NA 1806-0325Recalls & reported problems
Stryker GmbH · Screw Gauge, Long · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540196408), received between Sep 13, 2021 and Jun 1, 2023 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Incorrect Measurement5
- Break1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 1, 2023Incorrect Measurement; Incorrect Measurement
AS REPORTED: "DURING SURGERY, THE SURGEON NOTICED A DISCREPANCY BETWEEN THE LENGTH MEASUREMENT USING A SCALED DRILL AND THE LENGTH GAUGE (DIFFERENCE APPROX. 4MM). AFTER CHECKING IN THE STERILIZATION DEPARTMENT AND COMPARISON WITH TWO OTHER REFERENCES, THIS ERROR APPEARS TO BE CONFIRMED AND A COMPARATIVELY LONGER LENGTH WAS DETERMINED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17038477 ↗MalfunctionJun 1, 2023Incorrect Measurement; Incorrect Measurement
AS REPORTED: "DURING SURGERY, THE SURGEON NOTICED A DISCREPANCY BETWEEN THE LENGTH MEASUREMENT USING A SCALED DRILL AND THE LENGTH GAUGE (DIFFERENCE APPROX. 4MM). AFTER CHECKING IN THE STERILIZATION DEPARTMENT AND COMPARISON WITH TWO OTHER REFERENCES, THIS ERROR APPEARS TO BE CONFIRMED AND A COMPARATIVELY LONGER LENGTH WAS DETERMINED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17038275 ↗MalfunctionMay 31, 2023Incorrect Measurement; Incorrect Measurement
AS REPORTED: "DURING SURGERY, THE SURGEON NOTICED A DISCREPANCY BETWEEN THE LENGTH MEASUREMENT USING A SCALED DRILL AND THE LENGTH GAUGE (DIFFERENCE APPROX. 4MM). AFTER CHECKING IN THE STERILIZATION DEPARTMENT AND COMPARISON WITH TWO OTHER REFERENCES, THIS ERROR APPEARS TO BE CONFIRMED AND A COMPARATIVELY LONGER LENGTH WAS DETERMINED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17034777 ↗MalfunctionMay 30, 2023Incorrect Measurement; Incorrect Measurement
AS REPORTED: "DURING SURGERY, THE SURGEON NOTICED A DISCREPANCY BETWEEN THE LENGTH MEASUREMENT USING A SCALED DRILL AND THE LENGTH GAUGE (DIFFERENCE APPROX. 4MM). AFTER CHECKING IN THE STERILIZATION DEPARTMENT AND COMPARISON WITH TWO OTHER REFERENCES, THIS ERROR APPEARS TO BE CONFIRMED AND A COMPARATIVELY LONGER LENGTH WAS DETERMINED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17025965 ↗MalfunctionMay 30, 2023Incorrect Measurement; Incorrect Measurement
AS REPORTED: "DURING SURGERY, THE SURGEON NOTICED A DISCREPANCY BETWEEN THE LENGTH MEASUREMENT USING A SCALED DRILL AND THE LENGTH GAUGE (DIFFERENCE APPROX. 4MM). AFTER CHECKING IN THE STERILIZATION DEPARTMENT AND COMPARISON WITH TWO OTHER REFERENCES, THIS ERROR APPEARS TO BE CONFIRMED AND A COMPARATIVELY LONGER LENGTH WAS DETERMINED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17025752 ↗MalfunctionSep 13, 2021Break; Fracture; Break; Fracture
IT WAS REPORTED THE DEPTH GAUGE WAS BROKEN WHILE MEASURING WITH THE T2 DEPTH GAUGE LONG TO INSERT THE DISTAL LOCKING SCREW. IT WAS AS IF THE THIN TIP HAD COME LOOSE FROM THE THICKER PART IN FRONT OF IT. THE BROKEN PART WAS NOT RETRIEVED FROM THE PATIENT'S BODY.
Patient problems: Foreign Body In Patient
MAUDE report 12459276 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page