Trident 542-11-46DRecalls & reported problems
Howmedica Osteonics Corp. · Cluster Acetabular Shell · Class 2 · Implant
All models of Trident (160)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04546540240460), received between May 11, 2017 and Mar 12, 2021 — classified as 7 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Osseointegration Problem2
- Malposition of Device1
- Material Deformation1
- Mechanical Problem1
Patient problems
- Injury6
- Inadequate Osseointegration3
- Joint Dislocation2
- Pain2
- Erosion1
- Inflammation1
- Reaction1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMar 12, 2021Malposition of Device; Malposition of Device
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION OF THE FEMORAL HEAD FROM THE CONSTRAINED ACETABULAR INSERT. AS THE PATIENT HAS HAD REPEATED REVISIONS FOR DISLOCATION, THE SURGEON DECIDED TO REVISE THE HEAD, LINER, SHELL, AND SCREWS IN ORDER TO RE-POSITION THE NEW SHELL.
Patient problems: Joint Dislocation
MAUDE report 11471805 ↗InjuryJun 4, 2019Osseointegration Problem; Osseointegration Problem
AS REPORTED: "LOOSENING OF ACETABULAR SHELL. REVISION RIGHT TOTAL HIP. REVISED ACETABULAR [COMPONENTS] AND FEMORAL HEAD ONLY". REP INDICATED THAT THE PATIENT WAS REVISED TO STRYKER AND COMPETITOR DEVICES.
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 8667844 ↗InjuryFeb 27, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
THIS PI IS FOR REVISION 1 DONE ON (B)(6) 2017 DUE TO DISLOCATION AND PAIN. IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY ON (B)(6) 2016 AND AGAIN ON (B)(6) 2016 THE PATIENT WAS ADMITTED FOR DISLOCATIONS AND UNDERWENT CLOSED REDUCTIONS OF THE HIP. IT IS FURTHER ALLEGED ON (B)(6) 2017 LESS THAN TWO MONTHS AFTER THE LAST CLOSED REDUCTION SHE WAS BACK IN THE HOSPITAL BECAUSE OF A DISLOCATION AND PAIN. THE CUP AND LINER WERE REMOVED AND REPLACED.
Patient problems: Injury; Joint Dislocation; Pain
MAUDE report 8376067 ↗InjuryFeb 7, 2019Osseointegration Problem; Osseointegration Problem
PATIENT PRESENTED TO CLINIC WITH HIP PAIN. DOCTOR ORDERED CT SCAN AND CUP APPEARED TO BE LOOSE. CUP, DOME HOLE PLUG, SCREW, LINER, AND BIOLOX HEAD TAKEN OUT. REIMPLANTED TRIDENT II CUP.
Patient problems: Inadequate Osseointegration; Injury; Pain
MAUDE report 8318989 ↗InjuryJan 23, 2019Mechanical Problem; Mechanical Problem
PATIENT HAD A RIGHT TOTAL HIP REPLACEMENT (B)(6) 2018. PATIENT TEMPLATED TO A 46MM CUP. DURING SURGERY IT WAS NOTED THAT THE FEMORAL HEAD MEASURED 40MM INDICATING THE MAGNIFICATION IN TEMPLATING WAS INACCURATE. THE SURGEON REAMED TO 46 MM WHICH WAS THE SMALLEST CUP AVAILABLE. HE IMPLANTED A 46 CUP AND HE GOT ONE SCREW IN THE DOME. AT THE TIME OF SURGERY HE WAS CONCERNED ABOUT FIXATION. HE KEPT THE PATIENT TOUCH DOWN WEIGHT BEARING. SHE ROLLED OVER IN BED AND DISLODGED HER CUP. SHE WAS SEEN IN THE ER AND SCHEDULED FOR SURGERY. SHE WAS REVISED TO A 42MM TRIDENT 2 MULTIHOLE CUP WITH 3 SCREWS. FIXATION APPEARED TO BE ADEQUATE AT THE TIME OF REVISION. THE PATIENT WAS STABLE THROUGH THE FULL RANGE OF MOTION. THE OPERATION WAS COMPLETED SUCCESSFULLY. SURGEON ALLEGES NO DEFECT WITH THE IMPLANTS AND SAID PATIENT FACTORS CONTRIBUTED TO THE FAILURE.
Patient problems: Injury
MAUDE report 8272584 ↗InjurySep 21, 2018Material Deformation; Material Deformation
IT WAS REPORTED THROUGH THE ATTORNEY FOR THE PATIENT, AS A RESULT OF A LAWSUIT, THAT ALLEGEDLY THE PATIENT RECEIVED A TRIDENT CERAMIC ON CERAMIC HIP SYSTEM. IT WAS FURTHER ALLEGED THAT, THE PATIENT BEGAN TO EXPERIENCE SOME DIFFICULTIES, REQUIRING A REVISION SURGERY ON (B)(6) 2008. UPDATE AS PER MEDICAL REVIEW: "ON (B)(6) 2008 A REVISION OF THE RIGHT TOTAL HIP WAS PERFORMED WITH REMOVAL OF LOOSE, FRACTURED CERAMIC PIECES. EXTENSIVE METALLOSIS WAS NOTED AND A REVISION TOTAL HIP WAS PERFORMED FOR A DIAGNOSIS OF FRACTURED CERAMIC HEAD AND LINER, DISPLACED METAL LINER, EROSION SUPERIOR ACETABULAR SHELL, AND EXTENSIVE WEAR DEBRIS."
Patient problems: Erosion; Inflammation; Injury; Reaction
MAUDE report 7897741 ↗InjuryMay 11, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED VIA E-MAIL FROM SALES REPRESENTATIVE THAT THE DR. SURGICALLY IMPLANTED AND DID A LEFT HIP REVISION DUE TO LOOSE CUP.
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 6559883 ↗
Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices
- STAGEONE 431209Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 recalls0 reports
- SELF CENTERING 1035-45-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented84 reports
- LFIT 6260-9-036Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented80 reports
- ACTIS 101011060Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented66 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165228Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall1 death12 reports
- SELF CENTERING 1035-44-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented60 reports
- TANDEM 71324046Smith & Nephew, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall9 reports
- TAPERLOC COMPLETE XR 123 MICRO FEMORAL 51-145170Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented59 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page