Skip to content

T2 1806-1006Recalls & reported problems

Stryker GmbH · Nail Handle · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540196552), received between Sep 29, 2015 and Sep 1, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Device Slipped2
  • Failure to Align1
  • Fracture1
  • Material Fragmentation1
  • Positioning Failure1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 1, 2026Fracture

    Patient problems: Foreign Body In Patient

    MAUDE report 50543812 ↗
  • MalfunctionOct 8, 2020Positioning Failure; Failure to Align; Positioning Failure; Failure to Align

    FEMUR TARGET DEVICE 1806-1008/1806-1006, MISSDRILLING. STRIKE PLATE 1806-0150 COULD NOT BE SCREWED IN REPLACEMENT WAS AVAILABLE. SURGICAL DELAY OF 30MIN. NOT LONGER. NO OTHER CONSEQUENCES REPORTED. ADDITIONAL INFORMATION RECEIVED FROM THE OP LEAD: THE MISDRILLING OCCURRED AT THE PROXIMAL END THE ISSUE WAS RESOLVED BY USING SECOND TARGETING ARM AND LATER ORDERED NO PATIENT IMPACT, ONLY 30MIN OP TIME INCREASE THE THREADING OF STRIKE PLATE IMPACTED IN A WAY THAT IT CANNOT BE THREADED INTO THE TARGETING ARM ANYMORE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10653188 ↗
  • MalfunctionMar 27, 2020Material Fragmentation; Material Fragmentation

    IT WAS REPORTED, METAL ABRASION OCCURRED WHEN CONNECTING THE STRIKE PLATE AND THE NAIL HANDLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9891025 ↗
  • MalfunctionDec 19, 2018Break

    BROKEN IMPACTOR IN TARGETING DEVICE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8182769 ↗
  • MalfunctionSep 29, 2015Break; Device Slipped

    IT WAS REPORTED THAT DURING A RETRO GRADE FEMORAL NAIL PROCEDURE, THE THREADS THAT HOLD THE IMPACTOR ON ARE STRIPPED ON BOTH OF THE T2 FEMORAL NAIL TRAYS - BOTH SLEEVES ARE STRIPPED AND DO NOT HOLD THE NAIL ANY LONGER. NO ADVERSE CONSEQUENCE TO PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5110749 ↗
  • MalfunctionSep 29, 2015Break; Device Slipped

    IT WAS REPORTED THAT DURING A RETRO GRADE FEMORAL NAIL PROCEDURE, THE THREADS THAT HOLD THE IMPACTOR ON ARE STRIPPED ON BOTH OF THE T2 FEMORAL NAIL TRAYS - BOTH SLEEVES ARE STRIPPED AND DO NOT HOLD THE NAIL ANY LONGER. NO ADVERSE CONSEQUENCE TO PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5110736 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page