AXSOS 703522Recalls & reported problems
Stryker GmbH · Plate Adapter Dist. Lat. Femur Right · Class 2
All models of AXSOS (184)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540693600), received between Mar 12, 2020 and Oct 25, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Fracture1
- Material Deformation1
Patient problems
- No Known Impact Or Consequence To Patient1
- Perforation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 25, 2024Break; Fracture; Break; Fracture
AS REPORTED BY REP: "CASE TYPE: RIGHT DISTAL FEMORAL PERIPROSTHETIC FRACTURE. PRH-TELFORD HAD ITEM: 703522 (PLATE ADAPTER / DIST LAT RIGHT) BREAK.". FURTHER INFORMATION FROM REP ON 09/27/2024: "WHEN ATTACHING ADAPTOR TO PLATE, ONE OF THE ¿TEETH¿ HAD SHEARED OFF RENDERING THE DEVICE UNSTABLE AND THEREFORE UNUSABLE. ATTEMPTED MINIMALLY INVASIVE TECHNIQUE IN COMPROMISED PATIENT; HOWEVER, ONCE THE ADAPTOR BROKE, THE SURGEON HAD TO MAKE A LARGE INCISION TO ATTACH THE PLATE TO THE BONE. DEBRIS REPORTED. INSTRUMENT BROKE AWAY FROM THE PATIENT. LARGE WOUND IN COMPROMISED PATIENT.
Patient problems: Perforation
MAUDE report 20536863 ↗MalfunctionMar 12, 2020Material Deformation; Material Deformation
REP LEFT A VOICEMAIL SAYING THE THREADS OF THE DEVICES STRIPPED INTRA-OPERATIVELY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9827488 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page