Stryker GmbH · Screwdriver, Self-Holding, Short · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540520432), received between Dec 9, 2019 and Jul 25, 2023 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture4
  • Break2
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2023Break; Fracture; Break; Fracture

    AS REPORTED: "THE UPPER PART OF THE SCREWDRIVER IS BROKEN OFF."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17393972 ↗
  • MalfunctionDec 8, 2022Break; Fracture; Break; Fracture

    AS REPORTED: "ONE SMALL TOOTH BROKEN OFF SCREWDRIVER."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15947175 ↗
  • MalfunctionDec 2, 2022Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE SCREWDRIVER AND SLEEVE STRIPPED THE HEAD OF A GAMMA SCREW.

    Patient problems: Insufficient Information

    MAUDE report 15901400 ↗
  • InjuryOct 28, 2020Fracture; Fracture

    CASE HAS OCCURRED DURING A G3 SURGERY. WHEN LOCKING THE DISTAL 5MM SCREW, HALF OF THE HEX WAS BROKEN OFF THE SELF-HOLDING SCREWDRIVER. THIS EVENT WAS NOT DETECTED DURING SURGERY. ONLY ONE DAY LATER IN THE CONTROL X-RAY IT WAS SEEN THAT PART OF THE SCREWDRIVER WAS INSIDE THE PATIENT. THE PATIENT WAS RE-OPERATED ON (B)(6) 2020 AND THE BROKEN PART WAS REMOVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 10748699 ↗
  • MalfunctionDec 9, 2019Fracture; Fracture

    THE TIP OF THE SCREWDRIVER BROKE. IT WAS NOTICED DURING INSTRUMENT PREPARATION.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 9441239 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates