NA 702459SRecalls & reported problems
Stryker GmbH · Threaded Guide Wire · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04546540381996), received between Jul 16, 2018 and Jun 23, 2026 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Fracture5
- Detachment of Device or Device Component2
Patient problems
- Foreign Body In Patient5
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJun 23, 2026Fracture
IT WAS REPORTED THAT, AFTER PLACING THE GUIDEWIRE, ASNIS CANNULATED DRILL WAS USED. AS THE DRILLING WAS PERFORMED WHILST THE GUIDEWIRE WAS SLIGHTLY CURVED, THE GUIDEWIRE BROKE. AS THE BROKEN GUIDEWIRE COULD NOT BE REMOVED, THE SURGERY WAS COMPLETED WHILE THE BROKEN GUIDEWIRE WAS LEFT IN THE PATIENT'S BODY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 25595529 ↗InjuryDec 17, 2025Fracture; Fracture
IT WAS REPORTED THAT, THE DRILL BROKE DURING USE LEAVING FRAGMENTS INSIDE THE BODY. IN ADDITION TO THE DRILL, THE GUIDE WIRE ALSO BROKE. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 23827725 ↗MalfunctionApr 22, 2025Break; Fracture; Break; Fracture
AS REPORTED: "THE GUIDE PIN: (702459S) BROKE DURING DRILLING WITH A HOLLOW DRILL, AND THE TIP OF THE CANNULATED DRILL (702449) BECAME DISTORTED. AS A RESULT, THE CANNULATED DRILL (702449) BECAME STUCK IN THE DRILL SLEEVE (702465). THE BROKEN PIECES WERE REMOVED FROM PATIENT BONE WITHOUT ANY PARTICULAR DIFFICULTY, AND THE PROCEDURE WAS RESUMED USING NEW DEVICES. THERE WAS NO IMPACT ON THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21883844 ↗MalfunctionAug 10, 2023Break; Fracture; Break; Fracture
AS REPORTED: "A 1.4MM DIAMETER GUIDE PIN WAS USED TO FIX THE ENDOCARP OF THE TIBIA, BUT THE TIP OF THE THREADED PORTION BROKE OFF WHEN IT WAS PULLED OUT AND REMAINED IN THE BONE."
Patient problems: Foreign Body In Patient
MAUDE report 17504970 ↗MalfunctionDec 10, 2021Break; Break
AS REPORTED: "THE GUIDEWIRE BROKE DURING DRILLING, LEAVING THE TIP IN THE PATIENT'S BODY."
Patient problems: Foreign Body In Patient
MAUDE report 12975200 ↗MalfunctionApr 18, 2019Break; Fracture; Detachment of Device or Device Component
IT WAS REPORTED THAT THE SURGEON WAS USING 4MM CANNULATED SCREWS FOR A FOOT FRACTURE. WHEN TAKING FINAL X-RAYS IT WAS OBSERVED THAT THE TIP OF ONE OF THE GUIDE WIRES HAD SNAPPED OFF EITHER DURING PRE-DRILLING OR SCREW INSERTION (UNKNOWN WHICH). IT WAS DECIDED BY THE SURGEON THAT THIS WOULD NOT AFFECT THE OUTCOME OF THE PATIENT AND THEREFORE IT WAS LEFT IN THE BONE. THE REMAINING LENGTH OF WIRE WAS APPROXIMATED TO BE NO MORE THAN 3MM. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND NO ADVERSE CONSEQUENCES OR SURGICAL DELAY WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8528432 ↗MalfunctionJul 16, 2018Break; Detachment of Device or Device Component
AFTER INSERTING THE GUIDE WIRE INTO THE DRILL OR - 702249, THE GUIDE WIRE BROKE WITHIN THE DRILL IN THE BONE. COULD NOT REMOVE THE TIP OF BROKEN AND REMAIN IT IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 7691867 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page