Skip to content

NA 1806-6171Recalls & reported problems

Stryker GmbH · Conical Extractor, male, left hand · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540510075), received between Aug 7, 2018 and Sep 10, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Collapse1
  • Fracture1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE STRYKER EXTRACTION SIEVE THREADED ATTACHMENT WITH THE REF: (B)(4) WAS BROKEN OFF AND COULD NO LONGER BE USED. THIS WAS NOTICED DURING DEVICE INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20179316 ↗
  • MalfunctionJul 13, 2020Material Twisted/Bent; Material Twisted/Bent

    AFTER REMOVING T2GTN, THE END CAP COULD BE REMOVED BY THE USUAL PROCEDURE, BUT THE SURGEON COULD NOT PULL THE NAIL OUT WITH SOFT TISSUE OR CALLUS, AND TRIED TO REMOVE IT WITH THE CONICAL EXTRACTOR (18066171), BUT THE INSERTED ANGLE SEEMED BAD, AND THE TIP WAS DAMAGED. IN THIS CASE, IT WAS REPLACED WITH T2ALPHAGT AT THE PSEUDO JOINT AFTER T2GTN. AFTER REMOVING THE NAIL, REAMING WAS DONE UP TO 15 MM AND INSERT A 14 MM DIAMETER NAIL. HOWEVER, THE SURGEON ATTEMPTED TO REMOVE IT IN THE MIDDLE BECAUSE THE NAIL COULD NOT BE INSERTED. THE SURGEON TRIED TO ATTACH THE UNIVERSAL ROD AND REMOVE THE NAIL, BUT THE NAIL DID NOT COME OUT AT ALL, AND WHEN THE SURGEON WAS OPERATING IT WITH A HAMMER SEVERAL TIMES, THE SURGEON BROKE THE THREAD PART OF THE UNIVERSAL ROD. THE SURGEON COULDN'T PULL OUT THE NAIL, SO HE COULD EXPAND THE INSIDE OF THE MEDULLARY CANAL WITH K-WIRE, AND MANAGED TO PULL OUT THE NAIL USING THE UNIVERSAL ROD AT THE HOSPITAL. THE, HE WAS ABLE TO REAM UP TO 16 MM AGAIN AND INSERT A NAIL. WHEN THE STRIKE PLATE WAS ATTACHED TO THE TIP OF THE UNIVERSAL ROD, THE TIP OF THE UNIVERSAL ROD WAS DAMAGED AND LEFT INSIDE THE STRIKE PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10268041 ↗
  • MalfunctionAug 7, 2018Break; Collapse

    THE PATIENT UNDERWENT T2 TIBIA SURGERY ON (B)(6) 2017, AND THE REMOVAL SURGERY WAS PERFORMED ON (B)(6) 2018. WHEN THE END CAP WAS ATTEMPTED TO BE REMOVED WITH A DRIVER BUT THE HEAD OF THE END CAP COLLAPSED. THEREFORE, THE CONICAL EXTRACTOR WAS USED TO REMOVED THE END CAP BUT THE TIP OF THE EXTRACTOR BROKE IN THE HEAD OF THE END CAP, AND THE BROKEN TIP COULD NOT BE REMOVED. THE SURGEON TRIED TO REMOVED THE END CAP BUT IT COMPLETELY STUCK IN THE NAIL, SO ONLY LOCKING SCREWS WERE REMOVED AND THE NAIL REMOVAL SURGERY WAS GAVE UP. BROKEN FRAGMENT REMAINING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7756163 ↗