GAMMA 1320-0070Recalls & reported problems

Stryker GmbH · Screwdriver Strike Plate · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540377647), received between Feb 27, 2020 and Nov 22, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture3
  • Device Difficult to Setup or Prepare1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2024Break; Fracture; Device Difficult to Setup or Prepare; Break; Fracture; Device Difficult to Setup or Prepare

    DURING INSERTION OF THE LOCKING SCREW OF GAMMA 3, THE TORQUE BECAME HARD, AND THE LOCKING SCREW WAS INSERTED USING PLIERS ON THE HANDLE PART OF THE SCREWDRIVER. THE HANDLE OF THE SCREWDRIVER WAS DAMAGED IN THE PROCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20757488 ↗
  • MalfunctionJul 9, 2024Break; Fracture; Material Deformation; Break; Fracture; Material Deformation

    NAIL HOLDING SCREW DID NOT LOOSEN WITH NORMAL FORCE. EVENTUALLY, THE DEVICE WAS REMOVED, BUT DAMAGE OF THE NAIL HOLDING SCREW AND STRIKE PLATE WAS OBSERVED. DAMAGED MATERIAL THAT COULD BE SEEN IN THE X-RAY WAS REMOVED FROM PATIENT BODY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19704706 ↗
  • MalfunctionSep 16, 2020Fracture; Fracture

    AS REPORTED: "WHEN OPERATING A LONG GAMMA NAIL, INSERT A NAIL 3 SCREW DRIVER STRIKE PLATE (13200070) INTO THE DEVICE WHEN INSERTING THE NAIL INTO THE MEDULLARY CAVITY AND HIT IT WITH A HAMMER. THE OPERATION PROCEEDED WITHOUT PROBLEMS AND ENTERED A CLOSED WOUND, BUT A METAL PIECE LIKE A NEEDLE POINT WAS FOUND ON THE OPERATION TABLE. CHECK IF THERE ARE ANY METAL PIECES IN THE BODY WITH C-ARM, ETC. SINCE NO SUCH METAL PIECES WERE FOUND, CAREFULLY WASH AND CLOSE THE WOUND. WHEN THE INSTRUMENT WAS INSPECTED AFTER AN X-RAY, IT WAS FOUND THAT THE RECESS ON THE TIP OF THE HAMMER 3 SCREW DRIVER STRIKE PLATE (13200070) AND THE METAL PIECE SUCH AS THE NEEDLE POINT WERE ALIGNED. PROBABLY THE CAUSE WAS HITTING WITH A HAMMER 3 SCREW DRIVER STRIKE PLATE (13200070) APPLIED TO THE UNIVERSAL ROD CONNECTION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10542439 ↗
  • MalfunctionFeb 27, 2020Break; Break

    THE SNAP RING ON THE THREAD IS BROKEN. THE SCREWDRIVER CAN NO LONGER BE SCREWED IN. THE BREAKAGE WAS OBSERVED AFTER THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9766460 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →