NA 700347Recalls & reported problems
Stryker GmbH · Drill Bit AO Fitting · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04546540368973), received between Oct 21, 2016 and Mar 5, 2026 — classified as 2 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Fracture” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture8
- Break5
- Difficult to Remove1
Patient problems
- Foreign Body In Patient5
- No Known Impact Or Consequence To Patient4
- Device Embedded In Tissue or Plaque2
- No Code Available1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryMar 5, 2026Fracture
IT WAS REPORTED THAT ""LEFT ANKLE FIXATION, USED THE 2.5 STRYKER DRILL BIT (700347), WHICH SNAPPED, LEAVING 2CM OF DRILL BIT WITHIN THE PATIENTS LEFT MEDIAL MALLEOLUS. THE SURGEON WAS MR FERGUSSON. THE INCIDENT OCCURRED DURING THE PRIMARY SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO SURGICAL DELAY AND THEREFORE NO ADVERSE CONSEQUENCES DUE TO DELAYS. THE CASE WAS COMPLETED SUCCESSFULLY USING AN ALTERNATIVE DRILL BIT AND THE INCIDENT DISCUSSED WITH ORTHOPAEDIC COLLEAGUES ON THE MAINLAND IMMEDIATELY AT THE END OF THE CASE. THE FINAL OUTCOME WAS THAT THE FRACTURE WAS REDUCED AND STABILISED WITH GOOD POSITIONING OF THE METALWORK AND JOINT SPACES. THE SURGEON DOES NOT ANTICIPATE ANY LONG-LASTING EFFECTS AS THERE IS THE ADDITIONAL FIXATION METALWORK ALSO IN SITU. THE OTHER IMPLANTS USED WERE STRYKER 8-HOLE 1/3 TUBULAR PLATE, CORTICAL AND CANCELLOUS SCREWS AND X2 K-WIRES FOR STABILISATION AND THEN REMOVED. I CANNOT BE CERTAIN HOW LONG THE DRILL BIT HAS BEEN IN USE. THE TRAY WHERE THE DRILL BIT HAS BEEN IN USE AS A REUSABLE ITEM WAS INTRODUCED IN 2008 AND THAT TRAY HAS BEEN PROCESSED 117 TIMES. WE HAVE HAD, UP UNTIL RECENTLY 2 OF THE DRILL BITS ON THE TRAY SO VERY DIFFICU
Patient problems: Foreign Body In Patient
MAUDE report 24521792 ↗InjuryDec 16, 2025Fracture; Fracture
AS REPORTED: "GOLD 2.5 DRILL BIT BROKE WITHIN PATELLA DURING USE BY SURGEON. ATTEMPT WAS MADE TO RETRIEVE BROKEN DRILL BIT. DRILL BIT COULD NOT BE RETRIEVED WITHOUT CAUSING DAMAGE TO PATELLA. DRILL BIT REMAINS EMBEDDED IN PATELLA. SURGEON WILL PROVIDE DISCLOSURE TO PATIENT."
Patient problems: Foreign Body In Patient
MAUDE report 23818565 ↗MalfunctionAug 22, 2025Break; Fracture; Break; Fracture
AS REPORTED: "WHEN THE PRODUCT IN QUESTION WAS USED FOR DRILLING TO PLACE EXETER MESH ON THE ACETABULAR SIDE FOR REVISION OF ANOTHER COMPANY'S THA, THE DRILL BROKE OFF DUE TO INTERFERENCE WITH ANOTHER SCREW. ABOUT 1.5 CM OF THE TIP WAS LEFT BEHIND IN THE BODY."
Patient problems: Foreign Body In Patient
MAUDE report 22877646 ↗MalfunctionJun 27, 2025Break; Fracture; Break; Fracture
IT WAS REPORTED: "DRILL BIT TIP BROKE OFF IN PATIENTS TIBIA".
Patient problems: Foreign Body In Patient
MAUDE report 22346884 ↗MalfunctionMar 14, 2020Fracture; Fracture
IT WAS REPORTED BY A NURSE THAT: " WHEN USING THE MOTOR WITH THE 2.5 DIAMETER DRILL BIT (STRAPPING BOX N° 3), THIS ONE BROKE OFF IN THE PATIENT'S PATELLA. CLINICAL CONSEQUENCES: THE PATIENT HAS A PIECE OF DRILL BIT IN THE PATELLA."
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9832760 ↗MalfunctionFeb 5, 2020Fracture; Fracture
AS REPORTED: "USE RIM MESH FOR ACETABULAR. IT BREAKS WHEN USING A DRILL TO FIX THE MESH. THE THIN AND HARD BONE AND THE LONG DRILLING MAY HAVE MADE IT MORE FRAGILE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9674148 ↗MalfunctionDec 17, 2019Fracture; Fracture
IT WAS REPORTED, DURING THA SURGERY, THE DRILL BIT BROKE WHEN THE SURGEON WAS DRILLING TO THE HOLE OF THE MESH. THE TIP OF THE DRILL BIT WAS REMAINED TO THE PATIENT. THIS CASE IS REVISION DUE TO INFECTION WITH NON-STRYKER IMPLANTS. THE INFORMATION DETAILS OF THE PRIMARY SURGERY ARE UNKNOWN. BEFORE THE REVISION SURGERY, CEMENT MOLDING WITH ANTIBIOTICS WAS PERFORMED.
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 9478333 ↗MalfunctionMar 28, 2019Break
THE DRILL BIT BROKE OFF DURING SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8461815 ↗MalfunctionJan 29, 2019Break
PRIMARY PROCEDURE, LEFT PROXIMAL HUMERUS. IT WAS REPORTED THAT A DRILL BIT BOUND UP IN THE 2.5 DRILL GUIDE AND GOT STUCK. WHEN TRYING TO REMOVE THE BIT, APPROXIMATELY 60MM BROKE OFF IN THE GUIDE. PROCEDURE WAS COMPLETED FREE-HANDED, WITH A SURGICAL DELAY OF APPROXIMATELY 2 MINUTES REPORTED. REP REPORTED THAT X-RAYS, MEDICAL RECORDS, AND FURTHER INFORMATION ARE NOT AVAILABLE DUE TO HOSPITAL POLICY.
Patient problems: No Code Available
MAUDE report 8287414 ↗MalfunctionAug 15, 2018Break; Difficult to Remove
DURING ACETABULAR IMPACTION BONE GRAFT SURGERY, THE DRILL BIT BROKE AND WAS LEFT INSIDE THE PATIENT WHEN DRILLING 2ND SCREW HOLE. SINCE IT WAS IMPOSSIBLE TO RETRIEVE THE BROKEN DRILL BIT, THE SURGERY WAS CONTINUED WITH ANOTHER DRILL BIT.
Patient problems: Foreign Body In Patient
MAUDE report 7784905 ↗MalfunctionOct 21, 2016Fracture; Fracture
DURING SPS SURGERY, THE DRILL BROKE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6048564 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page