NA 702495Recalls & reported problems
Stryker GmbH · Direct Measuring Gauge · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540369154), received between Mar 9, 2020 and Dec 16, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement2
Patient problems
- Injury1
- No Known Impact Or Consequence To Patient1
- Pain1
- Perforation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 16, 2025Incorrect Measurement; Incorrect Measurement
AS REPORTED: "AN EXCESSIVELY LONG ASNIS SCREW WAS SELECTED, AND IT PIERCED THE HEAD OF THE COLLUM. THE PATIENT IS IN PAIN AND SAYS THAT THE SPECIALIST USED SCREWS THAT WERE TOO LONG. THE SPECIALIST SAYS THAT IT IS DUE TO THE STRYKER MEASURING INSTRUMENT. THE SCREW WAS REVISED ON (B)(6)."
Patient problems: Pain; Perforation
MAUDE report 23818249 ↗MalfunctionMar 9, 2020Incorrect Measurement; Incorrect Measurement
THE SURGEON REPORTED THAT AS HE WAS USING THE MEASURING DEVICE WITH A GUIDE WIRE, THE DEVICE DID NOT APPEAR TO BE MEASURING CORRECTLY AND IT RESULTED IN THE SURGEON CHOOSING A SCREW THAT WAS TOO LONG. THE SCREW WAS REMOVED AND REPLACED WITH A SHORTER SCREW. THE SURGEON IS NOT SURE WHETHER THE MEASURING DEVICE, K-WIRE OR SCREW ARE AT ISSUE. THERE WAS A DELAY (LENGTH OF DELAY UNKNOWN) BUT THIS WAS LESS THAN 30 MINUTES.
Patient problems: Injury; No Known Impact Or Consequence To Patient
MAUDE report 9806411 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page