Quick summary

This Stryker T2 nail handle is a manual orthopedic tool used to hold and position tibia nails during surgery. The 25 reports that name its identifier mostly describe the screw and adapter jamming or sticking together, making it hard to attach or detach the nail and lengthening surgery. The reports mention no deaths and 2 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 04546540196538), received between Sep 25, 2015 and Feb 21, 2026 — classified as 2 injuries, 23 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam8
  • Device Difficult to Setup or Prepare5
  • Difficult or Delayed Separation4
  • Difficult to Insert2
  • Separation Failure2
  • Break1
  • Device Handling Problem1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • No Known Impact Or Consequence To Patient5
  • Insufficient Information3
  • Discomfort1
  • Myocardial Infarction1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 21, 2026Mechanical Jam

    IT WAS REPORTED THAT A NAIL HOLDING SCREW JAMMED INTO THE NAIL HOLDING ADAPTER DURING A T2 TIBIA NAIL INSERTION. THIS WAS DURING A CASE WHEN THE NAIL WOULD NOT DE-THREAD FROM THE JIG, THIS RESULTED IN THE NAIL SNAPPING PROXIMALLY INSIDE THE PATIENT, THIS WAS REMOVED AND REPLACED WITH ANOTHER NAIL USING ANOTHER SET. NO REPORTED ADVERSE EFFECTS OTHER THAN A SIGNIFICANT TIME DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24410254 ↗
  • MalfunctionFeb 6, 2026Mechanical Jam

    IT WAS REPORTED THAT "BEFORE SURGERY THE STAFF WANTED TO TRY OUT THE NAIL. THE CONNECTED IT WITH TARGETING DEVICE AND IT WAS NOT POSSIBLE TO DISCONNECT THEM. DIFFERENT SYSTEM WAS USED FOR SURGERY".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24284026 ↗
  • MalfunctionMay 21, 2025Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    AS REPORTED: "THE INSTRUMENTS, ONCE THE PATIENT HAD ALREADY BEEN ANESTHETIZED, WERE FOUND TO BE DEFECTIVE AND UNUSABLE. OUR AGENT HAD TO URGENTLY DELIVER ANOTHER INSTRUMENT, WHICH WAS WITHDRAWN IN AN EMERGENCY FROM ONE OF OUR DEPOSIT ACCOUNTS, CAUSING CONSIDERABLE INCONVENIENCE TO BOTH CUSTOMERS AND SIGNIFICANTLY LENGTHENING THE DURATION OF THE OPERATION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22064169 ↗
  • MalfunctionMay 13, 2025Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    AS REPORTED: "T2 TIBIA NAIL HOLDER BOLT REF 18060370 THAT NO LONGER FITS IN THE T2 TIBIA NAIL HOLDER REF 18061002."

    Patient problems: Insufficient Information

    MAUDE report 22015057 ↗
  • MalfunctionApr 8, 2025Mechanical Jam; Mechanical Jam

    AS REPORTED: "INTRAOPERATIVELY, THE NAIL RETAINING PIN (1806-0370) COULD NO LONGER BE DETACHED FROM THE NAIL. THIS MEANT THAT THERE WAS A NON-DETACHABLE CONNECTION BETWEEN THE NAIL RETAINING PIN, THE NAIL ADAPTER 1806-1002) AND THE NAIL ITSELF. THE OPERATION HAD TO BE CONTINUED WITHOUT THE TARGETING DEVICE AND WAS COMPLETED MANUALLY. OP TIME EXTENDED (90 MIN), INCREASED RADIATION DUE TO HANDS-FREE LOCKING."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21794740 ↗
  • InjuryAug 22, 2024Mechanical Jam; Mechanical Jam

    THE SALES REP REPORTED THAT: "DURING THE OPERATION, THE SURGICAL TEAM WAS UNABLE TO COMPLETE THE PROCEDURE DUE TO A TECHNICAL PROBLEM WITH THE ANCILLARY PARTS. AFTER REAMING THE PATIENT'S TIBIA AND IDENTIFYING THE IMPLANT TO BE PLACED, THE OPERATORS WERE UNABLE TO CONTINUE THE PROCEDURE DUE TO A PROBLEM WITH THE ASSEMBLY AND DISASSEMBLY OF THE BOLT, NAIL HOLDER AND IMPLANT. THE BOLT AND NAIL HOLDER REMAINED ATTACHED, MAKING IT IMPOSSIBLE TO DISMANTLE THE IMPLANT ONCE THE PROCEDURE HAD BEEN COMPLETED. FOR SAFETY REASONS, THE OPERATORS PREFERRED NOT TO IMPLANT THE IMPLANT AND TO POSTPONE THE OPERATION TO A LATER DATE. INTERVENTION POSTPONED TO 07/31 INSTEAD OF 07/27. THE OPERATION WAS CARRIED OUT BY AN EXPERIENCED SURGICAL TEAM. THE ¿RECOVERY¿ OPERATION TOOK PLACE IN THE PRESENCE OF THE LOCAL SALES MANAGER, AND WENT SMOOTHLY. THE FAULTY EQUIPMENT WAS TAKEN BACK BY THE SALES REPRESENTATIVE FOR EXPERT EXAMINATION."

    Patient problems: Insufficient Information

    MAUDE report 20043007 ↗
  • MalfunctionApr 17, 2023Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    CUSTOMER REPORTED THAT: "THIS DEVICE PRESENTS THE FOLLOWING ANOMALY: IMPOSSIBLE TO LOWER THE NAIL FIXING SCREW INTO THE NAIL, PREVENTING IT FROM BEING FIXED ON THE NAIL HOLDER."

    Patient problems: Insufficient Information

    MAUDE report 16751380 ↗
  • MalfunctionNov 8, 2022Mechanical Jam; Mechanical Jam

    AS REPORTED: "TIBIA NAIL T2 LOCKED WITH NAIL HOLDING SCREW. COLD WELDING. NAIL COULD NOT BE COMPLETELY FIXED WITH NAIL ADAPTER AND NAIL HOLDING SCREW."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15756465 ↗
  • MalfunctionSep 5, 2022Mechanical Jam; Mechanical Jam

    AS REPORTED: "THE INCIDENT OCCURRED IN THE OPERATING ROOM ON (B)(6) 2022 AND CONCERNED THE PLACEMENT OF A T2 TIBIAL NAIL. AFTER THE NAIL WAS PLACED, IT WAS IMPOSSIBLE TO REMOVE THE NAIL ANCILLARY. THE FIXATION SCREW ASSEMBLY, TIBIA NAIL HANDLE AND NAIL SEIZED. THE NAIL HAD TO BE REMOVED FROM THE PATIENT (WITH THE HANDLE STILL PRESENT AT THE END) AND ANOTHER ONE HAD TO BE INSERTED (SMALLER SIZE, AS THERE IS ONLY ONE OF EACH SIZE) USING THE SECOND BOX OF ANCILLARY EQUIPMENT PRESENT IN THE OPERATING ROOM. THE OPERATION ENDED WELL FOR THE PATIENT, BUT THE PROCEDURE WAS LENGTHENED AND THE NAIL PLACED WAS SLIGHTLY SHORTER THAN INITIALLY PLANNED. THE NAIL - NAIL HANDLE DISASSEMBLY WAS PERFORMED AFTER THE PROCEDURE: THE NAIL TIP BROKE OFF AND SO DID THE FOLLOWING ANCILLARIES DUE TO THE FACT THAT THEY HAD TO BE FORCED OUT DUE TO SEIZURE. THE EXPLANTED NAIL, THE GIMBAL WRENCH, THE HANDLE AND THE FIXATION SCREW WERE RECOVERED AND DECONTAMINATED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15354732 ↗
  • MalfunctionMay 24, 2022Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    AS REPORTED: "THE NAIL HOLDING SCREW CAN NO LONGER BE PUSHED THROUGH THE NAIL HOLDING BRACKET."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14487508 ↗
  • MalfunctionOct 7, 2021Separation Failure; Misassembly by Users; Separation Failure; Misassembly by Users

    IT WAS REPORTED THE NAIL WAS ATTACHED TO THE HANDLE BACKWARDS AND WOULD NOT COME OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12590166 ↗
  • MalfunctionSep 9, 2021Device Handling Problem; Device Handling Problem

    AS REPORTED: "THERE WERE PROBLEMS WITH INSERTION WHEN THE TIBIAL NAIL WAS DRIVEN IN. IMPACT DOME AND NAIL PICK-UP PRESUMABLY DEFECTIVE. ITEM IS 10 YEARS OLD. ITEM HAS BEEN RE-PROCESSED 100 TIMES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12446961 ↗
  • MalfunctionAug 10, 2021Difficult to Insert; Device Difficult to Setup or Prepare; Difficult to Insert; Device Difficult to Setup or Prepare

    IT WAS REPORTED THAT THE NAIL HOLDING SCREW (1806-0370) WOULD NOT PASS THROUGH THE TARGETER ATTACHMENT (1806-1002). THE SURGEON HAD TO SWITCH NAIL SYSTEMS AS THE CASE COULD NOT BE COMPLETED WITHOUT THESE INSTRUMENTS. THE SURGEON TRIED THE FIRST SCREW AND IT DIDN¿T PASS THROUGH. THE SURGEON TRIED THE SECOND WITH THE SAME OUT COME. THE SCREWS LOST SOME OF THE THREADS WHILE ATTEMPTING TO PASS THEM THROUGH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12304123 ↗
  • MalfunctionApr 9, 2021Separation Failure; Mechanical Jam; Separation Failure; Mechanical Jam

    THE CUSTOMER REPORTED THAT : "STRYKER TIBIA NAIL FIXATION SLEEVE STUCK IN THE TIBIA NAIL HOLDER (1806-1002 - N°KH171892) AND IN THE NAIL: IMPOSSIBLE TO UNSCREW THEM AND THEREFORE TO PLACE THE NAIL (1822-1034S - N°K03F75C) IN THE PATIENT, THEREFORE CHANGE THE SIZE OF THE NAIL AND PLACE A NAIL WITHOUT AN ANCILLARY. IN THE END, THERE IS NO CLINICAL IMPACT FOR THE PATIENT. "

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11641957 ↗
  • MalfunctionFeb 18, 2021Fracture; Fracture

    "IT WAS REPORTED THAT DURING A FEMORAL NAIL CASE, A PIECE OF METAL BROKE OFF OF THE TARGETING JIG. NO STRYKER REP WAS PRESENT FOR THE CASE, REP WAS NOTIFIED DAYS AFTER THE CASE. THE HOSPITAL DECIDED TO GO WITH EXTERNAL FIXATION INSTEAD OF CONTINUING THE NAIL CASE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11346647 ↗
  • MalfunctionDec 30, 2020Difficult or Delayed Separation; Difficult or Delayed Separation

    AS REPORTED: "T2 TIBIA NAIL UNABLE TO BE DISCONNECTED FROM NAIL ADAPTER."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11098087 ↗
  • MalfunctionDec 4, 2020Difficult or Delayed Separation; Difficult or Delayed Separation

    SALES REP REPORTED THAT: " THE FIXING SCREW HAS REMAINED STUCK IN THE NAIL HOLDER." SURGERY WAS COMPLETED SUCCESSFULLY "BY OPENING THE SECOND DEPOSIT AT THE ESTABLISHMENT." CLINICAL CONSEQUENCES FOR THE PATIENT: "NO, BECAUSE THE TEAM CHECKED THE ASSEMBLY OF THE DEVICES BEFORE WORKING ON THE PATIENT AND THAT BY CHANCE THE SECOND SET OF INSTRUMENTS IN STORAGE HAD NOT BEEN USED." DELAYED BY 15 MIN. ADDITIONAL MEDICAL INTERVENTION NECESSARY: NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10946162 ↗
  • MalfunctionSep 25, 2020Difficult or Delayed Separation; Difficult or Delayed Separation

    THE END USER HAS REPORTED: EVENT DESCRIPTION : IT¿S A NEW DEPOT FOR T2 TIBIA, ONLY USE ONCE. THE NAIL HOLDING SCREW FOR TIBIA (REF 1806-0370) GOT STUCK IN THE TARGETING DEVICE (REF 1806-1002). WITH GREAT DIFFICULTY WE WERE ABLE TO REMOVE IT TO USE THE SECOND NAIL HOLDING SCREW AND WE HAVE THE SAME PROBLEM¿ AND IMPOSSIBLE TO CONNECT THE NAIL TO THE TARGET DEVICE ¿ AND IMPOSSIBLE TO USE THE T2 TIBIA NAIL. THE SURGEON HAD TO USE AN IMPLANT FROM AN OTHER COMPANY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10586510 ↗
  • MalfunctionMar 20, 2020Difficult or Delayed Separation; Difficult or Delayed Separation

    CUSTOMER REPORTED THAT : "PROCEDURE: DOUBLE TIBIA-FIBULA FRACTURE. PLACEMENT OF A TIBIAL NAIL WITH A HANDPIECE. AT THE END OF THE OPERATION, DURING THE REMOVAL OF THE HANDPIECE, THE TIBIAL NAIL REMAINED STUCK ON THE NAIL HOLDER OF THE ANCILLARY. AFTER MULTIPLE ATTEMPTS, IT WAS IMPOSSIBLE TO DETACH THE NAIL FROM THE ANCILLARY. THE SURGEON WAS FORCED TO EXPLANT THE NAIL AND SCREWS PUT IN PLACE AND PERFORM A NEW BORE. CHANGE OF NAIL WITH OBLIGATION TO DRIVE A BIGGER NAIL BECAUSE NO OTHER NAILS OF THE SAME SIZE IN STOCK." DELAY : 1H OPENING OF A NEW ANCILLARY INSTRUMENT SET AND PLACEMENT OF A BIGGER NAIL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9862995 ↗
  • MalfunctionJan 28, 2020Difficult to Remove; Mechanical Jam; Difficult to Remove; Mechanical Jam

    IT WAS REPORTED THAT THE "TIBIAL ALIGNMENT GUIDE IN QUESTION EXHIBITED GREAT DIFFICULTY IN ATTACHING TO THE NAIL (VIA THE NAIL HOLDING BOLTS). SURGEON TRIED FIRST BOLT, IT WOULD NOT ATTACH AND NEARLY GOT PERMANENTLY STUCK INSIDE THE ALIGNMENT HANDLE. 2ND BOLT WAS FIRMLY ATTACHED TO THE NAIL HOWEVER COULD NOT BE SUBSEQUENTLY DISASSEMBLED." SURGICAL DELAY OF 30-35 MINUTES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9637500 ↗
  • MalfunctionJul 24, 2019Material Erosion; Unraveled Material

    IT WAS REPORTED TO THE REP THAT UPON INSPECTION, THE HOLE FOR THE STRIKE PLATE IN THE STANDARD T2 TIBIA NAIL HOLDING ARM WAS OBSERVED TO BE COUNTER-THREADED AND FRAYING. NO ASSOCIATED PROCEDURE.

    Patient problems: No Patient Involvement

    MAUDE report 8824863 ↗
  • InjuryJun 7, 2019Therapeutic or Diagnostic Output Failure

    THE PHARMACIST FROM THE HOSPITAL REPORTED THE FOLLOWING EVENT : "PATIENT PAS, FEMALE, 23 YEARS OLD, 60 KG. WHEN USING THE NAIL, THE NAIL DOES NOT COME OFF THE NAIL HOLDER. THE PROBLEM HAD ALREADY OCCURRED, THE EQUIPMENT WAS EXCHANGED WITH THE LABORATORY AND THE PROBLEM RECURRED. LABORATORY NOTIFIED AND BROUGHT US A NEW NAIL HOLDER WITH LOCKING SCREWS. USE OF ANOTHER INSTRUMENT PANEL. NEW INFORMATION RECEIVED ON MAY 13TH 2019 INDICATED THAT THERE WAS AN IMPACT ON THE PATIENT, ESPECIALLY DURING THE FIRST PROCEDURE. AS A REMINDER, IT WAS AN OPEN LEG FRACTURE, BUT TOTALLY ACCESSIBLE TO A CENTRO-MEDULLARY NAILING. SURGEON APPROACHED AND DRILLED THE UPPER END OF THE TIBIA, HE MADE A COMPLETE BORE OF THE TIBIA (FOR NOTHING. FOR INFORMATION: RISK OF INCREASED FAT EMBOLISM DUE TO THE USE OF A SHEET), BEFORE HE REALIZED THAT THE NAIL INSTRUMENT WAS NOT WORKING AND THEREFORE CHANGE THE MEANS OF FIXATION (SO: EXTERNAL FIXATOR). THERE WAS A LOSS OF TIME OF ABOUT 30 MINUTES, THE TIME TO LOOK FOR OTHER BOXES, TO REALIZE THAT THEY WERE NOT WORKING EITHER, TO CALL LYON SOUTH TO "HELP US OUT", TO HAVE THE EXTERNAL FIXER BOX LOOKED FOR. DURING THE 1ST OPERATION, AFTER REMOVING THE INSTRUMENT SET FROM

    Patient problems: Discomfort; Myocardial Infarction

    MAUDE report 8680262 ↗
  • MalfunctionJan 2, 2019Break

    DURING A ROUTINE INSPECTION OF HOSPITAL'S INVENTORY, ONE OF THE TWO PRONGS ON THE TIBIAL NAIL HOLDING ARM THAT KEEP THE ARM FROM ROTATING WAS FOUND TO BE BROKEN OFF. NO ASSOCIATED PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8210633 ↗
  • MalfunctionDec 6, 2018Difficult to Insert

    IN ORDER TO ATTACH THE NAIL TO THE NAIL HANDLE, TRIED TO INSERT A HOLDING SCREW FROM THE POST. THE SCREW DID NOT ENTER ALL THE WAY AND THE THREAD PART DID NOT REACH THE NAIL. THERE WAS ANOTHER SPARE SCREW, BUT IT COULD NOT ENTER INTO THE POST AGAIN. FORCED IT IN, BUT COULD NOT ATTACH THE NAIL TO THE HANDLE. SHAVED OFF BOTH THE HANDLE AND THE SCREW WITH RASP, AND MANAGED TO CONTINUE THE OPERATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8139825 ↗
  • MalfunctionSep 25, 2015Fitting Problem

    IT WAS REPORTED THAT WHILE TRYING TO ASSEMBLE A T2 TIBIAL NAIL ON TO THE NAIL HOLDING ASSEMBLY, THE NAIL HOLDING BOLT WOULD NOT DROP THROUGH THE NAIL HOLDING ASSEMBLY SO WE COULD NOT ATTACH THE NAIL. WHEN INSPECTING THE ASSEMBLY, THERE IS A NOTICEABLE "ETCHING" IN THE ASSEMBLY CANAL WHERE THE BOLT DROPS THROUGH PROHIBITING THE SCREW FROM DROPPING COMPLETELY DOWN TO ITS DESIRED POSITION. AS A RESULT, THE SURGEON CHOSE TO USE A DIFFERENT VENDOR TO COMPLETE THE SURGERY. IT WAS A PRIMARY SURGERY (ORIF TIBIA).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5105212 ↗

FDA clearance and classification

The FDA regulates product code LXH (orthopedic manual surgical instrument) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the T2 1806-1002

Has the T2 1806-1002 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04546540196538), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the T2 1806-1002?

The FDA's MAUDE database lists 25 adverse event reports for this device, received between Sep 25, 2015 and Feb 21, 2026. In those reports, the most common specific device problems are “Mechanical Jam” (8), “Device Difficult to Setup or Prepare” (5) and “Difficult or Delayed Separation” (4); the most common patient problems are “Discomfort” (1) and “Myocardial Infarction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the T2 1806-1002?

None of the reports are classified as deaths, 2 as injuries and 23 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the T2 1806-1002 FDA approved?

Devices of this type (product code LXH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker GmbH.

Who makes the T2 1806-1002, and is it still sold?

The FDA's device database lists Stryker GmbH as the company responsible for it (brand T2, model 1806-1002). Its record shows it as in commercial distribution.