NA 62-18330Recalls & reported problems
Stryker Leibinger GmbH & Co. KG · IN-SITU PLATE CUTTER · Class 1
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 04546540366719), received between Jul 26, 2018 and Jun 25, 2021 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Fracture” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture9
- Break3
- Detachment of Device or Device Component3
- Material Fragmentation3
- Crack1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Known Impact Or Consequence To Patient7
- Insufficient Information6
- No Consequences Or Impact To Patient3
- No Patient Involvement1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 25, 2021Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
IT WAS REPORTED BY THE COMPANY REPRESENTATIVE THAT DURING THE SURGICAL PROCEDURE, THE SPRING ON THE DEVICE BROKE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12068367 ↗MalfunctionMay 6, 2021Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
IT WAS REPORTED BY A COMPANY REPRESENTATIVE THAT DURING THE SURGICAL PROCEDURE, THE IN-SITU PLATE CUTTER BROKE WHILE CUTTING A PLATE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11787520 ↗MalfunctionMay 6, 2021Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
IT WAS REPORTED BY A COMPANY REPRESENTATIVE THAT DURING THE SURGICAL PROCEDURE, THE IN-SITU PLATE CUTTER BROKE WHILE CUTTING A PLATE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11787519 ↗MalfunctionJan 29, 2021Crack; Crack
IT WAS REPORTED BY THE COMPANY REPRESENTATIVE THAT IT WAS NOTICED DURING THE PROCEDURE, THE DEVICE WAS CRACKED. THERE IS NO OTHER INFORMATION KNOWN AT THIS TIME.
Patient problems: Insufficient Information
MAUDE report 11253820 ↗MalfunctionJan 28, 2021Fracture; Fracture
IT WAS REPORTED BY THE COMPANY REPRESENTATIVE THAT THE BLADE OF THE DEVICE BROKE DURING THE PROCEDURE. A BACKUP WAS USED TO COMPLETE THE SURGERY. NO OTHER INFORMATION IS KNOWN AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11246144 ↗MalfunctionJan 15, 2021Fracture; Fracture
IT WAS REPORTED BY THE COMPANY REPRESENTATIVE THAT WHILE TRYING TO CUT A HOLE IN THE PLATE, THE TIP OF THE BLADE BROKE OFF.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11181248 ↗MalfunctionJan 11, 2021Fracture; Fracture
IT WAS REPORTED BY THE COMPANY REPRESENTATIVE THAT WHILE TRYING TO CUT A HOLE IN HE PLATE, THE TIP OF THE BLADE BROKE OFF.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11151008 ↗MalfunctionAug 13, 2020Fracture; Material Fragmentation; Fracture; Material Fragmentation
IT WAS REPORTED BY THE COMPANY REPRESENTATIVE THAT THE CUTTERS HAVE A CHIP OFF THE TIP OF THE CUTTING BLADE AND THE CUTTERS WILL NOT CLOSE COMPLETELY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10404894 ↗MalfunctionJun 9, 2020Fracture; Fracture
UPON INSPECTION, THE COMPANY REPRESENTATIVE NOTICED THAT THE PLATE CUTTER'S CUTTING BLADES AND TIP ARE BROKEN.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 10134827 ↗MalfunctionJan 27, 2020Material Fragmentation; Material Fragmentation
IT WAS REPORTED BY A COMPANY REPRESENTATIVE THAT THE PLATE CUTTERS HAVE CHIPS ON BOTH TIPS AND ALONG THE CUTTING EDGE.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 9634102 ↗MalfunctionJan 10, 2020Material Fragmentation; Material Fragmentation
IT WAS REPORTED BY THE COMPANY REPRESENTATIVE THAT WHEN THE SURGEON WAS USING THE CUTTERS TO CUT THROUGH A PLATE, THE PLATE DID NOT SHEAR AND INSTEAD THE CUTTERS BEGAN STRIPPING AWAY/FRACTURING IN SMALL INCREMENTS.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9575534 ↗MalfunctionMar 29, 2019Fracture
IT WAS REPORTED THAT THE IN-SITU CUTTER FRACTURED ON THE RIVET HOLE DURING A CRANIOTOMY CASE. THE CUTTER WAS USED TO CUT 1.7MM PLATE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8464852 ↗MalfunctionMar 19, 2019Break; Mechanical Problem
DURING THE PRODUCT QUALITY INVESTIGATION IT WAS FOUND THAT A SMALL FRAGMENT BROKE OUT OF THE UPPER CUTTING EDGE, AND THE SURFACE REVEALS A FRACTURE DUE TO A MECHANICAL OVERLOAD. THIS OCCURRED DURING THE INVESTIGATION, SO THERE WAS NO ASSOCIATED PROCEDURE, AND THERE WERE NO ADVERSE CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8434364 ↗MalfunctionAug 21, 2018Break
IT WAS REPORTED THAT THE IN-SITU PLATE CUTTER BROKE INTRA-OPERATIVELY, WHILE TRYING TO CUT A PLATE. NO MEDICAL INTERVENTION, DELAY OF SURGERY, OR ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7800795 ↗MalfunctionJul 26, 2018Break
IT WAS REPORTED THAT DURING A PROCEDURE THE SURGEON WAS CUTTING A PLATE WHEN THE CUTTERS BROKE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7723418 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.