VARIAX 703822Recalls & reported problems
Stryker GmbH · Straight Reduction Clamp · Class 1
All models of VARIAX (345)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04546540703774), received between Oct 30, 2018 and Feb 21, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Fracture7
- Detachment of Device or Device Component1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 21, 2025Break; Fracture; Break; Fracture
AS REPORTED: "BROKEN TIP OF BONE GRASPING FORCEPS". THE PROCEDURE COULD BE COMPLETED SUCCESSFULLY. THE BROKEN TIP COULD BE RETRIEVED FROM THE PATIENT'S BODY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21437518 ↗MalfunctionAug 28, 2023Break; Fracture; Break; Fracture
AS REPORTED: "THE TIP OF THE BONE GRASPING FORCEPS WAS BROKEN WHEN USED TO REPAIR THE FRACTURE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17633403 ↗MalfunctionApr 11, 2023Break; Fracture; Break; Fracture
AS REPORTED: "THE INSTRUMENT BROKE THE TIP DURING SURGERY, THE FRAGMENT WAS REMOVED AND DISCARDED AT THE HOSPITAL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16715460 ↗MalfunctionSep 11, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGERY TO ADDRESS A RIGHT ANKLE FRACTURE, THE TIP OF THE CLAMP SNAPPED OFF IN THE PATIENT'S BONE. THE PIECE WAS RETRIEVED AND INTRA-OP X-RAY WAS USED TO ENSURE NO PIECES REMAINED IN THE PATIENT. SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10525199 ↗MalfunctionMay 15, 2020Fracture; Fracture
AS REPORTED: "DOCTOR CLAMPED DOWN ON FRAGMENT AND ONE OF THE ARMS OF THE CLAMP SNAPPED OFF."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10062228 ↗MalfunctionOct 21, 2019Break; Fracture; Fracture
BOTH POINT TO POINT FORCEPS ON THE VARIAX 2 FOOT AND ANKLE TRAY BROKE. THE BROKEN PIECES WERE REMOVED AND AN X-RAY CONFIRMED THEY WERE NO LONGER PRESENT IN THE PATIENT. THE SURGEON WAS AWARE AND CONTINUED THE CASE USING A HOSPITAL ALTERNATIVE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES OR SURGICAL DELAY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9216228 ↗MalfunctionOct 21, 2019Break; Fracture; Fracture
BOTH POINT TO POINT FORCEPS ON THE VARIAX 2 FOOT AND ANKLE TRAY BROKE. THE BROKEN PIECES WERE REMOVED AND AN X-RAY CONFIRMED THEY WERE NO LONGER PRESENT IN THE PATIENT. THE SURGEON WAS AWARE AND CONTINUED THE CASE USING A HOSPITAL ALTERNATIVE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES OR SURGICAL DELAY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9216224 ↗MalfunctionOct 31, 2018Break
DURING REDUCTION OF CLAVICLE USING STRAIGHT REDUCTION CLAMP, THE TIP OF THE CLAMP WAS BROKEN OFF. THE BROKEN PIECE OF THE CLAMP FELL OUTSIDE THE PATIENT. THE SURGERY WAS COMPLETED WITH ANOTHER CLAMP.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8025156 ↗MalfunctionOct 30, 2018Break; Detachment of Device or Device Component
DURING REDUCTION OF CLAVICLE USING STRAIGHT REDUCTION CLAMP, THE TIP OF THE CLAMP WAS BROKEN OFF. THE BROKEN PIECE OF THE CLAMP FELL OUTSIDE THE PATIENT. THE SURGERY WAS COMPLETED WITH ANOTHER CLAMP.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8021012 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page