VARIAX 628008SRecalls & reported problems
Stryker GmbH · Superior Plate - Decreased Curvature · Class 2 · Implant
All models of VARIAX (345)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540686879), received between Dec 3, 2018 and Mar 30, 2026 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Loosening of Implant Not Related to Bone-Ingrowth1
- Positioning Failure1
- Unstable1
Patient problems
- Failure of Implant1
- No Code Available1
- No Known Impact Or Consequence To Patient1
- Nonunion/Delayed-union Bone Fracture1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2026Unstable; Loosening of Implant Not Related to Bone-Ingrowth
THE SURGEON GAVE PERMISSION TO MOVE, BUT CARRYING HEAVY OBJECTS WAS PROHIBITED. THE BONE HAD NOT HEALED, AND THE PATIENT DID NOT FOLLOW THE POSTOPERATIVE INSTRUCTIONS GIVEN BY THE PHYSICIAN AND CARRIED HEAVY OBJECTS. SUBSEQUENTLY, THE PATIENT DEVELOPED PAIN, AND IT WAS FOUND THAT THE DISTAL SCREW HAD DISENGAGED. NON-UNION WAS NOTICED. THE FOLLOWING SCREWS WERE REPORTED: (b)(6). IT WAS REPORTED THAT THEY WERE UNABLE TO DETERMINE WHICH SCREW WAS LOOSE. HOWEVER, IT WAS REPORTEDLY THE DISTAL SCREW, ALTHOUGH IT IS UNKNOWN WHICH SPECIFIC SCREW. AS A RESULT, A REVISION PROCEDURE WAS PERFORMED TO REPLACE THE PLATE AND THE SCREWS.
Patient problems: Nonunion/Delayed-union Bone Fracture; Pain
MAUDE report 24710461 ↗InjuryAug 13, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THE MOST MEDIAL SCREW LOOSENED FROM 3 WEEKS AFTER THE OPERATION, AND FINALLY THE LOCK WAS RELEASED AND ALL THE IMPLANTS CAME OFF. ON (B)(6) 2021, REVISION SURGERY WAS PERFORMED AND IT WAS REPLACED WITH A LONG PLATE.
Patient problems: Failure of Implant
MAUDE report 12318002 ↗MalfunctionAug 4, 2020Positioning Failure; Positioning Failure
AS REPORTED: "WHEN USING THE VARIAX LOCKING SCREW, THE PLATE AND SCREW DID NOT LOCK." ADDITIONALLY REPORTED: "ONLY ONE HOLE IN THE PLATE COULD NOT BE LOCKED. THE PROCEDURE ENDED WITHOUT INSERTING A SCREW INTO THE HOLE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10365319 ↗InjuryDec 3, 2018Device Dislodged or Dislocated
POSTOPERATIVE, IMPLANTED SCREW WAS BACKOUT . AS PRIMARY PROCEDURE, IMPLANTED THE FIRST 8 HOLE SUPERIOR ROW BEND PLATE, BUT REPLACES USING SUPERIOR LOW BEND 10 HOLES AS REVISION SURGERY.
Patient problems: No Code Available
MAUDE report 8126112 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page