T2 1806-1006Recalls & reported problems
Stryker GmbH · Nail Handle · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540196552), received between Sep 29, 2015 and Sep 1, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Device Slipped2
- Failure to Align1
- Fracture1
- Material Fragmentation1
- Positioning Failure1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 1, 2026Fracture
Patient problems: Foreign Body In Patient
MAUDE report 50543812 ↗MalfunctionOct 8, 2020Positioning Failure; Failure to Align; Positioning Failure; Failure to Align
FEMUR TARGET DEVICE 1806-1008/1806-1006, MISSDRILLING. STRIKE PLATE 1806-0150 COULD NOT BE SCREWED IN REPLACEMENT WAS AVAILABLE. SURGICAL DELAY OF 30MIN. NOT LONGER. NO OTHER CONSEQUENCES REPORTED. ADDITIONAL INFORMATION RECEIVED FROM THE OP LEAD: THE MISDRILLING OCCURRED AT THE PROXIMAL END THE ISSUE WAS RESOLVED BY USING SECOND TARGETING ARM AND LATER ORDERED NO PATIENT IMPACT, ONLY 30MIN OP TIME INCREASE THE THREADING OF STRIKE PLATE IMPACTED IN A WAY THAT IT CANNOT BE THREADED INTO THE TARGETING ARM ANYMORE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10653188 ↗MalfunctionMar 27, 2020Material Fragmentation; Material Fragmentation
IT WAS REPORTED, METAL ABRASION OCCURRED WHEN CONNECTING THE STRIKE PLATE AND THE NAIL HANDLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9891025 ↗MalfunctionDec 19, 2018Break
BROKEN IMPACTOR IN TARGETING DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8182769 ↗MalfunctionSep 29, 2015Break; Device Slipped
IT WAS REPORTED THAT DURING A RETRO GRADE FEMORAL NAIL PROCEDURE, THE THREADS THAT HOLD THE IMPACTOR ON ARE STRIPPED ON BOTH OF THE T2 FEMORAL NAIL TRAYS - BOTH SLEEVES ARE STRIPPED AND DO NOT HOLD THE NAIL ANY LONGER. NO ADVERSE CONSEQUENCE TO PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5110749 ↗MalfunctionSep 29, 2015Break; Device Slipped
IT WAS REPORTED THAT DURING A RETRO GRADE FEMORAL NAIL PROCEDURE, THE THREADS THAT HOLD THE IMPACTOR ON ARE STRIPPED ON BOTH OF THE T2 FEMORAL NAIL TRAYS - BOTH SLEEVES ARE STRIPPED AND DO NOT HOLD THE NAIL ANY LONGER. NO ADVERSE CONSEQUENCE TO PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5110736 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page