Skip to content

NA 1806-6172Recalls & reported problems

Stryker GmbH · Conical Extractor, male, left hand · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540510082), received between May 11, 2020 and Oct 16, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Break2
  • Mechanical Jam1
  • Off-Label Use1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 16, 2025Fracture; Off-Label Use; Mechanical Jam; Fracture; Off-Label Use; Mechanical Jam

    AS REPORTED: "'DESCRIPTION ACCORDING TO THE OPERATING ROOM: INTRAOPERATIVELY, THE THREAD OF INSERT 1806-6172 BROKE. THE INSERT BECAME STUCK IN THE THREAD AND COULD NO LONGER BE DISCONNECTED.'' UPDATE SEPTEMBER 22, 2025. HOW WAS THE EVENT RESOLVED? OTHER SURGICAL TRAYS AND INSTRUMENTS WERE OPENED. WAS THE EXPLANTED IMPLANT A STRYKER/WRIGHT MEDICAL OR FROM COMPETITION? THE PRODUCT WAS ONE FROM THE COMPETITION. IF STRYKER/WRIGHT MEDICAL, WHY DID THE SURGEON DECIDE TO EXPLANT THE DEVICE? PATIENT REQUEST. IF THE DEVICE WAS BROKEN OR DAMAGED: WAS THERE HARD/DENSE BONE? IT WAS HARD DENSE BONE. WAS THERE ANY DEBRIS REPORTED? WERE THEY ALL REMOVED FROM THE SURGICAL FIELD/PATIENT? ALL DEBRIS COULD BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23316684 ↗
  • MalfunctionApr 5, 2023Break; Fracture; Break; Fracture

    AS REPORTED BY THE SALES REP, "SURGEON ENCOUNTERED DIFFICULTIES IN THE REMOVAL OF A T2 TIBIA NAIL. IT PROVED IMPOSSIBLE TO REMOVE THE OBTURATOR PLUG AND, BY EXTENSION, THE NAIL DESPITE THE USE OF STRYKER EXTRACTION KITS. A LEFT-HANDED SCREWDRIVER, REFERENCE 1806-6172, BROKE OFF IN THE PLUG."

    Patient problems: Insufficient Information

    MAUDE report 16685981 ↗
  • MalfunctionFeb 7, 2023Break; Fracture; Break; Fracture

    AS REPORTED: "ONE ITEM BROKE DURING THE SURGERY, MALE EXTRACTOR 4MM. THE ITEM SNAPPED DURING USE, REMOVING A SCREW. THE PROCEDURE WAS DELAYED BY 40 MIN." IT IS UNCLEAR IF ANY DEBRIS WAS LEFT IN THE PATIENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16311921 ↗
  • MalfunctionMar 9, 2021Fracture; Fracture

    "THE CUSTOMER REPORTED THAT WHILST ATTEMPTING TO EXTRACT A CANNULATED HIP SCREW OUT OF A PATIENT, A PIECE OF THE INSTRUMENT SHEARED OFF INTO THE CANNULATED SCREW."

    Patient problems: Insufficient Information

    MAUDE report 11448980 ↗
  • MalfunctionMay 11, 2020Fracture; Fracture

    IT WAS REPORTED, TIP OF EXTRACTION TOOL BROKE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10044642 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page