Skip to content

NA 619905SRecalls & reported problems

Stryker GmbH · Washer · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540430335), received between Aug 18, 2025 and Aug 18, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fracture1
  • Material Deformation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 18, 2025Break; Fracture; Material Deformation; Break; Fracture; Material Deformation

    IT WAS REPORTED THAT: "A SCREW FROM THE ASNIS III RANGE BECAME DEFORMED AND EVENTUALLY BROKE AT THE HEAD DURING ABLATION ON A 14-YEAR-OLD PATIENT, MAKING ABLATION IMPOSSIBLE." MEASURES/ACTIONS TAKEN TO COMPLETE THE PROCEDURE: BREAK THE SCREW FLUSH WITH THE CORTEX AND LEAVE THE REMAINING PART IN PLACE. 3/10/2025 - ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER: "DURING THE REMOVAL PROCEDURE, I EASILY FOUND THE END OF THE SCREW, WITH THE HELP OF A GUIDE PIN, BUT I TRIED TO REMOVE IT FOR 45 MIN (TOURNIQUET TIME 69 MIN INSTEAD OF THE PLANNED 20-30 MIN), WITH A SHARP BROKEN END OF THE SCREW WHEN IT BROKE OFF, POTENTIALLY IRRITATING THE OVERLYING TENDONS." 7/23/2025: UPON DEEP INSPECTION OF THE RETURNED SCREW, IT HAS BEEN NOTICED THAT A WASHER WAS ALSO RETURNED. THE WASHER IS ALSO BROKEN.

    Patient problems: Foreign Body In Patient

    MAUDE report 22835358 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page