VariAx 53-27214SRecalls & reported problems
Stryker GmbH · BONE SCREW, T7 · Class 2 · Implant
All models of VARIAX (345)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540593870), received between Sep 18, 2015 and Nov 14, 2023 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Difficult to Insert2
- Adverse Event Without Identified Device or Use Problem1
- Material Twisted/Bent1
Patient problems
- Device Embedded In Tissue or Plaque1
- Implant Pain1
- Loss of Range of Motion1
- No Code Available1
- No Information1
- No Known Impact Or Consequence To Patient1
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IN 2018, THE CLAIMANT HAD AN ORIF AND WAS IMPLANTED WITH UNKNOWN PINS AND A PLATE, AROUND (B)(6) 2021/(B)(6) 2022, SHE FELT SOME DISCOMFORT/IRRITATION AND SOUGHT MEDICAL CARE. THE REVISION WAS SCHEDULED FOR (B)(6) 2022 TO REMOVE THE IMPLANTED DEVICES. DURING THE SURGERY, THE SURGEON DISCOVERED THAT HE DID NOT HAVE THE CORRECT INSTRUMENTS TO REMOVE THE HARDWARE. THE HOSPITAL SPENT SOME TIME TRYING TO FIND A SUITABLE TO ASSIST IN THE REMOVAL, BUT COULD NOT LOCATE A SUITABLE SCREWDRIVER. HOWEVER, ONE SCREW WAS ABLE TO BE REMOVED THROUGH CAREFUL USE OF A SMALLER SCREWDRIVER HEAD. SINCE THE FAILED SURGERY, THE PATIENT HAS BEEN LEFT WITH A RAISED RED RASH ON HER WRIST, AND IS IN MORE PAIN THAN BEFORE THE FAILED SURGERY. ADDITIONALLY, THIS IS AFFECTING HER WRIST MOVEMENT MORE THAT BEFORE THE SURGERY.
Patient problems: Implant Pain; Loss of Range of Motion; Skin Inflammation/ Irritation
MAUDE report 18132908 ↗MalfunctionMar 29, 2018Difficult to Insert
THE SURGEON REPORTED THAT WHEN IMPLANTING A DISTAL RADIUS PLATE, TWO 2.7MM SCREWS WERE TWISTING IN THE HOLE AND NOT GETTING PURCHASE. THE CASE WAS COMPLETED USING 2 NEW 2.3MM SCREWS AND BY DRILLING 2 NEW HOLES ALONG THE OBLONG. THE PATIENT WAS YOUNG AND WITH GOOD BONE. THERE WAS A SHORT DELAY OF A FEW MINUTES WHILE ADDITIONAL HOLES WERE DRILLED AND NEW SCREWS WERE OBTAINED. THE SURGEON WAS SATISFIED WITH THE OUTCOME OF THE CASE.
Patient problems: No Information
MAUDE report 7383027 ↗MalfunctionMar 29, 2018Difficult to Insert; Material Twisted/Bent
THE SURGEON REPORTED THAT WHEN IMPLANTING A DISTAL RADIUS PLATE, TWO 2.7MM SCREWS WERE TWISTING IN THE HOLE AND NOT GETTING PURCHASE. THE CASE WAS COMPLETED USING 2 NEW 2.3MM SCREWS AND BY DRILLING 2 NEW HOLES ALONG THE OBLONG. THE PATIENT WAS YOUNG AND WITH GOOD BONE. THERE WAS A SHORT DELAY OF A FEW MINUTES WHILE ADDITIONAL HOLES WERE DRILLED AND NEW SCREWS WERE OBTAINED. THE SURGEON WAS SATISFIED WITH THE OUTCOME OF THE CASE.
Patient problems: No Code Available
MAUDE report 7383022 ↗MalfunctionSep 18, 2015Break
DURING VARIAX DR SURGERY, THE SURGEON DRILLED SCREW HOLE WITH A 2.0MM DRILL BIT. THE SURGEON INSERTED THE CORTICAL SCREW INTO THE PLATE SCREW HOLE. HOWEVER, THE SHAFT OF CORTICAL SCREW BROKE. THE SURGEON REMOVED THE BROKEN SCREW. THE SURGEON DRILLED ANOTHER SCREW HOLE WITH A 2.3MM DRILL BIT. THE SURGEON INSERTED THE LOCKING SCREW IN ANOTHER SCREW HOLE. THE SHAFT OF SCREW BROKE AGAIN. THE SURGEON WAS NOT ABLE TO REMOVE THE SHAFT OF BROKEN SCREW. THE SURGEON IS REQUESTING AN IMMEDIATE INVESTIGATION OF THE BROKEN SCREWS.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 5088814 ↗MalfunctionSep 18, 2015Break
DURING VARIAX DR SURGERY, THE SURGEON DRILLED SCREW HOLE WITH A 2.0MM DRILL BIT. THE SURGEON INSERTED THE CORTICAL SCREW INTO THE PLATE SCREW HOLE. HOWEVER, THE SHAFT OF CORTICAL SCREW BROKE. THE SURGEON REMOVED THE BROKEN SCREW. THE SURGEON DRILLED ANOTHER SCREW HOLE WITH A 2.3MM DRILL BIT. THE SURGEON INSERTED THE LOCKING SCREW IN ANOTHER SCREW HOLE. THE SHAFT OF SCREW BROKE AGAIN. THE SURGEON WAS NOT ABLE TO REMOVE THE SHAFT OF BROKEN SCREW. THE SURGEON IS REQUESTING AN IMMEDIATE INVESTIGATION OF THE BROKEN SCREWS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5088774 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page