Skip to content

NA 1806-0292Recalls & reported problems

Stryker GmbH · Screwdriver Shaft 3,5 x 85 mm · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540353283), received between Dec 14, 2021 and Dec 14, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 14, 2021Break; Break

    AS REPORTED: "DURING REMOVAL SURGERY, THE TIP OF THE SCREWDRIVER WAS BROKEN OFF WHEN REMOVING THE END CAP. THE TIP OF THE SCREWDRIVER BROKE OFF JUST SHORT OF THE SURFACE OF THE END CAP, SO IT COULD NOT BE REMOVED. ONLY THE LATERAL LOCKING SCREW WAS REMOVED, AND THE NAIL AND END CAP WERE LEFT IN THE PATIENT BODY."

    Patient problems: Foreign Body In Patient

    MAUDE report 13003416 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page