENT OCCUR DURING THE PROCEDURE? IF YES: WAS THE PROCEDURE COMPLETED CORRECTLY? YES, DURING THE PROCEDURE, IT WAS COMPLETED CORRECTLY. WAS THERE ANY IMPACT ON THE PATIENT?/ANY CLINICAL CONSEQUENCES FOR THE PATIENT? NO IMPACT ON THE PATIENT. WAS THE PROCEDURE DELAYED? IF YES: BY HOW MUCH TIME (IN MINUTES)? THE PROCEDURE WAS DELAYED BY 25-30 MINUTES WHILE WE FETCHED AN ANCILLARY DEVICE TO EXTRACT THE NAIL AND REPOSITION ANOTHER ONE. WAS ADDITIONAL MEDICAL INTERVENTION NECESSARY? NO."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24300025 ↗
  • MalfunctionDec 18, 2019Material Fragmentation; Material Fragmentation

    AS REPORTED: "STAFF WAS TRYING TO AFFIX THE T2 HUMERAL NAIL TO TARGETER WITH NAIL HOLDING BOLT. BOLT WOULD NOT THREAD INTO NAIL. UPON INSPECTION WE NOTICED THREADS ON THE NAIL HOLDING BOLT ARE BROKEN.". ADDITIONALLY REPORTED: EVENT OCCURRED INTRA-OPERATIVELY. SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9488556 ↗
  • InjuryOct 26, 2018Break; Connection Problem

    IT WAS REPORTED THAT THE NAIL WAS NOT CONNECTED TO TARGET DEVICE WITH T HANDLE. AFTER THAT, THE TARGET DEVICE (1806-0143) AND HOLDING SCREW (1806-0163) WAS LOCKED AND BOTH DEVICES WERE NOT ABLE TO BE REMOVED. THEREFORE, THE TARGET DEVICE WAS NOT USED AND SAME NAIL WAS IMPLANTED AND THE FOUR SCREWS WERE INSERTED WITH FREE HAND TECHNIC. WHEN THE DISTAL SCREW WAS INSERTED, THE BLOOD VESSEL DAMAGED AT THE DISTAL PART OF THE NAIL AND THE VASCULAR SUTURED. CARDIC ARREST OCCURRED DUE TO VASCULAR INJURY, BUT PERFORMED CARDIOPULMONARY RESUSCITATION, PATIENT'S RESUSCITATED. PATIENT DID NOT HAVE RESIDUAL EFFECT. THE DOCTOR REQUESTED THAT CAUSE TARGET DEVICE AND HOLDING SCREW LOCKED AND TROUBLESHOOTING.

    Patient problems: Cardiac Arrest

    MAUDE report 8010619 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →