NA 1806-0263Recalls & reported problems
Stryker GmbH · Screwdriver Shaft, Compression · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04546540196330), received between Jan 24, 2019 and Jan 7, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Fracture4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 7, 2026Fracture
AS REPORTED: "WHILE USING THAT DRIVER TO APPLY COMPRESSION, IT BROKE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23989501 ↗MalfunctionFeb 7, 2025Break; Fracture; Break; Fracture
AS REPORTED: DRIVER BROKEN WHILE PERFORMING COMPRESSION WITH COMPRESSION SCREWDRIVER. A SECOND SCREWDRIVER WAS USED AND PROCEDURE WAS COMPLETED SUCCESSFULLY".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21339904 ↗MalfunctionApr 13, 2023Break; Fracture; Break; Fracture
AS REPORTED: "THE DRIVER BROKE OFF AT THE BASE WHILE TURNING THE COMPRESSION SCREW WITH A COMPRESSION DRIVER TO APPLY COMPRESSION TO THE FRACTURE. AT THE POINT, COMPRESSION WAS SUFFICIENT, SO THE PROCEDURE WAS FINISHED AS IT WAS."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16730805 ↗MalfunctionSep 23, 2020Fracture; Fracture
THE PRODUCT SPECIALIST REPORTED THAT THE IT BROKE DURING USE. IT BROKE DURING FINAL TIGHTENING OF THE COMPRESSION SCREW BUT HE WAS ABLE TO FINISH. IT PROLONGED THE OPERATION WITH 15-20 MIN BUT HE IS HAPPY WITH THE RESULTS. NO IMPACT ON PATIENT. IT BROKE UP AT THE ATTACHMENT TO THE HANDLE. HE BENT IT MORE AND COULD USE A CHUCK TO HOLD IT TO BE ABLE TO TIGHTEN THE LAST BIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10576678 ↗MalfunctionJun 12, 2020Break; Break
AS REPORTED, DURING THE OPERATION, WHEN TRYING TO APPLY COMPRESSION, THE WELDED PART OF THE DRIVER WAS DAMAGED. THE SR EXPLAINED TO THE SURGEON ABOUT THE POSSIBILITY OF DAMAGE BEFORE APPLYING COMPRESSION. THE SURGEON USED THE SCREWDRIVER FOR COMPRESSION SCREW AFTER CONSENT. AFTER ALL, THE WELDED PART WAS EASILY DAMAGED. SINCE THERE ARE TWO DRIVERS INSTALLED IN THE INSTRUMENT, ANOTHER DRIVER WAS USED. THE WELDED PART WAS ALSO EASILY DAMAGED. THE SURGEON WAS STATED THAT HE WANTED MORE COMPRESSION, BUT BECAUSE THERE WAS NO LONGER A DRIVER TIP, HE HAD TO BEND THE DAMAGED DRIVER 90 DEGREES WITH PLIERS AND USED IT TO PUT MORE PRESSURE. PRESSURE WAS APPLIED UNTIL THE SURGEON WAS SATISFIED, AND THE OPERATION ENDED SAFELY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10151086 ↗MalfunctionJun 12, 2020Break; Break
AS REPORTED, DURING THE OPERATION, WHEN TRYING TO APPLY COMPRESSION, THE WELDED PART OF THE DRIVER WAS DAMAGED. THE SR EXPLAINED TO THE SURGEON ABOUT THE POSSIBILITY OF DAMAGE BEFORE APPLYING COMPRESSION. THE SURGEON USED THE SCREWDRIVER FOR COMPRESSION SCREW AFTER CONSENT. AFTER ALL, THE WELDED PART WAS EASILY DAMAGED. SINCE THERE ARE TWO DRIVERS INSTALLED IN THE INSTRUMENT, ANOTHER DRIVER WAS USED. THE WELDED PART WAS ALSO EASILY DAMAGED. THE SURGEON WAS STATED THAT HE WANTED MORE COMPRESSION, BUT BECAUSE THERE WAS NO LONGER A DRIVER TIP, HE HAD TO BEND THE DAMAGED DRIVER 90 DEGREES WITH PLIERS AND USED IT TO PUT MORE PRESSURE. PRESSURE WAS APPLIED UNTIL THE SURGEON WAS SATISFIED, AND THE OPERATION ENDED SAFELY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10150706 ↗MalfunctionJan 24, 2019Break
IT IS REPORTED THAT THE SCREWDRIVER BROKEE OFF AT COUPLING PIECE DURING OPERATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8276287 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.