T2 1806-4260Recalls & reported problems
Stryker GmbH · Drill, AO · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker orthopedic drill is used during surgery to place screws for bone nails. Across 24 reports, it mostly describes drill bits snapping or jamming while drilling near the nail, sometimes leaving broken fragments in the patient's bone. There are 3 reported injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 04546540196767), received between Jan 8, 2020 and Apr 7, 2026 — classified as 3 injuries, 21 malfunctions.
The most frequently reported specific device problem is “Fracture” (22 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture22
- Break10
- Failure to Align2
- Material Deformation1
- Mechanical Jam1
Patient problems
- Insufficient Information7
- No Clinical Signs, Symptoms or Conditions6
- No Known Impact Or Consequence To Patient6
- Foreign Body In Patient5
- Device Embedded In Tissue or Plaque4
- Perforation1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 7, 2026Fracture
AS REPORTED: "DURING SURGERY TO INSERT A T2 ANKLE NAIL, A 4.2 MM DRILL BIT BROKE AT BOTH THE PROXIMAL AND DISTAL FIXATION SCREW INSERTION SITES AFTER COMING INTO CONTACT WITH THE NAIL. THIS RESULTED IN THE BREAKAGE OF TWO DRILL BITS. AS THE HOSPITAL HAD A GMMA4 FIXATION DRILL OF THE SAME DIAMETER IN STOCK, THIS WAS USED AS A SUBSTITUTE TO COMPLETE THE OPERATION.".
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 24807306 ↗MalfunctionApr 1, 2026Fracture
AS REPORTED: "WHEN A DRILL WAS USED TO INSERT A SCREW FROM THE POSTERIOR SIDE OF THE CALCANEUS, THE DRILL DID NOT PASS THROUGH THE NAIL HOLE AND INSTEAD CONTACTED THE FRONT OF THE NAIL. THE DRILL BROKE AND THE BROKEN PIECES REMAINED INSIDE THE CALCANEUS. AN ADDITIONAL SKIN CUT WAS REQUIRED TO REMOVE THE DRILL FROM THE BONE, WHICH EXTENDED THE SURGERY TIME."
Patient problems: Perforation
MAUDE report 24755385 ↗MalfunctionMar 6, 2026Failure to Align; Material Deformation; Mechanical Jam
AS REPORTED: "DURING PROXIMAL TIBIAL DRILLING, THE MIS-DRILL WAS OCCURRED AND THE DRILL TIP BECAME STUCK IN THE HOLE. THE DRILL WAS CONTINUED TO BE ROTATED WHILE STUCK, RESULTING IN DAMAGE TO THE TIP."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24532803 ↗InjuryMay 6, 2025Failure to Align; Failure to Align
AS REPORTED: "ANKLE NAIL WAS INTRODUCED. SCREWS WERE LOCKED. MOST DISTAL CALCANEAL SCREW TRAJECTORY WAS DRILLED WITH A 4.2MM DRILL, TROUGH THE TARGETING DEVICE. THE DRILL DIDN'T END UP IN THE HOLE BUT SCRATCHED THE NAIL ON THE SIDE. THE NAIL WAS REMOVED, TESTED AND RE-INSERTED. THE INCIDENT HAPPENED AGAIN. THIS WAS TRIED 3 TIMES. SURGERY WAS DELAYED BY 2 HOURS. A PLATE WAS USED TO FUSE THE ANKLE, WHICH LEAD TO EXTRA INCISION AND DUE TO THE PLATE, NO BONE GRAFT COULD BE USED, WHICH WAS PLANNED."
Patient problems: Insufficient Information
MAUDE report 21966995 ↗MalfunctionApr 22, 2025Break; Fracture; Break; Fracture
AS REPORTED: "WHEN DRILLING FOR INSERTION OF A PA SCREW INTO THE CALCANEUS, THE DRILL MISSED THE SCREW HOLE AND GRAZED THE NAIL. WHEN TRYING TO REMOVE THE DRILL WHILE TURNING, THE DRILL BIT BROKE OFF. WHEN DRILLING AGAIN WITH A SPARE DRILL, IT BROKE OFF IN THE SAME WAY".
Patient problems: Insufficient Information
MAUDE report 21883314 ↗MalfunctionApr 22, 2025Break; Fracture; Break; Fracture
AS REPORTED: "WHEN DRILLING FOR INSERTION OF A PA SCREW INTO THE CALCANEUS, THE DRILL MISSED THE SCREW HOLE AND GRAZED THE NAIL. WHEN TRYING TO REMOVE THE DRILL WHILE TURNING, THE DRILL BIT BROKE OFF. WHEN DRILLING AGAIN WITH A SPARE DRILL, IT BROKE OFF IN THE SAME WAY".
Patient problems: Insufficient Information
MAUDE report 21883053 ↗MalfunctionFeb 26, 2025Break; Fracture; Break; Fracture
AS REPORTED: "THE DRILL MADE CONTACT WITH THE NAIL DURING DRILLING AND BROKE OFF."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21470245 ↗InjuryDec 12, 2024Break; Fracture; Break; Fracture
IT WAS REPORTED THAT: "DURING DRILLING OF THE DISTAL LOCKING OF A GAMMA NAIL, THE DRILL BIT BROKE IN 2 IN THE PATIENT. THE DISTAL FRAGMENT, MEASURING APPROXIMATELY 3CM, REMAINED IN THE PATIENT, AS EXTRACTION WAS IMPOSSIBLE. THUS, THE DISTAL LOCKING IS IMPOSSIBLE." ON (B)(6) 2024 - ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER: THE DRILL GUIDE (TARGET DEVICE) OF THE STRYKER ANCILLARY IS USED TO GUIDE THE DRILL FOR DISTAL SCREW PLACEMENT. BEFORE PLACING THE NAIL IN THE PATIENT'S FEMUR, WE SYSTEMATICALLY CHECK THAT THE ANCILLARY IS FUNCTIONING CORRECTLY, WITH THE NAIL IN PLACE ON THE ANCILLARY. YES, THEORETICALLY, THE GUIDE POSITIONS THE DRILL BIT OPPOSITE THE NAIL'S LOCKING HOLE, BUT IN THIS CASE THAT WASN'T THE CASE, AS THE DRILL BIT PIERCED THE FIRST CORTEX OF THE FEMUR, THEN BROKE WHEN PERFORATING THE SECOND. IT MUST HAVE RUBBED ABNORMALLY ON THE NAIL'S LOCKING HOLE, RESULTING IN EXCESSIVE STRESS AND CONSEQUENT BREAKAGE.
Patient problems: Foreign Body In Patient
MAUDE report 20921056 ↗InjuryFeb 26, 2023Break; Fracture; Break; Fracture
AS REPORTED: "AFTER COMPRESSION OF BOTH THE TALOFEMORAL AND TALOCALCANEAL JOINTS, A DRILL FRACTURE OCCURRED WHEN DRILLING INTO THE CALCANEUS FROM THE LATERAL SIDE. THEY GAVE UP ON THE HOLE AND MOVED ON TO THE POSTERIOR CALCANEAL SCREW INSERTION TECHNIQUE. AFTER DRILLING AND MEASURING WITH A DEPTH GAUGE, THE SCREW WAS INSERTED, BUT IT WAS FOUND THAT THE SCREW HAD OB'D. THE SCREW WAS REMOVED AND THE DRILL WAS BROKEN WHEN DRILLING AGAIN. TWO SCREWS THAT COULD BE INSERTED INTO THE CALCANEUS THROUGH THE NAIL WERE NOT INSERTED. INSTEAD, TWO SCREWS WERE INSERTED BY FREEHAND DRILLING FROM OUTSIDE THE NAIL. NO BROKEN PIECES OF THE DRILL WERE LEFT BEHIND IN THE BODY." . THIS RESULTED IN A 30 MINUTE SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16441208 ↗MalfunctionFeb 26, 2023Break; Fracture; Break; Fracture
AS REPORTED: "AFTER COMPRESSION OF BOTH THE TALOFEMORAL AND TALOCALCANEAL JOINTS, A DRILL FRACTURE OCCURRED WHEN DRILLING INTO THE CALCANEUS FROM THE LATERAL SIDE. THEY GAVE UP ON THE HOLE AND MOVED ON TO THE POSTERIOR CALCANEAL SCREW INSERTION TECHNIQUE. AFTER DRILLING AND MEASURING WITH A DEPTH GAUGE, THE SCREW WAS INSERTED, BUT IT WAS FOUND THAT THE SCREW HAD OB'D. THE SCREW WAS REMOVED AND THE DRILL WAS BROKEN WHEN DRILLING AGAIN. TWO SCREWS THAT COULD BE INSERTED INTO THE CALCANEUS THROUGH THE NAIL WERE NOT INSERTED. INSTEAD, TWO SCREWS WERE INSERTED BY FREEHAND DRILLING FROM OUTSIDE THE NAIL. NO BROKEN PIECES OF THE DRILL WERE LEFT BEHIND IN THE BODY." THIS RESULTED IN A 30 MINUTE SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16441193 ↗MalfunctionJan 11, 2023Break; Fracture; Break; Fracture
AS REPORTED: "DURING A ANKLE ARTHRODESIS T2 NAIL, DRILL BIT BROKE OFF INSIDE THE PATIENT AND COULD NOT BE REMOVED. THE SURGERY DID NOT CAUSE ANY HARM TO THE PATIENT, BUT IT WAS NOT POSSIBLE TO INSERT THE POSTERO-ANTERIOR SCREW IN THE CALCANEUS. THERE WAS A DELAY OF ABOUT 20 MINUTES TRYING TO GET THE BROKEN PIECE OUT".
Patient problems: Foreign Body In Patient
MAUDE report 16141784 ↗MalfunctionApr 20, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A MEDICAL PROCEDURE, 2 CUTTERS OF 4.2 LENGTH 340 BROKE.
Patient problems: Insufficient Information
MAUDE report 14161130 ↗MalfunctionApr 20, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A MEDICAL PROCEDURE, 2 CUTTERS OF 4.2 LENGTH 340 BROKE.
Patient problems: Insufficient Information
MAUDE report 14161129 ↗MalfunctionApr 7, 2022Break; Fracture; Break; Fracture
AS REPORTED: "TIP OF DRILL BIT BROKEN DURING PROCEDURE."
Patient problems: Insufficient Information
MAUDE report 14043797 ↗MalfunctionNov 12, 2021Fracture; Fracture
CUSTOMER REPORTED THE FOLLOWING ISSUE: "DURING PROXIMAL LOCKING, THE DIAM 3.2 LONG 340 DRILL BIT OF THE T2 FEMUR NAIL ANCILLARY BROKE. THE END OF THE DRILL BIT THEREFORE REMAINED IN THE LESSER TROCHANTER. NO POSSIBILITY TO REMOVE IT, IT REPLACED THE DYNAMIC SCREW. HUMAN CONSEQUENCES: THE BROKEN DRILL BIT WAS LEFT IN THE BONE. A PRIORI, TO DATE, NO IMPACT ON THE PATIENT'S CURRENT CONDITION (THE INCIDENT TOOK PLACE ON (B)(6) AT 1AM, AND THE BREAK WAS CLEAN). THIS INCIDENT DID NOT REQUIRE ANY PARTICULAR FOLLOW-UP OF THE PATIENT. IT WAS NOT POSSIBLE TO RECOVER THE REST OF THE DILL BIT AS IT WAS DISPOSED OF IN A NEEDLE CONTAINER".
Patient problems: Foreign Body In Patient
MAUDE report 12802287 ↗MalfunctionSep 2, 2021Fracture; Fracture
THE CUSTOMER REPORTED THAT: "WE HAD A PROBLEM DURING AN INTERVENTION." ON 08/09/2021 THE CUSTOMER ADDED THAT: "DURING A SURGERY USING THE "FEMUR NAIL" ANCILLARY, THE LOCKING BIT BROKE OFF IN THE BONE DURING DISTAL LOCKING. THE BROKEN DRILL BIT WAS REMOVED WITHOUT DIFFICULTY. NO CLINICAL CONSEQUENCES TO DATE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12413916 ↗MalfunctionJan 13, 2021Fracture; Fracture
IT WAS REPORTED VIA MEDWATCH #: (B)(4): "THE DRILL BROKE OFF IN THE PATIENT'S FEMUR." UPDATE 30/DEC/2020 : FROM OP REPORT: ¿THE LATERAL IMAGE APPEARED TO SHOW THAT THE MOST DISTAL SCREW WAS NOT WITHIN THE NAIL. THIS WAS REMOVED. THE END CAP WAS ALSO REMOVED AND THE JOG WAS REPLACED. THE MOST DISTAL SCREW WAS RE-DRILLED. HOWEVER, THIS MET WITH METALLIC RESISTANCE AND WHEN THE DRILL WAS WITHDRAWN, IT APPEARED THAT THE TIP OF THE DRILL HAD BROKEN OFF. FLUOROSCOPIC IMAGES SHOWED THE DRILL TIP EMBEDDED WITHIN THE BONE. A NEW DRILL WAS USED, WHICH ALSO MET WITH METALLIC RESISTANCE. IT APPEARED THAT THE SCREW WAS HITTING ONE OF THE POSTS FROM THE FEMORAL COMPONENT. THE DECISION WAS MADE NOT TO PLACE THE MOST DISTAL SCREW. A MORE PROXIMAL TRANSVERSE SCREW WAS PLACED THROUGH THE JIG. FINAL FLUOROSCOPIC IMAGES SHOWED GOOD FEMORAL ALIGNMENT AND THE DRILL TIP WAS CONTAINED ENTIRELY WITHIN THE BONE.¿ COMPLICATIONS: ¿BROKEN DRILL BIT WITHIN THE DISTAL FEMUR.¿
Patient problems: Foreign Body In Patient
MAUDE report 11168032 ↗MalfunctionJan 6, 2021Fracture; Fracture
THE CUSTOMER REPORTED TO OUR TECHNICAL SERVICE THAT : "IN EARLY AUGUST, ONE OF OUR OPERATING ROOM TEAMS HAD A PROBLEM WITH A DRILL BIT FOR NAILING, REF. 1806-4260, SN (B)(6). WHEN THE DRILL BIT WAS USED TO MAKE THE HOLE FOR THE FEMORAL NAIL, IT BROKE DURING INSERTION AND A PIECE REMAINED IN THE PATIENT'S BONE. IN DIRECT ACTION, OUR BIOMEDICAL WORKSHOP PROCEEDED TO ACQUIRE A NEW DRILL BIT. THE BROKEN DRILL BIT IS IN MY POSSESSION IF IT IS NECESSARY TO RECOVER IT FOR EXPERTISE."
Patient problems: Foreign Body In Patient
MAUDE report 11127659 ↗MalfunctionApr 3, 2020Fracture; Fracture
PHARMACIST RESIDENT REPORTED THAT: " WHILE DRILLING THE FEMUR, THE TIP OF THE DRILL BIT BROKE OFF AND REMAINED IN THE FEMUR. CLINICAL CONSEQUENCES: INCREASED OPERATING TIME (TIME TO REMOVE THE TIP) WITH RISK WITH A PATIENT WITH A COMPLICATED FEMUR FRACTURE. THE SURGEON WAS ABLE TO REMOVE THE TIP USING A NEEDLE HOLDER. THE DM WAS RETAINED AND IS AVAILABLE FOR EXPERT EVALUATION." 10 MINUTE DELAY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9922645 ↗MalfunctionMar 27, 2020Fracture; Fracture
PHARMACIST, REPORTED BY PHONE: "THE DRILL BROKE DURING THE PROCEDURE. THERE WAS A PROCEDURE DELAY. NO FURTHER INFORMATION AVAILABLE, WE WILL PROVIDE YOU MORE DETAILS". ADDITIONALLY REPORTED BY PHARMACIST: "I CONFIRM MATERIOVIGILANCE NOTIFICATION ON PRODUCT 18064260 LOT K227443. THE EVENT HAPPENED DURING PROCEDURE, WHEN INSERTING THE DRILL BIT INTO THE LOCKING HOLE OF THE NAIL. THERE WERE NO CONSEQUENCES ON THE PROCEDURE TIME (REPLACEMENT OF THE DRILL WITH ANOTHER AVAILABLE DRILL) OR ON THE PATIENT (EASY REMOVAL OF THE 2 PIECES). NO MEDICAL INTERVENTION WAS NECESSARY."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9891479 ↗MalfunctionJan 13, 2020Fracture; Fracture
DRILLS BROKE DURING MEDICAL PROCEDURE AND REMAIN IN THE PATIENTS BONE.
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 9584525 ↗MalfunctionJan 13, 2020Fracture; Fracture
DRILLS BROKE DURING MEDICAL PROCEDURE AND REMAIN IN THE PATIENTS BONE.
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 9584524 ↗MalfunctionJan 8, 2020Fracture; Fracture
AS REPORTED: "DURING THE FEMUR OSTEOSYNTHESIS PROCEDURE, THE DOCTOR TRIES TO PERFORM THE PERFORATIONS FOR THE PROXIMAL LOCKING HOLES AND THE DRILL BIT IS FRACTURED, ANOTHER ONE OF THE SAME REFERENCE IS USED WHICH ALSO FRACTURES."
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 9567104 ↗MalfunctionJan 8, 2020Fracture; Fracture
AS REPORTED: "DURING THE FEMUR OSTEOSYNTHESIS PROCEDURE, THE DOCTOR TRIES TO PERFORM THE PERFORATIONS FOR THE PROXIMAL LOCKING HOLES AND THE DRILL BIT IS FRACTURED, ANOTHER ONE OF THE SAME REFERENCE IS USED WHICH ALSO FRACTURES."
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 9567101 ↗
FDA clearance and classification
The FDA regulates product code LXH (orthopedic manual surgical instrument) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the T2 1806-4260
Has the T2 1806-4260 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04546540196767), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the T2 1806-4260?
The FDA's MAUDE database lists 24 adverse event reports for this device, received between Jan 8, 2020 and Apr 7, 2026. In those reports, the most common specific device problems are “Fracture” (22), “Break” (10) and “Failure to Align” (2); the most common patient problems are “Foreign Body In Patient” (5), “Device Embedded In Tissue or Plaque” (4) and “Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the T2 1806-4260?
None of the reports are classified as deaths, 3 as injuries and 21 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the T2 1806-4260 FDA approved?
Devices of this type (product code LXH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker GmbH.
Who makes the T2 1806-4260, and is it still sold?
The FDA's device database lists Stryker GmbH as the company responsible for it (brand T2, model 1806-4260). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,989 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,199 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,163 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT852 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument821 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument559 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT492 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.