XIA 3 482318550Recalls & reported problems

Stryker Corporation · POLYAXIAL SCREW · Class 2 · Implant

All models of XIA 3 (93)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540598608), received between May 15, 2017 and Apr 14, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Mechanical Problem2
  • Appropriate Device Problem Term/Code Not Available1
  • Detachment Of Device Component1
  • Loose or Intermittent Connection1
  • Material Deformation1
  • Material Separation1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient2
  • Injury1
  • Insufficient Information1
  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2026Device Dislodged or Dislocated

    A COMPANY REPRESENTATIVE REPORTED THAT THE TULIP OF A XIA 3 POLYAXIAL SCREW DISENGAGED INTRA-OPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24877426 ↗
  • MalfunctionAug 4, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING ROD INSERTION THE TULIP OF THE SCREW SPLAYED WHEN A BLOCKER WAS INSERTED INTO THE SCREW TULIP. IT WAS FURTHER REPORTED THAT THE BLOCKER WAS NOT CROSS THREADED. THE SCREW AND BLOCKER WERE LEFT IMPLANTED. NO ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION WERE REPORTED.  THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12276025 ↗
  • MalfunctionOct 27, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT THE TULIP OF A XIA 3 TITANIUM POLYAXIAL SCREW AT L4 DISENGAGED LESS THAN A MONTH POST-OPERATIVELY. REVISION SURGERY IS NOT SCHEDULED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10742591 ↗
  • InjuryAug 28, 2019Appropriate Device Problem Term/Code Not Available; Migration; Migration

    IT WAS REPORTED THAT PATIENT UNDERWENT A REVISION SURGERY TO REPLACE A XIA 3 TITANIUM POLYAXIAL SCREW WITH A DISENGAGED TULIP HEAD.

    Patient problems: Injury

    MAUDE report 8942147 ↗
  • MalfunctionNov 21, 2017Loose or Intermittent Connection; Mechanical Problem; Device Dislodged or Dislocated; Naturally Worn; Mechanical Problem; Device Dislodged or Dislocated

    IT WAS REPORTED THAT; L5/S1 BY PLIF WAS PERFORMED. WHEN APPLYING COMPRESSION ON THE LEFT SIDE TEMPORARILY FIXED WITH A ROD, THE HEAD OF L5 IS COMPRESSING, THE POSITION OF THE HEAD MOVES CONSIDERABLY AND THEN WHEN TRYING TO FASTEN THE FINAL TIGHTENING WITH A TORQUE WRENCH THE LEFT SIDE OF L5 IDLES AND THE HEAD MOVES. THEREFORE, ONCE THE BLOCKER AND ROD ARE REMOVED, THE HEAD DETACHES. FOR THAT REASON, THE SCREW WAS REPLACED WITH 8.5 X 50 MM OF THE SAME SIZE, COMPRESSION WAS PERFORMED AND FINAL TIGHTENING. WHEN THE DAMAGED SCREW WAS CONFIRMED, THE WASHER BUILT IN THE HEAD WORN OUT AND WAS PARTIALLY DAMAGED. IN ADDITION, PART OF THE SIX POINTER WAS ALSO DAMAGED.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 7049716 ↗
  • MalfunctionMay 15, 2017Detachment Of Device Component; Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT; DURING L5-S TLIF SURGERY, THE SCREW HEAD DETACHED FROM THE SCREW SHAFT WHILE THE FINAL TIGHTENING OF S1 RIGHT SCREW. THEN THE SURGEON REPLACE THE SCREW AND FINISHED THE SURGERY WITHOUT PROBLEM.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 6567328 ↗