Skip to content

NA 91-14008Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · DRILL FOR 8MM SCREWS, STRYKER-SHAFT · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540493675), received between Jul 14, 2023 and Jul 14, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 14, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE TIP OF THE DRILL BROKE OFF INTO THE PATIENT AND REMAINS IN PATIENT. THERE IS NO SIGNIFICANT SURGERY DELAY AND NO REPORTED ADVERSE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17325418 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page